Real-time out-of-plane artifact subtraction tomosynthesis imaging using prior CT for scanning beam digital x-ray system.

Meng Wu1, Rebecca Fahrig2.   

Abstract

PURPOSE: The scanning beam digital x-ray system (SBDX) is an inverse geometry fluoroscopic system with high dose efficiency and the ability to perform continuous real-time tomosynthesis in multiple planes. This system could be used for image guidance during lung nodule biopsy. However, the reconstructed images suffer from strong out-of-plane artifact due to the small tomographic angle of the system.
METHODS: The authors propose an out-of-plane artifact subtraction tomosynthesis (OPAST) algorithm that utilizes a prior CT volume to augment the run-time image processing. A blur-and-add (BAA) analytical model, derived from the project-to-backproject physical model, permits the generation of tomosynthesis images that are a good approximation to the shift-and-add (SAA) reconstructed image. A computationally practical algorithm is proposed to simulate images and out-of-plane artifacts from patient-specific prior CT volumes using the BAA model. A 3D image registration algorithm to align the simulated and reconstructed images is described. The accuracy of the BAA analytical model and the OPAST algorithm was evaluated using three lung cancer patients' CT data. The OPAST and image registration algorithms were also tested with added nonrigid respiratory motions.
RESULTS: Image similarity measurements, including the correlation coefficient, mean squared error, and structural similarity index, indicated that the BAA model is very accurate in simulating the SAA images from the prior CT for the SBDX system. The shift-variant effect of the BAA model can be ignored when the shifts between SBDX images and CT volumes are within ±10 mm in the x and y directions. The nodule visibility and depth resolution are improved by subtracting simulated artifacts from the reconstructions. The image registration and OPAST are robust in the presence of added respiratory motions. The dominant artifacts in the subtraction images are caused by the mismatches between the real object and the prior CT volume.
CONCLUSIONS: Their proposed prior CT-augmented OPAST reconstruction algorithm improves lung nodule visibility and depth resolution for the SBDX system.

Entities:  

Mesh:

Year:  2014        PMID: 25370638      PMCID: PMC4218694          DOI: 10.1118/1.4896818

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  26 in total

Review 1.  Digital x-ray tomosynthesis: current state of the art and clinical potential.

Authors:  James T Dobbins; Devon J Godfrey
Journal:  Phys Med Biol       Date:  2003-10-07       Impact factor: 3.609

2.  Image quality assessment: from error visibility to structural similarity.

Authors:  Zhou Wang; Alan Conrad Bovik; Hamid Rahim Sheikh; Eero P Simoncelli
Journal:  IEEE Trans Image Process       Date:  2004-04       Impact factor: 10.856

3.  A comparison of reconstruction algorithms for breast tomosynthesis.

Authors:  Tao Wu; Richard H Moore; Elizabeth A Rafferty; Daniel B Kopans
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

4.  Respiratory motion estimation from slowly rotating x-ray projections: theory and simulation.

Authors:  Rongping Zeng; Jeffrey A Fessler; James M Balter
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

5.  Optimization of a tomosynthesis system for the detection of lung nodules.

Authors:  Angel R Pineda; Sungwon Yoon; David S Paik; Rebecca Fahrig
Journal:  Med Phys       Date:  2006-05       Impact factor: 4.071

6.  Comparison of entrance exposure and signal-to-noise ratio between an SBDX prototype and a wide-beam cardiac angiographic system.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Timothy D Betts; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

7.  Optimization of slice sensitivity profile for radiographic tomosynthesis.

Authors:  Baojun Li; Gopal B Avinash; Jeffrey W Eberhard; Bernhard E H Claus
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

8.  A theoretical study of some maximum likelihood algorithms for emission and transmission tomography.

Authors:  K Lange; M Bahn; R Little
Journal:  IEEE Trans Med Imaging       Date:  1987       Impact factor: 10.048

9.  A technique for estimating 4D-CBCT using prior knowledge and limited-angle projections.

Authors:  You Zhang; Fang-Fang Yin; W Paul Segars; Lei Ren
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

10.  Calibration-free device sizing using an inverse geometry x-ray system.

Authors:  Michael T Tomkowiak; Michael A Speidel; Amish N Raval; Michael S Van Lysel
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.