Segmentation and quantification of materials with energy discriminating computed tomography: a phantom study.

Huy Q Le1, Sabee Molloi.   

Abstract

PURPOSE: To experimentally investigate whether a computed tomography (CT) system based on CdZnTe (CZT) detectors in conjunction with a least-squares parameter estimation technique can be used to decompose four different materials.
METHODS: The material decomposition process was divided into a segmentation task and a quantification task. A least-squares minimization algorithm was used to decompose materials with five measurements of the energy dependent linear attenuation coefficients. A small field-of-view energy discriminating CT system was built. The CT system consisted of an x-ray tube, a rotational stage, and an array of CZT detectors. The CZT array was composed of 64 pixels, each of which is 0.8 x 0.8 x 3 mm. Images were acquired at 80 kVp in fluoroscopic mode at 50 ms per frame. The detector resolved the x-ray spectrum into energy bins of 22-32, 33-39, 40-46, 47-56, and 57-80 keV. Four phantoms were constructed from polymethylmethacrylate (PMMA), polyethylene, polyoxymethylene, hydroxyapatite, and iodine. Three phantoms were composed of three materials with embedded hydroxyapatite (50, 150, 250, and 350 mg/ml) and iodine (4, 8, 12, and 16 mg/ml) contrast elements. One phantom was composed of four materials with embedded hydroxyapatite (150 and 350 mg/ml) and iodine (8 and 16 mg/ml). Calibrations consisted of PMMA phantoms with either hydroxyapatite (100, 200, 300, 400, and 500 mg/ml) or iodine (5, 15, 25, 35, and 45 mg/ml) embedded. Filtered backprojection and a ramp filter were used to reconstruct images from each energy bin. Material segmentation and quantification were performed and compared between different phantoms.
RESULTS: All phantoms were decomposed accurately, but some voxels in the base material regions were incorrectly identified. Average quantification errors of hydroxyapatite/iodine were 9.26/7.13%, 7.73/5.58%, and 12.93/8.23% for the three-material PMMA, polyethylene, and polyoxymethylene phantoms, respectively. The average errors for the four-material phantom were 15.62% and 2.76% for hydroxyapatite and iodine, respectively.
CONCLUSIONS: The calibrated least-squares minimization technique of decomposition performed well in breast imaging tasks with an energy resolving detector. This method can provide material basis images containing concentrations of the relevant materials that can potentially be valuable in the diagnostic process.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21361191      PMCID: PMC3021555          DOI: 10.1118/1.3525835

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  27 in total

1.  An analytic model for the response of a CZT detector in diagnostic energy dispersive x-ray spectroscopy.

Authors:  Robert J LeClair; Yinkun Wang; Peiying Zhao; Michel Boileau; Lilie Wang; Fabrice Fleurot
Journal:  Med Phys       Date:  2006-05       Impact factor: 4.071

2.  Evaluation of the minimum iodine concentration for contrast-enhanced subtraction mammography.

Authors:  P Baldelli; A Bravin; C Di Maggio; G Gennaro; A Sarnelli; A Taibi; M Gambaccini
Journal:  Phys Med Biol       Date:  2006-08-08       Impact factor: 3.609

3.  Technique factors and their relationship to radiation dose in pendant geometry breast CT.

Authors:  John M Boone; Alexander L C Kwan; J Anthony Seibert; Nikula Shah; Karen K Lindfors; Thomas R Nelson
Journal:  Med Phys       Date:  2005-12       Impact factor: 4.071

4.  K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors.

Authors:  E Roessl; R Proksa
Journal:  Phys Med Biol       Date:  2007-07-17       Impact factor: 3.609

5.  Feasibility of real time dual-energy imaging based on a flat panel detector for coronary artery calcium quantification.

Authors:  Tong Xu; Justin L Ducote; Jerry T Wong; Sabee Molloi
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

Review 6.  Tumour angiogenesis and its relation to contrast enhancement on computed tomography: a review.

Authors:  K A Miles
Journal:  Eur J Radiol       Date:  1999-06       Impact factor: 3.528

7.  Mammographically detected breast arterial calcifications: indicators for arteriosclerotic diseases?

Authors:  Füsun Taşkin; Alev Akdilli; Can Karaman; Alparslan Unsal; Kutsi Köseoğlu; Filiz Ergin
Journal:  Eur J Radiol       Date:  2006-08-04       Impact factor: 3.528

8.  Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies.

Authors:  W A Kalender; W H Perman; J R Vetter; E Klotz
Journal:  Med Phys       Date:  1986 May-Jun       Impact factor: 4.071

9.  Mammograms may convey more than breast cancer risk: breast arterial calcification and arterio-sclerotic related diseases in women of the DOM cohort.

Authors:  P A van Noord; D Beijerinck; J M Kemmeren; Y van der Graaf
Journal:  Eur J Cancer Prev       Date:  1996-12       Impact factor: 2.497

10.  The composition of body tissues.

Authors:  H Q Woodard; D R White
Journal:  Br J Radiol       Date:  1986-12       Impact factor: 3.039

View more
  13 in total

1.  Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study.

Authors:  H-M Cho; H Ding; B P Ziemer; S Molloi
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  Quantification of breast density with spectral mammography based on a scanned multi-slit photon-counting detector: a feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Phys Med Biol       Date:  2012-07-06       Impact factor: 3.609

3.  K-edge ratio method for identification of multiple nanoparticulate contrast agents by spectral CT imaging.

Authors:  H Ghadiri; M R Ay; M B Shiran; H Soltanian-Zadeh; H Zaidi
Journal:  Br J Radiol       Date:  2013-08-09       Impact factor: 3.039

4.  Energy dispersive CdTe and CdZnTe detectors for spectral clinical CT and NDT applications.

Authors:  W C Barber; J C Wessel; E Nygard; J S Iwanczyk
Journal:  Nucl Instrum Methods Phys Res A       Date:  2015-06-01       Impact factor: 1.455

5.  Model-based material decomposition with a penalized nonlinear least-squares CT reconstruction algorithm.

Authors:  Steven Tilley; Wojciech Zbijewski; J Webster Stayman
Journal:  Phys Med Biol       Date:  2019-01-22       Impact factor: 3.609

6.  Quantitative material characterization from multi-energy photon counting CT.

Authors:  Adam M Alessio; Lawrence R MacDonald
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

7.  Breast composition measurement with a cadmium-zinc-telluride based spectral computed tomography system.

Authors:  Huanjun Ding; Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

8.  Dual-energy cone-beam CT with a flat-panel detector: effect of reconstruction algorithm on material classification.

Authors:  W Zbijewski; G J Gang; J Xu; A S Wang; J W Stayman; K Taguchi; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

9.  Breast tissue decomposition with spectral distortion correction: a postmortem study.

Authors:  Huanjun Ding; Bo Zhao; Pavlo Baturin; Farnaz Behroozi; Sabee Molloi
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

10.  Microcalcification detectability using a bench-top prototype photon-counting breast CT based on a Si strip detector.

Authors:  Hyo-Min Cho; Huanjun Ding; William C Barber; Jan S Iwanczyk; Sabee Molloi
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071