Spatial resolution properties in cone beam CT: a simulation study.

Lingyun Chen1, Chris C Shaw, Mustafa C Altunbas, Chao-Jen Lai, Xinming Liu.   

Abstract

This work is intended to investigate the spatial resolution properties in cone beam CT by estimating the point spread functions (PSFs) in the reconstructed 3D images through simulation. The point objects were modeled as 3D delta functions. Their projections onto the detector plane were analytically derived and blurred with 2D PSFs estimated and used to represent the detector and focal spot blurring effects. The 2D PSF for detector blurring was computed from the line spread function measured for a typical a-Si/CsI flat panel detector used for general radiography. The focal spot blurring effect was simulated for an x-ray source with a nominal focal spot size of 0.6 mm and 1.33 x magnification at the rotating center. Projection images were computed and sampled with an interval significantly smaller than the detector pixel size to avoid aliasing. Images were reconstructed using the Feldkamp algorithm with the five different filter functions. Reconstructed PSFs were plotted and analyzed to investigate the effects of detector blurring alone, focal spot blurring alone, or a combination of the two on the PSFs and their variations with the radial distance and z-level. Effects of binning and reconstruction filters were also studied. Our results show that the PSFs due to detector blurring are largely symmetric and vary little with the locations of the point objects. With focal spot blurring only or added to detector blurring, the PSFs along the rotation axis were largely symmetric but became increasingly asymmetric as the point objects were moved away from the rotation axis. The PSFs were found to become wider in the axial (anode to cathode) direction as the objects were moved toward the cathode side. The 3D PSFs may be approximated by an ellipsoid with three different axial lengths. They were found to point upright along the rotating axis but tilt toward the rotating axis as the point object was moved away from the axis.

Entities:  

Mesh:

Year:  2008        PMID: 18383694      PMCID: PMC2857694          DOI: 10.1118/1.2829867

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: image-based correction of gantry motion nonidealities.

Authors:  R Fahrig; D W Holdsworth
Journal:  Med Phys       Date:  2000-01       Impact factor: 4.071

2.  Cone-beam computed tomography with a flat-panel imager: initial performance characterization.

Authors:  D A Jaffray; J H Siewerdsen
Journal:  Med Phys       Date:  2000-06       Impact factor: 4.071

3.  Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation.

Authors:  Yong Yang; Eduard Schreibmann; Tianfang Li; Chuang Wang; Lei Xing
Journal:  Phys Med Biol       Date:  2007-01-12       Impact factor: 3.609

4.  Evaluation of the spatial resolution characteristics of a cone-beam breast CT scanner.

Authors:  Alexander L C Kwan; John M Boone; Kai Yang; Shih-Ying Huang
Journal:  Med Phys       Date:  2007-01       Impact factor: 4.071

5.  Derivation and analysis of a filtered backprojection algorithm for cone beam projection data.

Authors:  X H Yan; R M Leahy
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

6.  Measurement of the PSF for a CT scanner: appropriate wire diameter and pixel size.

Authors:  E L Nickoloff
Journal:  Phys Med Biol       Date:  1988-01       Impact factor: 3.609

7.  Evaluation of the spatial resolution of a CT scanner by direct analysis of the edge response function.

Authors:  S M Bentzen
Journal:  Med Phys       Date:  1983 Sep-Oct       Impact factor: 4.071

8.  Intraoperative cone-beam CT for guidance of head and neck surgery: Assessment of dose and image quality using a C-arm prototype.

Authors:  M J Daly; J H Siewerdsen; D J Moseley; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

9.  A-Si:H/CsI(Tl) flat-panel versus computed radiography for chest imaging applications: image quality metrics measurement.

Authors:  Xinming Liu; Chris C Shaw
Journal:  Med Phys       Date:  2004-01       Impact factor: 4.071

10.  Cone-beam CT with megavoltage beams and an amorphous silicon electronic portal imaging device: potential for verification of radiotherapy of lung cancer.

Authors:  E C Ford; J Chang; K Mueller; K Sidhu; D Todor; G Mageras; E Yorke; C C Ling; H Amols
Journal:  Med Phys       Date:  2002-12       Impact factor: 4.071

View more
  5 in total

1.  Radiation doses in cone-beam breast computed tomography: a Monte Carlo simulation study.

Authors:  Ying Yi; Chao-Jen Lai; Tao Han; Yuncheng Zhong; Youtao Shen; Xinming Liu; Shuaiping Ge; Zhicheng You; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

2.  Impact of flat panel-imager veiling glare on scatter-estimation accuracy and image quality of a commercial on-board cone-beam CT imaging system.

Authors:  Dimitrios Lazos; Jeffrey F Williamson
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

3.  Cone beam CT for determining breast cancer margin: an initial experience and its comparison with mammography and specimen radiograph.

Authors:  Juan Yao; Chris Shaw; C J Lai; John Rong; Jian Wang; Wenya Liu
Journal:  Int J Clin Exp Med       Date:  2015-09-15

4.  Influence of acquisition parameters on MV-CBCT image quality.

Authors:  Olivier Gayou
Journal:  J Appl Clin Med Phys       Date:  2012-01-05       Impact factor: 2.102

5.  Tridimensional assessment of the dental follicle dimensions of impacted mandibular third molars using cone-beam CT.

Authors:  Marlene Barroso; Luis-Ernesto Arriola-Guillén; Yalil-Augusto Rodríguez-Cárdenas; Gustavo-Armando Ruíz-Mora; Maria-Eugenia Guerrero; Carlos Flores-Mir
Journal:  J Clin Exp Dent       Date:  2018-08-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.