Literature DB >> 21361181

Spectral optimization for dedicated breast CT.

Michaela Weigel1, Sabrina V Vollmar, Willi A Kalender.   

Abstract

PURPOSE: The authors investigated the choice of spectra for the optimization of the dose-weighted contrast-to-noise ratio (CNRD) for a dedicated breast CT scanner. The objective is to provide the desired image quality level at minimal dose values. The CNRD was investigated as a function of energy or tube voltage and filtrations for various breast sizes and contrasts.
METHODS: The authors performed simulations of the pendant female breast as cylinders consisting of a homogeneous mixture of adipose and glandular tissue with diameters from 6 to 18 cm. The contrasts of adipose tissue, calcium hydroxyapatite, and iodine contrast agent relative to glandular tissue were analyzed using inserts of 9 mm in diameter. Simulations were conducted for monochromatic and polychromatic radiation with a 3 mm Al or a 0.3 mm Cu filter. Simulations and measurements on an experimental micro-CT scanner were performed for validation purposes with a 6 cm water-equivalent cylinder, with inserts representing a pure density difference of 10%, calcium hydroxyapatite, and iodine contrast agent. A breast tissue sample embedded in paraffin was investigated to confirm the simulation results.
RESULTS: Optimal tube voltages were found to be in the range of 30-55 kV for breast CT imaging. For example, with 3 mm Al or 0.3 mm Cu filtration, optimal tube voltages were about 53 and 48 kV, respectively, for the contrast iodine/glandular tissue for all diameters. With 3 mm Al filtration, optimal tube voltages increased from 30 to 37 kV and from 30 to 47 kV for the contrast calcium hydroxyapatite/glandular tissue and adipose/glandular tissue, respectively. Tube power requirements were estimated and the change in tube current relative to 80 kV was found to be between 4 and 14 and between 11 and 214 with 3 mm Al filtration or 0.3 mm Cu filtration, respectively. These numbers show that realizing the low optimal tube voltages may not be feasible in most cases due to power requirements.
CONCLUSIONS: Depending on the filtration, the authors assume that a compromise solution has to be found between the highest potential dose reduction and a solution working with available x-ray sources. In view of the tube power constraints, the authors recommend aiming for tube voltages in the range of 50 kV and higher.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21361181     DOI: 10.1118/1.3523599

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Investigation of x-ray spectra for iodinated contrast-enhanced dedicated breast CT.

Authors:  Stephen J Glick; Andrey Makeev
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-26

2.  Measurement of breast tissue composition with dual energy cone-beam computed tomography: a postmortem study.

Authors:  Huanjun Ding; Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

3.  High-resolution spiral CT of the breast at very low dose: concept and feasibility considerations.

Authors:  Willi A Kalender; Marcel Beister; John M Boone; Daniel Kolditz; Sabrina V Vollmar; Michaela C C Weigel
Journal:  Eur Radiol       Date:  2011-06-09       Impact factor: 5.315

4.  Characterization of scatter magnitude and distribution in dedicated breast computed tomography with bowtie filters.

Authors:  Kimberly Kontson; Robert J Jennings
Journal:  J Med Imaging (Bellingham)       Date:  2014-12-18

5.  Technical feasibility proof for high-resolution low-dose photon-counting CT of the breast.

Authors:  Willi A Kalender; Daniel Kolditz; Christian Steiding; Veikko Ruth; Ferdinand Lück; Ann-Christin Rößler; Evelyn Wenkel
Journal:  Eur Radiol       Date:  2016-06-15       Impact factor: 5.315

6.  Characteristic performance evaluation of a photon counting Si strip detector for low dose spectral breast CT imaging.

Authors:  Hyo-Min Cho; William C Barber; Huanjun Ding; Jan S Iwanczyk; Sabee Molloi
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

7.  Microcalcification detectability using a bench-top prototype photon-counting breast CT based on a Si strip detector.

Authors:  Hyo-Min Cho; Huanjun Ding; William C Barber; Jan S Iwanczyk; Sabee Molloi
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

8.  Optimization of the energy for Breast monochromatic absorption X-ray Computed Tomography.

Authors:  Pasquale Delogu; Vittorio Di Trapani; Luca Brombal; Giovanni Mettivier; Angelo Taibi; Piernicola Oliva
Journal:  Sci Rep       Date:  2019-09-11       Impact factor: 4.379

9.  Association between power law coefficients of the anatomical noise power spectrum and lesion detectability in breast imaging modalities.

Authors:  Lin Chen; Craig K Abbey; John M Boone
Journal:  Phys Med Biol       Date:  2013-02-19       Impact factor: 3.609

10.  Investigation of energy weighting using an energy discriminating photon counting detector for breast CT.

Authors:  Kesava S Kalluri; Mufeed Mahd; Stephen J Glick
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.