Characterization of scatter magnitude and distribution in dedicated breast computed tomography with bowtie filters.

Kimberly Kontson1, Robert J Jennings1.   

Abstract

Scatter contamination of projection images in cone-beam computed tomography (CT) degrades the image quality. The use of bowtie filters in dedicated breast CT can decrease this scatter contribution. Three bowtie filter designs that compensate for one or more aspects of the beam-modifying effects due to differences in path length in a projection were studied. These designs have been investigated in terms of their ability to reduce the scatter contamination in projection images acquired in a dedicated breast CT geometry. The scatter magnitude was measured as the scatter-to-primary ratio (SPR) using experimental and Monte Carlo techniques for various breast phantom diameters and tube voltages. The results show that a 55% reduction in the center SPR value could be obtained with the bowtie filter designs. On average, the bowtie filters reduced the center SPR by approximately 18% over all breast diameters. The distribution of the scatter was calculated at a range of different locations to produce scatter distribution maps for all three bowtie filter designs. With the inclusion of the bowtie filters, the scatter distribution was more uniform for all breast diameters. The results of this study will be useful in designing scatter correction methods and understanding the benefits of bowtie filters in dedicated breast CT.

Keywords:  bowtie filters; breast computed tomography; scatter; scatter-to-primary ratio

Year:  2014        PMID: 26158057      PMCID: PMC4478986          DOI: 10.1117/1.JMI.1.3.033505

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  29 in total

1.  A PENELOPE-based system for the automated Monte Carlo simulation of clinacs and voxelized geometries-application to far-from-axis fields.

Authors:  Josep Sempau; Andreu Badal; Lorenzo Brualla
Journal:  Med Phys       Date:  2011-11       Impact factor: 4.071

2.  Experimental spectral measurements of heavy K-edge filtered beams for x-ray computed mammotomography.

Authors:  D J Crotty; R L McKinley; M P Tornai
Journal:  Phys Med Biol       Date:  2007-01-10       Impact factor: 3.609

3.  The myth of the 50-50 breast.

Authors:  M J Yaffe; J M Boone; N Packard; O Alonzo-Proulx; S Y Huang; C L Peressotti; A Al-Mayah; K Brock
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

4.  Scatter correction method for x-ray CT using primary modulation: phantom studies.

Authors:  Hewei Gao; Rebecca Fahrig; N Robert Bennett; Mingshan Sun; Josh Star-Lack; Lei Zhu
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

5.  Spectral optimization for dedicated breast CT.

Authors:  Michaela Weigel; Sabrina V Vollmar; Willi A Kalender
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

6.  Photon counting spectral breast CT: effect of adaptive filtration on CT numbers, noise, and contrast to noise ratio.

Authors:  Justin D Silkwood; Kenneth L Matthews; Polad M Shikhaliev
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

7.  A method for comparing beam-hardening filter materials for diagnostic radiology.

Authors:  R J Jennings
Journal:  Med Phys       Date:  1988 Jul-Aug       Impact factor: 4.071

8.  Performance of glass fiber antiscatter devices at mammographic energies.

Authors:  R Fahrig; J G Mainprize; N Robert; A Rogers; M J Yaffe
Journal:  Med Phys       Date:  1994-08       Impact factor: 4.071

9.  Experimentally determined spectral optimization for dedicated breast computed tomography.

Authors:  Nicolas D Prionas; Shih-Ying Huang; John M Boone
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

10.  Compensators for dose and scatter management in cone-beam computed tomography.

Authors:  S A Graham; D J Moseley; J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.