Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*.

Liane H van der Jagt1, Nomdo M Jansonius.   

Abstract

Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, were compared with each other and to Goldmann applanation tonometry in a large group of healthy subjects and patients with ocular hypertension or glaucoma (n = 103). Measurements performed with the ICARE and the Tonopen XL were in good agreement with that of the Goldmann tonometer. Intraocular pressure (IOP) values measured with the TGDc-01 were significantly lower and showed more variability. The 95% limits of agreement (portable tonometer - Goldmann) were -17 to +10 mmHg for the TGDc-01, -6 to +7 mmHg for the ICARE and -6 to +8 mmHg for the Tonopen. Corneal thickness could not explain the differences between an IOP measured with the portable tonometers and an IOP measured with the Goldmann tonometer. Patient comfort was slightly higher for ICARE when compared with the Tonopen.

Entities:  

Mesh:

Year:  2005        PMID: 16101949     DOI: 10.1111/j.1475-1313.2005.00318.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  26 in total

Review 1.  Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.

Authors:  A Cervino
Journal:  Br J Ophthalmol       Date:  2006-12       Impact factor: 4.638

2.  Evaluation of the impedance tonometers TGDc-01 and iCare according to the international ocular tonometer standards ISO 8612.

Authors:  Peter Christian Ruokonen; Thomas Schwenteck; Jörg Draeger
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-02-21       Impact factor: 3.117

3.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

Review 4.  [Measuring intraocular pressure by different methods].

Authors:  J Lamparter; E M Hoffmann
Journal:  Ophthalmologe       Date:  2009-08       Impact factor: 1.059

5.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

6.  [Rebound tonometry after vitreoretinal surgery].

Authors:  A Rosentreter; R Hoerster; T Schick; N Eter; T S Dietlein; S Fauser
Journal:  Ophthalmologe       Date:  2015-11       Impact factor: 1.059

Review 7.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

8.  [Rebound tonometry and applanation tonometry during narcosis investigation of pediatric glaucoma].

Authors:  A Rosentreter; A M Schild; A Lappas; G K Krieglstein; T S Dietlein
Journal:  Ophthalmologe       Date:  2011-04       Impact factor: 1.059

9.  Effect of prolonged supine position on the intraocular pressure in patients with obstructive sleep apnea syndrome.

Authors:  Sin Yee Fang; Wan Haslina Wan Abdul Halim; Marina Mat Baki; Norshamsiah Md Din
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2018-02-28       Impact factor: 3.117

10.  Central corneal thickness and Diaton transpalpebral tonometry.

Authors:  Mustafa Ilker Toker; Ayse Vural; Haydar Erdogan; Aysen Topalkara; Mustafa Kemal Arici
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-02-20       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.