Literature DB >> 24999298

Calibration-Free Coronary Artery Measurements for Interventional Device Sizing using Inverse Geometry X-ray Fluoroscopy: In Vivo Validation.

Michael T Tomkowiak1, Amish N Raval2, Michael S Van Lysel3, Tobias Funk4, Michael A Speidel3.   

Abstract

Proper sizing of interventional devices to match coronary vessel dimensions improves procedural efficiency and therapeutic outcomes. We have developed a novel method using inverse geometry x-ray fluoroscopy to automatically determine vessel dimensions without the need for magnification calibration or optimal views. To validate this method in vivo, we compared results to intravascular ultrasound (IVUS) and coronary computed tomography angiography (CCTA) in a healthy porcine model. Coronary angiography was performed using Scanning-Beam Digital X-ray (SBDX), an inverse geometry fluoroscopy system that performs multiplane digital x-ray tomosynthesis in real time. From a single frame, 3D reconstruction of the arteries was performed by localizing the depth of vessel lumen edges. The 3D model was used to directly calculate length and to determine the best imaging plane to use for diameter measurements, where out-of-plane blur was minimized and the known pixel spacing was used to obtain absolute vessel diameter. End-diastolic length and diameter measurements were compared to measurements from CCTA and IVUS, respectively. For vessel segment lengths measuring 6 mm to 73 mm by CCTA, the SBDX length error was -0.49 ± 1.76 mm (SBDX - CCTA, mean ± 1 SD). For vessel diameters measuring 2.1 mm to 3.6 mm by IVUS, the SBDX diameter error was 0.07 ± 0.27 mm (SBDX - minimum IVUS diameter, mean ± 1 SD). The in vivo agreement between SBDX-based vessel sizing and gold standard techniques supports the feasibility of calibration-free coronary vessel sizing using inverse geometry x-ray fluoroscopy.

Entities:  

Keywords:  cardiac interventions; inverse geometry; quantitative coronary angiography; x-ray fluoroscopy

Year:  2014        PMID: 24999298      PMCID: PMC4079058          DOI: 10.1117/12.2044078

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  23 in total

1.  Vessel extraction in medical images by wave-propagation and traceback.

Authors:  F K Quek; C Kirbas
Journal:  IEEE Trans Med Imaging       Date:  2001-02       Impact factor: 10.048

2.  Assessment of coronary stent deployment using computer enhanced x-ray images-validation against intravascular ultrasound and best practice recommendations.

Authors:  Andrew G Davies; Dwayne Conway; Scott Reid; Arnold R Cowen; Mohan Sivananthan
Journal:  Catheter Cardiovasc Interv       Date:  2012-01-19       Impact factor: 2.692

3.  Stent underexpansion and residual reference segment stenosis are related to stent thrombosis after sirolimus-eluting stent implantation: an intravascular ultrasound study.

Authors:  Kenichi Fujii; Stéphane G Carlier; Gary S Mintz; Yi-ming Yang; Issam Moussa; Giora Weisz; George Dangas; Roxana Mehran; Alexandra J Lansky; Edward M Kreps; Michael Collins; Gregg W Stone; Jeffrey W Moses; Martin B Leon
Journal:  J Am Coll Cardiol       Date:  2005-04-05       Impact factor: 24.094

4.  Comparison of entrance exposure and signal-to-noise ratio between an SBDX prototype and a wide-beam cardiac angiographic system.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Timothy D Betts; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

5.  Automated analysis of coronary arterial morphology in cineangiograms: geometric and physiologic validation in humans.

Authors:  S R Fleagle; M R Johnson; C J Wilbricht; D J Skorton; R F Wilson; C W White; M L Marcus; S M Collins
Journal:  IEEE Trans Med Imaging       Date:  1989       Impact factor: 10.048

6.  Calibration-free device sizing using an inverse geometry x-ray system.

Authors:  Michael T Tomkowiak; Michael A Speidel; Amish N Raval; Michael S Van Lysel
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

7.  Morphometric analysis in three-dimensional intracoronary ultrasound: an in vitro and in vivo study performed with a novel system for the contour detection of lumen and plaque.

Authors:  C von Birgelen; C Di Mario; W Li; J C Schuurbiers; C J Slager; P J de Feyter; J R Roelandt; P W Serruys
Journal:  Am Heart J       Date:  1996-09       Impact factor: 4.749

8.  Multidetector-row computed tomography vs. angiography and intravascular ultrasound for the evaluation of the diameter of proximal coronary arteries.

Authors:  Anil-Martin Sinha; Andreas H Mahnken; Anja Borghans; Stefan Krüger; Ralf Koos; Katrin Dedden; Joachim E Wildberger; Rainer Hoffmann
Journal:  Int J Cardiol       Date:  2005-09-19       Impact factor: 4.164

9.  Influence of inflation pressure and balloon size on the development of intimal hyperplasia after balloon angioplasty. A study in the atherosclerotic rabbit.

Authors:  I J Sarembock; P J LaVeau; S L Sigal; I Timms; J Sussman; C Haudenschild; M D Ezekowitz
Journal:  Circulation       Date:  1989-10       Impact factor: 29.690

10.  Influence of balloon size on initial success, acute complications, and restenosis after percutaneous transluminal coronary angioplasty. A prospective randomized study.

Authors:  G S Roubin; J S Douglas; S B King; S F Lin; N Hutchison; R G Thomas; A R Gruentzig
Journal:  Circulation       Date:  1988-09       Impact factor: 29.690

View more
  1 in total

1.  Calibration-free coronary artery measurements for interventional device sizing using inverse geometry x-ray fluoroscopy: in vivo validation.

Authors:  Michael T Tomkowiak; Amish N Raval; Michael S Van Lysel; Tobias Funk; Michael A Speidel
Journal:  J Med Imaging (Bellingham)       Date:  2014-10
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.