Calibration-free device sizing using an inverse geometry x-ray system.

Michael T Tomkowiak1, Michael A Speidel, Amish N Raval, Michael S Van Lysel.   

Abstract

PURPOSE: Quantitative coronary angiography (QCA) can be used to support device size selection for cardiovascular interventions. The accuracy of QCA measurements using conventional x-ray fluoroscopy depends on proper calibration using a reference object and avoiding vessel foreshortening. The authors have developed a novel interventional device sizing method using the inverse geometry scanning-beam digital x-ray (SBDX) fluoroscopy system. The proposed method can measure the diameter and length of vessel segments without imaging a reference object and when vessels appear foreshortened.
METHODS: SBDX creates multiple tomosynthetic x-ray images corresponding to planes through the patient volume. The structures that lie in the plane are in focus and the features above and below the plane are blurred. Three-dimensional localization of the vessel edges was performed by examining the degree of blurring at each image plane. A 3D vessel centerline was created and used to determine vessel magnification and angulation relative to the image planes. Diameter measurements were performed using a model-based method and length measurements were calculated from the 3D centerline. Phantom validation was performed by measuring the diameter and length of vessel segments with nominal diameters ranging from 0.5 to 2.8 mm and nominal lengths of 42 mm. The phantoms were imaged at a range of positions between the source and the detector (+/- 16 cm relative to isocenter) and with a range of foreshortening angles (0 degrees-75 degrees).
RESULTS: Changes in vessel phantom position created magnifications ranging from 87% to 118% relative to isocenter magnification. Average diameter errors were less than 0.15 mm. Average length measurements were within 1% (0.3 mm) of the true length. No trends were observed between measurement accuracy and magnification. Changes in vessel phantom orientation resulted in decreased apparent length down to 28% of the original nonforeshortened length. Average diameter errors were less than 0.25 mm across all vessel angulations; errors were less than 0.1 mm for smaller diameter vessels and low to moderate vessel angles. Diameter errors increased with true diameter and vessel angle relative to the image plane. Average length measurement errors were also within 1% (0.3 mm) for each angulation.
CONCLUSIONS: Tomosynthetic imaging with SBDX can accurately measure dimensions of vessels in various magnifications and angulations without calibration. This method may be more accurate and convenient than conventional QCA techniques.

Entities:  

Mesh:

Year:  2011        PMID: 21361197      PMCID: PMC3021558          DOI: 10.1118/1.3528227

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

Review 1.  A new model-based technique for enhanced small-vessel measurements in X-ray ciné-angiograms.

Authors:  R C Chan; W C Karl; R S Lees
Journal:  IEEE Trans Med Imaging       Date:  2000-03       Impact factor: 10.048

2.  3-D reconstruction of coronary arterial tree to optimize angiographic visualization.

Authors:  S J Chen; J D Carroll
Journal:  IEEE Trans Med Imaging       Date:  2000-04       Impact factor: 10.048

3.  High-quality 3-D coronary artery imaging on an interventional C-arm x-ray system.

Authors:  Eberhard Hansis; John D Carroll; Dirk Schäfer; Olaf Dössel; Michael Grass
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

4.  Impact of stent overlap on angiographic and long-term clinical outcome in patients undergoing drug-eluting stent implantation.

Authors:  Lorenz Räber; Peter Jüni; Lukas Löffel; Simon Wandel; Stéphane Cook; Peter Wenaweser; Mario Togni; Rolf Vogel; Christian Seiler; Franz Eberli; Thomas Lüscher; Bernhard Meier; Stephan Windecker
Journal:  J Am Coll Cardiol       Date:  2010-03-23       Impact factor: 24.094

5.  Impact of multiple and long sirolimus-eluting stent implantation on 3-year clinical outcomes in the j-Cypher Registry.

Authors:  Shinichi Shirai; Takeshi Kimura; Masakiyo Nobuyoshi; Takeshi Morimoto; Kenji Ando; Yoshimitsu Soga; Kyohei Yamaji; Katsuhiro Kondo; Koyu Sakai; Takeshi Arita; Masahiko Goya; Masashi Iwabuchi; Hiroyoshi Yokoi; Hideyuki Nosaka; Kazuaki Mitsudo
Journal:  JACC Cardiovasc Interv       Date:  2010-02       Impact factor: 11.195

6.  Automatic generation of 3D coronary artery centerlines using rotational X-ray angiography.

Authors:  Uwe Jandt; Dirk Schäfer; Michael Grass; Volker Rasche
Journal:  Med Image Anal       Date:  2009-08-03       Impact factor: 8.545

7.  Optimal stent-sizing with intravascular ultrasound contributes to complete neointimal coverage after sirolimus-eluting stent implantation assessed by angioscopy.

Authors:  Fusako Sera; Masaki Awata; Masaaki Uematsu; Jun-ichi Kotani; Shinsuke Nanto; Seiki Nagata
Journal:  JACC Cardiovasc Interv       Date:  2009-10       Impact factor: 11.195

8.  Three-dimensional tracking of cardiac catheters using an inverse geometry x-ray fluoroscopy system.

Authors:  Michael A Speidel; Michael T Tomkowiak; Amish N Raval; Michael S Van Lysel
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

9.  Continued improvement of clinical outcome and cost effectiveness following intravascular ultrasound guided PCI: insights from a prospective, randomised study.

Authors:  A L Gaster; U Slothuus Skjoldborg; J Larsen; L Korsholm; C von Birgelen; S Jensen; P Thayssen; K E Pedersen; T H Haghfelt
Journal:  Heart       Date:  2003-09       Impact factor: 5.994

10.  Predictors of subacute stent thrombosis: results of a systematic intravascular ultrasound study.

Authors:  Edouard Cheneau; Laurent Leborgne; Gary S Mintz; Jun-ichi Kotani; Augusto D Pichard; Lowell F Satler; Daniel Canos; Marco Castagna; Neil J Weissman; Ron Waksman
Journal:  Circulation       Date:  2003-06-23       Impact factor: 29.690

View more
  5 in total

1.  Inverse geometry x-ray imaging: application in interventional procedures.

Authors:  Michael A Speidel
Journal:  J Am Coll Radiol       Date:  2011-01       Impact factor: 5.532

2.  X-ray and optical stereo-based 3D sensor fusion system for image-guided neurosurgery.

Authors:  Duk Nyeon Kim; You Seong Chae; Min Young Kim
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-09-26       Impact factor: 2.924

3.  Real-time out-of-plane artifact subtraction tomosynthesis imaging using prior CT for scanning beam digital x-ray system.

Authors:  Meng Wu; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

4.  Calibration-free coronary artery measurements for interventional device sizing using inverse geometry x-ray fluoroscopy: in vivo validation.

Authors:  Michael T Tomkowiak; Amish N Raval; Michael S Van Lysel; Tobias Funk; Michael A Speidel
Journal:  J Med Imaging (Bellingham)       Date:  2014-10

5.  Calibration-Free Coronary Artery Measurements for Interventional Device Sizing using Inverse Geometry X-ray Fluoroscopy: In Vivo Validation.

Authors:  Michael T Tomkowiak; Amish N Raval; Michael S Van Lysel; Tobias Funk; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.