The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness.

Ali Poostchi1, Robert Mitchell, Simon Nicholas, Gordon Purdie, Anthony Wells.   

Abstract

PURPOSE: To assess agreement between the iCare rebound tonometer and Goldmann tonometry and to assess the influence of central corneal thickness and the value of scleral rebound tonometer readings.
METHODS: Prospective single-centre cross-sectional study comparing iCare rebound tonometer (RT) intraocular pressure (IOP) readings taken from corneal and scleral locations to Goldmann applanation tonometer (GAT) readings in 100 subjects attending ophthalmology clinics.
RESULTS: There was a significant difference between RT and GAT, with RT tending to overestimate IOP. The mean difference between RT and GAT measurements was 3.36 mmHg. The mean difference between the log of RT and the log of GAT measurements was 0.2356, a ratio of 1.27 (P < 0.0001). A formula derived from a linear regression analysis suggested that a 10% increase in CCT increased the RT IOP reading by 9.9%. Scleral RT readings showed no relationship to GAT readings.
CONCLUSIONS: The rebound tonometer cannot replace the Goldmann tonometer in the office setting given the wide limits of agreement between the two devices. Corneal rebound tonometer readings are influenced by CCT whereas scleral rebound tonometer readings are of no value.

Entities:  

Mesh:

Year:  2009        PMID: 19788665     DOI: 10.1111/j.1442-9071.2009.02109.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  22 in total

1.  Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects.

Authors:  Ioannis Halkiadakis; Aimilianos Stratos; George Stergiopoulos; Eleni Patsea; Sotiris Skouriotis; Panagiotis Mitropoulos; Dimitrios Papaconstantinou; Gerasimos Georgopoulos
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08       Impact factor: 3.117

2.  [Influence of corneal thickness on intraocular pressure measurements following Descemet's stripping automated endothelial keratoplasty (DSAEK)].

Authors:  M K J Klamann; A-K B Maier; J Gonnermann; N Torun; P C Ruokonen
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

3.  Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Fatih Özcura; Nilgün Yildirim; Afsun Şahin; Ertuğrul Çolak
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

4.  [Rebound tonometry after vitreoretinal surgery].

Authors:  A Rosentreter; R Hoerster; T Schick; N Eter; T S Dietlein; S Fauser
Journal:  Ophthalmologe       Date:  2015-11       Impact factor: 1.059

5.  Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects.

Authors:  Yoshitake Kato; Shunsuke Nakakura; Naoko Matsuo; Kayo Yoshitomi; Marina Handa; Hitoshi Tabuchi; Yoshiaki Kiuchi
Journal:  Int Ophthalmol       Date:  2017-04-09       Impact factor: 2.031

6.  Limbal rebound tonometry: clinical comparisons and applications.

Authors:  Georgios Bontzos; Michail Agiorgiotakis; Zoi Kapsala; Efstathios Detorakis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-06-28       Impact factor: 3.117

7.  Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients.

Authors:  Milena Pahlitzsch; Jeanette Brünner; Johannes Gonnermann; Anna-Karina B Maier; Necip Torun; Eckart Bertelmann; Matthias Kj Klamann
Journal:  Int J Ophthalmol       Date:  2016-11-18       Impact factor: 1.779

8.  [Comparison of the iCare tonometer with the Goldmann tonometer in Malawi].

Authors:  J Hohmann; M Schulze-Schwering; T Chirambo Nyaka; V Moyo; P C Kayange; D Doycheva; N H Batumba; M S Spitzer
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

9.  Does patient comfort influence the choice of tonometer for the measurement of intraocular pressure?

Authors:  Mary O Ugalahi; Mukaila A Seidu; Bolutife A Olusanya; Aderonke M Baiyeroju
Journal:  Int Ophthalmol       Date:  2015-10-16       Impact factor: 2.031

10.  Validation of the Icare® TONOVET plus rebound tonometer in normal rabbit eyes.

Authors:  Shawna Gloe; Abby Rothering; Julie A Kiland; Gillian J McLellan
Journal:  Exp Eye Res       Date:  2019-06-12       Impact factor: 3.467

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.