Microcalcification detectability using a bench-top prototype photon-counting breast CT based on a Si strip detector.

Hyo-Min Cho1, Huanjun Ding1, William C Barber2, Jan S Iwanczyk2, Sabee Molloi1.   

Abstract

PURPOSE: To investigate the feasibility of detecting breast microcalcification (μCa) with a dedicated breast computed tomography (CT) system based on energy-resolved photon-counting silicon (Si) strip detectors.
METHODS: The proposed photon-counting breast CT system and a bench-top prototype photon-counting breast CT system were simulated using a simulation package written in matlab to determine the smallest detectable μCa. A 14 cm diameter cylindrical phantom made of breast tissue with 20% glandularity was used to simulate an average-sized breast. Five different size groups of calcium carbonate grains, from 100 to 180 μm in diameter, were simulated inside of the cylindrical phantom. The images were acquired with a mean glandular dose (MGD) in the range of 0.7-8 mGy. A total of 400 images was used to perform a reader study. Another simulation study was performed using a 1.6 cm diameter cylindrical phantom to validate the experimental results from a bench-top prototype breast CT system. In the experimental study, a bench-top prototype CT system was constructed using a tungsten anode x-ray source and a single line 256-pixels Si strip photon-counting detector with a pixel pitch of 100 μm. Calcium carbonate grains, with diameter in the range of 105-215 μm, were embedded in a cylindrical plastic resin phantom to simulate μCas. The physical phantoms were imaged at 65 kVp with an entrance exposure in the range of 0.6-8 mGy. A total of 500 images was used to perform another reader study. The images were displayed in random order to three blinded observers, who were asked to give a 4-point confidence rating on each image regarding the presence of μCa. The μCa detectability for each image was evaluated by using the average area under the receiver operating characteristic curve (AUC) across the readers.
RESULTS: The simulation results using a 14 cm diameter breast phantom showed that the proposed photon-counting breast CT system can achieve high detection accuracy with an average AUC greater than 0.89 ± 0.07 for μCas larger than 120 μm in diameter at a MGD of 3 mGy. The experimental results using a 1.6 cm diameter breast phantom showed that the prototype system can achieve an average AUC greater than 0.98 ± 0.01 for μCas larger than 140 μm in diameter using an entrance exposure of 1.2 mGy.
CONCLUSIONS: The proposed photon-counting breast CT system based on a Si strip detector can potentially offer superior image quality to detect μCa with a lower dose level than a standard two-view mammography.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26133636      PMCID: PMC4491016          DOI: 10.1118/1.4922680

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  35 in total

1.  Detective quantum efficiency dependence on x-ray energy weighting in mammography.

Authors:  R N Cahn; B Cederström; M Danielsson; A Hall; M Lundqvist; D Nygren
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

Review 2.  Digital x-ray tomosynthesis: current state of the art and clinical potential.

Authors:  James T Dobbins; Devon J Godfrey
Journal:  Phys Med Biol       Date:  2003-10-07       Impact factor: 3.609

3.  Microcalcification detection using cone-beam CT mammography with a flat-panel imager.

Authors:  Xing Gong; Aruna A Vedula; Stephen J Glick
Journal:  Phys Med Biol       Date:  2004-06-07       Impact factor: 3.609

4.  A comprehensive analysis of DgN(CT) coefficients for pendant-geometry cone-beam breast computed tomography.

Authors:  J M Boone; N Shah; T R Nelson
Journal:  Med Phys       Date:  2004-02       Impact factor: 4.071

5.  Scatter rejection in multislit digital mammography.

Authors:  Magnus Aslund; Björn Cederström; Mats Lundqvist; Mats Danielsson
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

Review 6.  Early detection of breast cancer: mammography.

Authors:  C J D'Orsi
Journal:  Breast Cancer Res Treat       Date:  1991-05       Impact factor: 4.872

7.  An improved method for simulating microcalcifications in digital mammograms.

Authors:  Federica Zanca; Dev Prasad Chakraborty; Chantal Van Ongeval; Jurgen Jacobs; Filip Claus; Guy Marchal; Hilde Bosmans
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

8.  Spectral optimization for dedicated breast CT.

Authors:  Michaela Weigel; Sabrina V Vollmar; Willi A Kalender
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

9.  Comprehensive assessment of the slice sensitivity profiles in breast tomosynthesis and breast CT.

Authors:  Anita Nosratieh; Kai Yang; Shadi Aminololama-Shakeri; John M Boone
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

10.  Occult malignant breast lesions in 114 patients: relationship to age and the presence of microcalcifications.

Authors:  G Hermann; C Janus; I S Schwartz; A Papatestas; D G Hermann; J G Rabinowitz
Journal:  Radiology       Date:  1988-11       Impact factor: 11.105

View more
  2 in total

1.  High-resolution μ CT imaging for characterizing microcalcification detection performance in breast CT.

Authors:  Andrew M Hernandez; Amy E Becker; Su Hyun Lyu; Craig K Abbey; John M Boone
Journal:  J Med Imaging (Bellingham)       Date:  2021-07-20

2.  Detection of Microcalcifications in Spiral Breast Computed Tomography with Photon-Counting Detector Is Feasible: A Specimen Study.

Authors:  Matthias Wetzl; Evelyn Wenkel; Eva Balbach; Ebba Dethlefsen; Arndt Hartmann; Julius Emons; Christiane Kuhl; Matthias W Beckmann; Michael Uder; Sabine Ohlmeyer
Journal:  Diagnostics (Basel)       Date:  2021-05-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.