Reproducibility and clinical evaluation of rebound tonometry.

Jose M Martinez-de-la-Casa1, Julian Garcia-Feijoo, Alfredo Castillo, Julian Garcia-Sanchez.   

Abstract

PURPOSE: To establish the reproducibility of a rebound tonometer in humans and the effect of corneal thickness on measurements, comparing it with Goldmann applanation tonometer.
METHODS: In a first study designed to examine the reliability of the RBT, three experienced ophthalmologists undertook three consecutive intraocular pressure (IOP) measurements in 12 eyes of 12 normal subjects. A cross-sectional study was then performed to compare measurements obtained using the two tonometers in 147 eyes of 85 patients with ocular hypertension or glaucoma.
RESULTS: Intraobserver coefficients of correlation obtained in the reproducibility study were 0.82, 0.73, and 0.87. Interobserver correlation was 0.82. There was a good correlation between IOP readings obtained by the RBT and the GAT (r = 0.865, P < 0.0001). RBT readings were consistently higher than GAT measurements (median difference, 1.8 +/- 2.8 mm Hg). A Bland-Altman plot indicated the 95% limits of agreement between the two methods were -3.7 to 7.3 mm Hg (slope = -0.022, P = 0.618). Using RBT, the point that best discriminated between patients with an IOP < or = 21 mm Hg and those with >21 mm Hg, as determined by the GAT was >23 mm Hg (sensitivity, 70.5%; specificity, 95.1%). In terms of pachymetry, the two tonometers behaved in a similar way, with correlation observed between IOP measurements and central corneal thickness.
CONCLUSIONS: Rebound tonometry is a reproducible method of determining IOP in humans. In general, it tends to overestimate IOP compared with Goldmann applanation tonometry. The tonometers used in both methods are similarly affected by pachymetry.

Entities:  

Mesh:

Year:  2005        PMID: 16303951     DOI: 10.1167/iovs.05-0586

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  41 in total

1.  Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects.

Authors:  Ioannis Halkiadakis; Aimilianos Stratos; George Stergiopoulos; Eleni Patsea; Sotiris Skouriotis; Panagiotis Mitropoulos; Dimitrios Papaconstantinou; Gerasimos Georgopoulos
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08       Impact factor: 3.117

2.  Effects of trabeculectomy on posture-induced intraocular pressure changes over time.

Authors:  Akira Sawada; Tetsuya Yamamoto
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-02-10       Impact factor: 3.117

3.  Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties.

Authors:  J Moreno-Montañés; I Gosende; J Caire; M García-Granero
Journal:  Eye (Lond)       Date:  2010-10-01       Impact factor: 3.775

Review 4.  Discovery of Molecular Therapeutics for Glaucoma: Challenges, Successes, and Promising Directions.

Authors:  Rebecca K Donegan; Raquel L Lieberman
Journal:  J Med Chem       Date:  2015-09-25       Impact factor: 7.446

Review 5.  Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.

Authors:  A Cervino
Journal:  Br J Ophthalmol       Date:  2006-12       Impact factor: 4.638

6.  Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Fatih Özcura; Nilgün Yildirim; Afsun Şahin; Ertuğrul Çolak
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

7.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

8.  IOP agreement between I-Care TA01 rebound tonometer and the Goldmann applanation tonometer in eyes with and without glaucoma.

Authors:  Gautam Sinha; Shikha Gupta; Shreyas Temkar; Veena Pandey; Ramanjit Sihota; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2014-12-16       Impact factor: 2.031

9.  [Rebound tonometry after vitreoretinal surgery].

Authors:  A Rosentreter; R Hoerster; T Schick; N Eter; T S Dietlein; S Fauser
Journal:  Ophthalmologe       Date:  2015-11       Impact factor: 1.059

10.  Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study.

Authors:  Senay Asik Nacaroglu; Emine Seker Un; Mehmet Giray Ersoz; Yelda Tasci
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.