[Rebound tonometry and applanation tonometry during narcosis investigation of pediatric glaucoma].

A Rosentreter1, A M Schild, A Lappas, G K Krieglstein, T S Dietlein.   

Abstract

BACKGROUND: A comparison of intraocular pressure (IOP) in cases of infantile glaucoma during general anesthesia was carried out by measurement with the iCare rebound tonometer (RBT) and a handheld applanation tonometer (Perkins).
METHODS: A total of 45 eyes from pediatric patients with childhood glaucoma were included in this prospective trial. Bland-Altman plots and linear regression were used for statistical analysis.
RESULTS: In almost two thirds of the eyes the difference between RBT and Perkins was 2 mmHg or less. A systemic bias of 1.96 mmHg and a 95% confidence interval of -3.35 to 7.26 was detected. Linear regression of the Bland-Altman data showed a proportional error (gradient=0.16; r(2)=0.23; p<0.01).
CONCLUSIONS: The iCare rebound tonometer is useful during general anesthesia in cases of childhood glaucoma. It provides comparable IOP values to applanation tonometry with a tendency to record higher values.

Entities:  

Mesh:

Year:  2011        PMID: 21174100     DOI: 10.1007/s00347-010-2286-5

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  18 in total

1.  Rebound tonometer compared with goldmann tonometer in normal and pathologic corneas.

Authors:  Javier Moreno-Montañés; Noelia García; Ana Fernández-Hortelano; Alfredo García-Layana
Journal:  Cornea       Date:  2007-05       Impact factor: 2.651

2.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

3.  [Clinical discrepancy between Schiötz and Perkins Tonometry in juvenile glaucoma].

Authors:  T S Dietlein; P C Jacobi; G K Krieglstein
Journal:  Klin Monbl Augenheilkd       Date:  1996-11       Impact factor: 0.700

4.  The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness.

Authors:  Ali Poostchi; Robert Mitchell; Simon Nicholas; Gordon Purdie; Anthony Wells
Journal:  Clin Exp Ophthalmol       Date:  2009-09       Impact factor: 4.207

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Effect of corneal thickness on dynamic contour, rebound, and goldmann tonometry.

Authors:  Jose M Martinez-de-la-Casa; Julian Garcia-Feijoo; Eva Vico; Ana Fernandez-Vidal; Jose M Benitez del Castillo; Mohamed Wasfi; J Garcia-Sanchez
Journal:  Ophthalmology       Date:  2006-09-25       Impact factor: 12.079

7.  Prognosis of primary ab externo surgery for primary congenital glaucoma.

Authors:  T S Dietlein; P C Jacobi; G K Krieglstein
Journal:  Br J Ophthalmol       Date:  1999-03       Impact factor: 4.638

8.  Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness.

Authors:  M E Iliev; D Goldblum; K Katsoulis; C Amstutz; B Frueh
Journal:  Br J Ophthalmol       Date:  2006-05-03       Impact factor: 4.638

9.  Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*.

Authors:  Liane H van der Jagt; Nomdo M Jansonius
Journal:  Ophthalmic Physiol Opt       Date:  2005-09       Impact factor: 3.117

10.  Reproducibility and tolerability of the ICare rebound tonometer in school children.

Authors:  Afsun Sahin; Hikmet Basmak; Leyla Niyaz; Nilgun Yildirim
Journal:  J Glaucoma       Date:  2007-03       Impact factor: 2.503

View more
  1 in total

1.  COMPARISON OF IOP READINGS USING REBOUND I CARE TONOMETER AND PERKINS APPLANATION TONOMETER IN AN AFRICAN POPULATION.

Authors:  D S Ademola-Popoola; A F Odi; T M Akande
Journal:  J West Afr Coll Surg       Date:  2014 Jan-Mar
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.