Comparison of iCare tonometer and Goldmann applanation tonometry in normal corneas and in eyes with automated lamellar and penetrating keratoplasty.

M L Salvetat1, M Zeppieri, F Miani, C Tosoni, L Parisi, P Brusini.   

Abstract

PURPOSE: To compare intraocular pressure (IOP) measurements with Goldmann applanation tonometry (GAT) and iCare tonometry in normal and post-keratoplasty corneas and to assess the influence of central corneal thickness (CCT), corneal curvature (CC), and corneal astigmatism (CA) on IOP.
METHODS: This prospective cross-sectional study included one eye of 101 subjects with normal corneas (58 healthy subjects, 43 glaucoma); and 90 post-keratoplasty patients: 34 penetrating keratoplasties (PK); 20 automated-lamellar-therapeutic keratoplasties (ALTK); 19 Descemet-stripping-automated-endothelial keratoplasties (DSAEK); 17 edematous grafts. All subjects underwent GAT and iCare IOP measurements in random order, and CCT, CC, and CA evaluation. The Bland-Altman method and multivariate regression analysis were used to assess inter-tonometer agreement and the influence of CCT, CC, and CA on IOP.
RESULTS: iCare significantly underestimated IOP in all groups compared with GAT (GAT minus iCare of 3.5±3.5 mm Hg, P<0.001), but overestimated IOP in the edematous grafts (GAT minus iCare of -6.5±1.9 mm Hg, P<0.001). In normal corneas, both tonometer measurements were directly related to CCT values; iCare readings appeared inversely related to CC. There was no significant relationship between IOP and CCT, CC and CA in post-keratoplasty eyes, except between CC and iCare measurements for PK eyes.
CONCLUSIONS: The agreement between GAT and iCare was clinically acceptable in control, ALTK and DSAEK groups, and poor in PK and edematous grafts eyes. In normal corneas, GAT was significantly affected by CCT; iCare was influenced by CCT and CC. The iCare appeared less influenced by corneal edema when compared with GAT. High IOP readings taken with both tonometers in grafts should raise suspicion of true elevated IOP.

Entities:  

Mesh:

Year:  2011        PMID: 21436848      PMCID: PMC3171271          DOI: 10.1038/eye.2011.60

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  37 in total

1.  Descemet's stripping with endothelial keratoplasty in 200 eyes: Early challenges and techniques to enhance donor adherence.

Authors:  Francis W Price; Marianne O Price
Journal:  J Cataract Refract Surg       Date:  2006-03       Impact factor: 3.351

2.  Determining in vivo biomechanical properties of the cornea with an ocular response analyzer.

Authors:  David A Luce
Journal:  J Cataract Refract Surg       Date:  2005-01       Impact factor: 3.351

3.  Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis.

Authors:  Jun Liu; Cynthia J Roberts
Journal:  J Cataract Refract Surg       Date:  2005-01       Impact factor: 3.351

4.  Effects of corneal thickness, curvature, astigmatism and direction of gaze on Goldmann applanation tonometry readings.

Authors:  Gunilla Rask; Anders Behndig
Journal:  Ophthalmic Res       Date:  2005-11-16       Impact factor: 2.892

5.  Effects of antiglaucoma drops on MMP and TIMP balance in conjunctival and subconjunctival tissue.

Authors:  Tadashi Ito; Hiroshi Ohguro; Kazuhisa Mamiya; Ikuyo Ohguro; Mitsuru Nakazawa
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-03       Impact factor: 4.799

6.  Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry.

Authors:  Markus Kohlhaas; Andreas G Boehm; Eberhard Spoerl; Antje Pürsten; Hans J Grein; Lutz E Pillunat
Journal:  Arch Ophthalmol       Date:  2006-04

7.  Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers.

Authors:  Makoto Nakamura; Urtogtah Darhad; Yasuko Tatsumi; Miyuki Fujioka; Azusa Kusuhara; Hidetaka Maeda; Akira Negi
Journal:  Am J Ophthalmol       Date:  2006-08       Impact factor: 5.258

8.  Reproducibility and clinical evaluation of rebound tonometry.

Authors:  Jose M Martinez-de-la-Casa; Julian Garcia-Feijoo; Alfredo Castillo; Julian Garcia-Sanchez
Journal:  Invest Ophthalmol Vis Sci       Date:  2005-12       Impact factor: 4.799

9.  Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients.

Authors:  Paolo Brusini; Maria Letizia Salvetat; Marco Zeppieri; Claudia Tosoni; Lucia Parisi
Journal:  J Glaucoma       Date:  2006-06       Impact factor: 2.503

10.  Comparison of posterior lamellar keratoplasty techniques to penetrating keratoplasty.

Authors:  Irit Bahar; Igor Kaiserman; Penny McAllum; Allan Slomovic; David Rootman
Journal:  Ophthalmology       Date:  2008-04-28       Impact factor: 12.079

View more
  14 in total

1.  IOP agreement between I-Care TA01 rebound tonometer and the Goldmann applanation tonometer in eyes with and without glaucoma.

Authors:  Gautam Sinha; Shikha Gupta; Shreyas Temkar; Veena Pandey; Ramanjit Sihota; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2014-12-16       Impact factor: 2.031

2.  Impact of corneal parameters on intraocular pressure measurements in different tonometry methods.

Authors:  Aleksandra Zakrzewska; Marta P Wiącek; Anna Machalińska
Journal:  Int J Ophthalmol       Date:  2019-12-18       Impact factor: 1.779

Review 3.  Glaucoma management after corneal transplantation surgeries.

Authors:  Helen L Kornmann; Steven J Gedde
Journal:  Curr Opin Ophthalmol       Date:  2016-03       Impact factor: 3.761

4.  Commentary: Glaucoma following penetrating keratoplasty: A double whammy.

Authors:  Mano Ranjan Das; Ramaswami Krishnadas
Journal:  Indian J Ophthalmol       Date:  2022-04       Impact factor: 2.969

5.  Intra-Ocular Pressure Measurement in a Patient with a Thin, Thick or Abnormal Cornea.

Authors:  Colin I Clement; Douglas G A Parker; Ivan Goldberg
Journal:  Open Ophthalmol J       Date:  2016-02-29

6.  Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation.

Authors:  Mahmoud Rateb; Mahmoud Abdel-Radi; Zeiad Eldaly; Mohamed Nagy Elmohamady; Asaad Noor El Din
Journal:  J Ophthalmol       Date:  2019-05-09       Impact factor: 1.909

7.  Risk Factors for the Development of Ocular Hypertension After Keratoplasty: A Systematic Review.

Authors:  Ilona Liesenborghs; Johannes S A G Schouten; Tos T J M Berendschot; Henny J M Beckers; Rudy M M A Nuijts; Nienke Visser; Carroll A B Webers
Journal:  Cornea       Date:  2020-03       Impact factor: 3.152

8.  The measurement of intraocular pressure over positive soft contact lenses by rebound tonometry.

Authors:  Fabrizio Zeri; Mario De Cusatis; Luigi Lupelli; Peter Graham Swann
Journal:  J Optom       Date:  2015-10-21

Review 9.  Glaucoma after penetrating keratoplasty.

Authors:  Mihail Zemba; Alina-Cristina Stamate
Journal:  Rom J Ophthalmol       Date:  2017 Jul-Sep

10.  Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty.

Authors:  Oded Ohana; David Varssano; Gabi Shemesh
Journal:  Indian J Ophthalmol       Date:  2017-07       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.