Monte Carlo and phantom study of the radiation dose to the body from dedicated CT of the breast.

Ioannis Sechopoulos1, Srinivasan Vedantham, Sankararaman Suryanarayanan, Carl J D'Orsi, Andrew Karellas.   

Abstract

PURPOSE: To prospectively determine the radiation dose absorbed by the organs and tissues of the body during a dedicated breast computed tomography (CT) study by using Monte Carlo methods and a phantom.
MATERIALS AND METHODS: By using the Geant4 Monte Carlo tool kit, the Cristy anthropomorphic phantom and the geometry of a dedicated breast CT prototype were simulated. The simulation was used to track x-rays emitted from the source until their complete absorption or exit from the simulation limits. The interactions of the x-rays with the 65 different volumes representing organs, bones, and other tissues of the phantom that resulted in energy deposition were recorded. These data were used to compute the radiation dose to the organs and tissues during a complete dedicated breast CT scan relative to the average glandular dose to the imaged breast (relative organ dose [ROD]), by using the x-ray spectra proposed for dedicated breast CT imaging. The effectiveness of a lead shield for reducing the dose to the organs was investigated.
RESULTS: The maximum ROD among the organs was for the ipsilateral lung with a maximum ROD of 3.25%, followed by ROD for the heart and the thymus. Of the skeletal tissues, the sternum received the highest dose with a maximum ROD to the bone marrow of 2.24% and to the bone surface of 7.74%. The maximum ROD to the uterus, representative of that of an early-stage fetus, was 0.026%. These maxima occurred for the highest-energy x-ray spectrum (80 kVp) that was analyzed. A lead shield does not substantially protect the organs that receive the highest dose from dedicated breast CT.
CONCLUSION: Although the dose to the organs from dedicated breast CT is substantially higher than that from planar mammography, it is comparable to or considerably lower than that reached by other radiographic procedures and much lower than that of other CT examinations.

Entities:  

Mesh:

Year:  2008        PMID: 18292479      PMCID: PMC2430761          DOI: 10.1148/radiol.2471071080

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  22 in total

1.  Glandular breast dose for monoenergetic and high-energy X-ray beams: Monte Carlo assessment.

Authors:  J M Boone
Journal:  Radiology       Date:  1999-10       Impact factor: 11.105

2.  Dedicated breast CT: radiation dose and image quality evaluation.

Authors:  J M Boone; T R Nelson; K K Lindfors; J A Seibert
Journal:  Radiology       Date:  2001-12       Impact factor: 11.105

3.  Beyond randomized controlled trials: organized mammographic screening substantially reduces breast carcinoma mortality.

Authors:  Daniel B Kopans
Journal:  Cancer       Date:  2002-01-15       Impact factor: 6.860

4.  Dosimetric considerations in mammography.

Authors:  K A Hatziioannou; K Psarrakos; E Molyvda-Athanasopoulou; G Kitis; E Papanastassiou; I Sofroniadis; O Kimoundri
Journal:  Eur Radiol       Date:  2000       Impact factor: 5.315

5.  Why should breast tumour detection go three dimensional?

Authors:  Zikuan Chen; Ruola Ning
Journal:  Phys Med Biol       Date:  2003-07-21       Impact factor: 3.609

6.  A comprehensive analysis of DgN(CT) coefficients for pendant-geometry cone-beam breast computed tomography.

Authors:  J M Boone; N Shah; T R Nelson
Journal:  Med Phys       Date:  2004-02       Impact factor: 4.071

7.  Normalized glandular dose (DgN) coefficients for flat-panel CT breast imaging.

Authors:  Samta C Thacker; Stephen J Glick
Journal:  Phys Med Biol       Date:  2004-12-21       Impact factor: 3.609

8.  Cone-beam volume CT breast imaging: feasibility study.

Authors:  Biao Chen; Ruola Ning
Journal:  Med Phys       Date:  2002-05       Impact factor: 4.071

9.  Photoelectron enhancement of the absorbed dose from X rays to human bone marrow: experimental and theoretical studies.

Authors:  S D King; F W Spiers
Journal:  Br J Radiol       Date:  1985-04       Impact factor: 3.039

10.  Effective doses to patients undergoing thoracic computed tomography examinations.

Authors:  W Huda; E M Scalzetti; M Roskopf
Journal:  Med Phys       Date:  2000-05       Impact factor: 4.071

View more
  16 in total

1.  Evaluation of an improved algorithm for producing realistic 3D breast software phantoms: application for mammography.

Authors:  K Bliznakova; S Suryanarayanan; A Karellas; N Pallikarakis
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

2.  Dosimetric characterization of a dedicated breast computed tomography clinical prototype.

Authors:  Ioannis Sechopoulos; Steve Si Jia Feng; Carl J D'Orsi
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

3.  Average glandular dose coefficients for pendant-geometry breast CT using realistic breast phantoms.

Authors:  Andrew M Hernandez; John M Boone
Journal:  Med Phys       Date:  2017-08-20       Impact factor: 4.071

4.  Investigation of x-ray spectra for iodinated contrast-enhanced dedicated breast CT.

Authors:  Stephen J Glick; Andrey Makeev
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-26

5.  X-ray scatter correction method for dedicated breast computed tomography.

Authors:  Ioannis Sechopoulos
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

6.  Radiation doses in cone-beam breast computed tomography: a Monte Carlo simulation study.

Authors:  Ying Yi; Chao-Jen Lai; Tao Han; Yuncheng Zhong; Youtao Shen; Xinming Liu; Shuaiping Ge; Zhicheng You; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

7.  Non-Gaussian statistical properties of breast images.

Authors:  Craig K Abbey; Anita Nosrateih; Jascha Sohl-Dickstein; Kai Yang; John M Boone
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

8.  Simulated lesion, human observer performance comparison between thin-section dedicated breast CT images versus computed thick-section simulated projection images of the breast.

Authors:  L Chen; J M Boone; C K Abbey; J Hargreaves; C Bateni; K K Lindfors; K Yang; A Nosratieh; A Hernandez; P Gazi
Journal:  Phys Med Biol       Date:  2015-03-31       Impact factor: 3.609

9.  Library based x-ray scatter correction for dedicated cone beam breast CT.

Authors:  Linxi Shi; Srinivasan Vedantham; Andrew Karellas; Lei Zhu
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

Review 10.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.