Temperature change in pig rib bone during implant site preparation by low-speed drilling.

Sun-Jong Kim1, Jaeyoung Yoo, Young-Soo Kim, Sang-Wan Shin.   

Abstract

OBJECTIVES: The purpose of this study was to evaluate the temperature change during low-speed drilling using infrared thermography.
MATERIAL AND METHODS: Pig ribs were used to provide cortical bone of a similar quality to human mandible. Heat production by three implant drill systems (two conventional drilling systems and one low-speed drilling system) was evaluated by measuring the bone temperature using infrared thermography. Each system had two different bur sizes. The drill systems used were twist drill (2.0 mm/2.5 mm), which establishes the direction of the implant, and finally a 3.0 mm-pilot drill. Thermal images were recorded using the IRI1001 system (Infrared Integrated Systems Ltd.). Baseline temperature was 31±1ºC. Measurements were repeated 10 times, and a static load of 10 kg was applied while drilling. Data were analyzed using descriptive statistics. Statistical analysis was conducted with two-way ANOVA. RESULTS AND
CONCLUSIONS: Mean values (n=10 drill sequences) for maximum recorded temperature (Max TºC), change in temperature (ΔTºC) from baseline were as follows. The changes in temperature (ΔTºC) were 1.57ºC and 2.46ºC for the lowest and the highest values, respectively. Drilling at 50 rpm without irrigation did not produce overheating. There was no significant difference in heat production between the 3 implant drill systems (p>0.05). No implant drill system produced heat exceeding 47ºC, which is the critical temperature for bone necrosis during low-speed drilling. Low-speed drilling without irrigation could be used during implant site preparation.

Entities:  

Mesh:

Year:  2010        PMID: 21085811      PMCID: PMC4246386          DOI: 10.1590/s1678-77572010000500016

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


INTRODUCTION

Drilling procedures during dental implant site preparation may cause not only mechanical damage to the bone involved but a temperature increase in the bone adjacent to the implant site. Significant temperature increases can result in heat-induced bone injury[2]. Thermonecrosis has been reported in the literatures[13-17]. Eriksson and Albrektsson[14] (1983) conducted an experiment on the rabbit tibia to evaluate the effects of heat production on bone regeneration. They found that heating the implants in the rabbit tibia to a temperature of 50ºC for 1 min was enough to cause 30% of the bone to be resorbed. This was a slow-developing process that extended over a period of 4 weeks. They reported that setting the temperature between 47 and 50ºC for 1 min reduced significantly bone formation around the implants, while no significant effects were observed after heating up to 47ºC for 1 min[13]. Eriksson and Adell[12] (1986) measured temperature change during drilling with a thermocouple, indicating that drilling using the osseointegration technique does not impair bone regeneration due to excessive heat production[4,12]. Watanabe, et al.[26] (1992) measured heat distribution in bone using thermography and found that the maximum temperature was generated when bone was cut using drills. Based on the result that maximum temperature during drilling was higher without irrigation than with irrigation, they indicated that bone temperature should not exceed the critical temperature without irrigation. Similar results have been reported by Yacker, et al.[27] (1996). Benington, et al.[7] (1996) have reported that temperature rises up to 130.1ºC without irrigation after monitoring changes in bone temperature during the sequence of drilling for implant site preparation by the osseointegration technique using infrared thermography. However, most studies have been performed to investigate temperature change at a conventional drilling speed, which is over 1,000 rpm until now. This study evaluated the safety of low-speed drilling at 50 rpm without irrigation due to temperature changes.

MATERIAL AND METHODS

Pig ribs with a 2.0 mm-thick cortical bone were obtained and sliced into 10 cm-long sections. The main reason for the selection of pig ribs was their superior homogeneity in the thickness of the cortical bone. The cow or pig mandible was considered first. However, the irregular thickness of their cortical bone specimen was a definite disadvantage. Although pig ribs are weaker than the human mandible, they are excellent in homogeneity of the thickness of cortical bone. They were embedded in a resin block, and immobilized to the table. In an attempt to duplicate clinical conditions, the lower half of the specimen was soaked in a water bath at 37±1ºC, and the upper half of the specimen at a room temperature of 25ºC. Initially, the temperature of the bone was set at 31±1ºC (Figure 1).
Figure 1

Pig rib immobilized to resin block

Pig rib immobilized to resin block For applying a static load, it was pressed 10 times with the same force as in the operation. About 10 kg of static load was applied. Bicon implant drill system (Bicon, Boston, MA, USA) was selected as the low-speed drill system in this study. Brånemark drill system (Nobelbiocare, Sweden) and Osstem drill system (Osstem, Seoul, Korea) were used as conventional speed drill systems. Table 1 shows the drilling systems, drill diameter size, irrigation and drill speed used in this study. The measurements were repeated 10 times and 10 kg of static load was applied while drilling.
Table 1

Implant drilling systems tested in the study

Implant SystemDrill sizeIrrigationDrill speed (rpm)
   TestedRecommended
 
Bränemark2.0 mm, 3.0 mmNo501,200
Osstem2.0 mm, 3.0 mmNo501,200
Bicon2.0 mm, 3.0 mmNo5050
Implant drilling systems tested in the study To ensure the consistency in the drilling procedure, one operator used a conventional dental handpiece (W&H, Bürmoos, Austria). The manufacturer's instructions were followed during implant site preparation with low-speed drilling protocol. No irrigation was performed since infrared radiation cannot transmit through water. Consequently, low-speed drilling at 50 rpm with high torque was performed because drilling without irrigation may cause bone necrosis. The temperature was recorded using a digital thermometer IRI1001 system (InfraRed Integrated Systems Ltd., UK) (Figure 2).
Figure 2

Infrared thermography measured the temperature changes

Infrared thermography measured the temperature changes To prevent distraction from bur wear, brand-new burs were used. However, since the experiment was repeated 10 times, bur wear might have been a variable as experiments progressed. Brånemark drill system was used only once according to the manufacturer's instructions. The schematic diagram of the experimental implant drill systems is shown in Figure 3.
Figure 3

Schematic diagram of the experimental system

Schematic diagram of the experimental system Data were analyzed using descriptive statistics. Statistical analysis was conducted with two-way ANOVA (α=0.05).

RESULTS

Drilling at 50 rpm without irrigation did not produce overheating at the tip of the drills; temperatures changes were 1.57ºC and 2.46ºC for the lowest and the highest values, respectively. The drill systems used were a 2.0 mm twist drill, which establishes the direction of the implant, and a 3.0 mm pilot drill, which progressively increases the diameter of the site. As expected, the conventional systems did not increase the temperature of pig cortical bone up to the critical point of 47ºC, which was reported by Eriksson, et al.[4] (2007). The low-speed drilling showed low temperature change as well. The results are summarized in Table 2.
Table 2

Mean baseline temperature, maxiamum temperature, and temperature difference according to implant drill systems and drill diameter

SystemDiameter (mm)Mean Baseline Temp. (ºC)Mean Max. Temp. (ºC)Mean T (ºC)
     
Bränemark230.9932.561.57
 330.5932.581.99
Osstem231.7433.531.79
 331.233.562.46
Bicon2.530.6132.281.67
 330.4532.171.72
Mean baseline temperature, maxiamum temperature, and temperature difference according to implant drill systems and drill diameter No implant drill systems produced heat exceeding the critical temperature. There were no significant differences in heat production among the 3 systems. Temperature became higher as drill diameter increased. The implant drill system and the drill diameter were defined as independent factors. There was no significant difference among the 3 implant drill systems (p>0.05). There was statistically significant difference in heat production according to drill diameter (p<0.05). The larger the drill diameter was, the higher the bone temperature because the greater bone volume to be cut created more friction during the drilling procedure. Figure 4 shows the relationship between drill diameter and bone temperature.
Figure 4

Relationship between drill diameter and bone temperature

Relationship between drill diameter and bone temperature

DISCUSSION

The purpose of this study was to investigate the heat produced by 3 individual implant drill systems after repeated drilling including low-speed drilling technique without irrigation. The concept of low-speed drilling has been suggested as an alternative to the conventional procedure. This technique has pros, including easy manipulation of drilling path, and cons, including longer drilling time. The strength of low-speed drilling can be easily controlled to maintain the path of drilling. Conventional drilling can change its drilling path on its own when the drill encounters a dense cortical bone. Low-speed drilling can inform the operator that the path has changed so that the operator can correct it. Ericsson and Albrektsson[14] (1983) demonstrated that heating bone to 47ºC for 1 min reduced the amount of bone that grew into the implant. They established that the threshold level for bone survival during implant site preparation is 47ºC, keeping drilling time below 1 min. Bachus, et al.[5] (2000) have demonstrated that through the application of a strong force to the drill, both the maximum cortical temperature above 50ºC and its duration can be effectively reduced and subsequently decrease the potential for thermal necrosis in the neighboring cortical bone. Multiple factors have been implicated in the production of heat during osteotomy preparation. These factors include drill speed[9,24,25], cortical thickness[3,23,27], drill sharpness[1,11], drill force[1,18], drill depth[11], drill design[11,18], drill diameter[11], graduated compared with 1-step drilling[11,27], and irrigation[6,15,22]. Bur wear has been considered as well. Chacon, et al.[10] (2006) have concluded that bur wear and bur geometry play a major role in heat production. From a heat generation standpoint, Misir, et al.[23] (1972) conclude that preparing an implant site with using surgical drill guides generates heat more than classical implant site preparation regardless of the irrigation type. Bone temperature can be measured by two methods. One is the method using a thermocouple embedded in bone, and the other is infrared thermography. Thermography detects the radiant infrared value of a subject, discriminates the distribution of temperature as a visible image, and expresses it in color on a monitor. Thermography has a few advantages: its ease of use and real time measurement. Although thermography is limited in accuracy with an error of up to ±1ºC, this error is not large enough to affect the conclusion. Infrared thermography was used in this study because the method using a thermocouple required too much effort to set up the instrument, and also because it was unclear whether this method was more accurate. Implant site preparation techniques currently in use involve drilling at speeds of 1,000 to 1,500 rpm to prepare potential recipient sites. The mechanical and thermal damage to the tissue surrounding the implant during drilling could have a destructive effect on the initial state of the cavity housing the implant[19,20]. In this study, gradual drilling caused less friction, and thus less trauma to the bone. Compared to conventional drilling, low-speed drilling took much longer duration of drilling time. A single drill with the same size (2 mm and 3 mm) was used for each implant drill system except for the Brånemark system. Different Brånemark system burs with the same size were used only once according to the manufacturer's protocol, as previously reported by Chacon, et al.[9] (1996). Those authors showed that temperature increases when drills were used multiple times. Clinicians sometimes need a small amount of bone graft material during implant surgery. Some use alloplastic material and others take autogenous bone from the patient's donor site. Others have used bone was collected with new devices during surgery for bone graft. This is particularly important because clinically autogenous bone has been reported to be the ideal defect filler. Bone collected during low-speed drilling may be easier to manipulate than bone collected by other methods, such as bone trap and bone collector. The effects of drill speed on heat production have been studied to improve irrigation delivery systems[19,20]. The fact that the drill is used at low speeds helps improving the quality of bone obtained, as shown by light microscopic and ultrastructural analysis[4], and the biological conditions of bone tissue in the new-alveolar sockets. However, this method has a disadvantage in that friction may possibly traumatize the collected bone as well as the surrounding bone because it cannot be irrigated during drilling. To prevent bone necrosis, clinicians should drill at a low speed between 20 and 100 rpm without irrigation. As the drill completes considerably more rotations at high speed than at low speed, yet advances the same distance, high-speed drilling shreds the tissue to a greater extent. The effects of drill speed on heat production have been studied to improve irrigation delivery systems[25]. Several authors[8,21] have compared bone particles obtained from different commercial bone traps and examined differences in the percentage of bone to coagulum. Bone trap design also affects the mass and the nature of the collected tissue[28]. The presence of microorganisms found during the collection process can be very high due to saliva retention with some bone traps[8]. However, the displaced tissue retained in the drill bit during low-speed drilling was very easy to collect, reducing saliva contamination. In the present study, low-speed drilling without irrigation did not increase bone temperature. The small number of data included in this study limits the extrapolation of results. A randomized clinical trial comparing low-speed drilling and conventional speed drilling under controlled clinical conditions should be conducted to provide further evidence of the effect of low-speed drilling protocols. Animal studies may also be helpful in gathering histological data on bone regeneration from collected bone during low-speed drilling.

CONCLUSIONS

Within the limitations of this study, it is suggested that low-speed drilling at 50 rpm without irrigation may not significantly increase bone temperature, and that there may be a direct relationship between bur diameter and bone temperature. This may explain why we start drilling with the smallest-sized bur and increase the bur size progressively during implant surgery.
  28 in total

1.  The effects of drilling force on cortical temperatures and their duration: an in vitro study.

Authors:  K N Bachus; M T Rondina; D T Hutchinson
Journal:  Med Eng Phys       Date:  2000-12       Impact factor: 2.242

2.  Heat generation during implant drilling: the significance of motor speed.

Authors:  Mohamed Sharawy; Carl E Misch; Norman Weller; Sherif Tehemar
Journal:  J Oral Maxillofac Surg       Date:  2002-10       Impact factor: 1.895

3.  Temperatures measured in human cortical bone when drilling.

Authors:  L S Matthews; C Hirsch
Journal:  J Bone Joint Surg Am       Date:  1972-03       Impact factor: 5.284

4.  Thermal injury to bone. A vital-microscopic description of heat effects.

Authors:  A Eriksson; T Albrektsson; B Grane; D McQueen
Journal:  Int J Oral Surg       Date:  1982-04

5.  Effect of internal irrigation on frictional heat generated from bone drilling.

Authors:  C Lavelle; D Wedgwood
Journal:  J Oral Surg       Date:  1980-07

6.  The effect of heat on bone regeneration: an experimental study in the rabbit using the bone growth chamber.

Authors:  R A Eriksson; T Albrektsson
Journal:  J Oral Maxillofac Surg       Date:  1984-11       Impact factor: 1.895

7.  Bone collected during dental implant surgery: a clinical and histological study.

Authors:  Malcolm Peter Jackson Young; Helen Victoria Worthington; Richard Ernest Lloyd; David Bernard Drucker; Philip Sloan; Duncan Howard Carter
Journal:  Clin Oral Implants Res       Date:  2002-06       Impact factor: 5.977

8.  Thermal changes observed at implant sites during internal and external irrigation.

Authors:  Ian C Benington; Paul A Biagioni; James Briggs; Sean Sheridan; Philip-John Lamey
Journal:  Clin Oral Implants Res       Date:  2002-06       Impact factor: 5.977

9.  Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man.

Authors:  T Albrektsson; P I Brånemark; H A Hansson; J Lindström
Journal:  Acta Orthop Scand       Date:  1981

10.  Viability of autogenous bone grafts obtained by using bone collectors: histological and microbiological study.

Authors:  Alberto Blay; Samy Tunchel; Wilson Roberto Sendyk
Journal:  Pesqui Odontol Bras       Date:  2003-12-16
View more
  22 in total

1.  Slow drilling speeds for single-drill implant bed preparation. Experimental in vitro study.

Authors:  R A Delgado-Ruiz; E Velasco Ortega; G E Romanos; S Gerhke; I Newen; J L Calvo-Guirado
Journal:  Clin Oral Investig       Date:  2017-04-22       Impact factor: 3.573

2.  Effect of irrigation and stainless steel drills on dental implant bed heat generation.

Authors:  B Bullon; E F Bueno; M Herrero; A Fernandez-Palacin; J V Rios; P Bullon; F J Gil
Journal:  J Mater Sci Mater Med       Date:  2015-01-29       Impact factor: 3.896

3.  [Study on bone volume harvested from the implant sites with different methods].

Authors:  W T Li; P Li; M Z Piao; F Zhang; J DI
Journal:  Beijing Da Xue Xue Bao Yi Xue Ban       Date:  2020-02-18

4.  Biomechanical Evaluation of Temperature Rising and Applied Force in Controlled Cortical Bone Drilling: an Animal in Vitro Study.

Authors:  Mohammad Javad Ein-Afshar; Mostafa Shahrezaee; Mohammad Hossein Shahrezaee; Seyed Reza Sharifzadeh
Journal:  Arch Bone Jt Surg       Date:  2020-09

5.  Comparison of heat production and bone architecture changes in the implant site preparation with compressive osteotomes, osseodensification technique, piezoelectric devices, and standard drills: an ex vivo study on porcine ribs.

Authors:  Nishith Bhargava; Vittoria Perrotti; Vito Carlo Alberto Caponio; Victor Haruo Matsubara; Diana Patalwala; Alessandro Quaranta
Journal:  Odontology       Date:  2022-07-19       Impact factor: 2.885

Review 6.  Low-speed drilling without irrigation versus conventional drilling for dental implant osteotomy preparation: a systematic review.

Authors:  Juan Carlos Bernabeu-Mira; David Soto-Peñaloza; Miguel Peñarrocha-Diago; Fabio Camacho-Alonso; Rebeca Rivas-Ballester; David Peñarrocha-Oltra
Journal:  Clin Oral Investig       Date:  2021-04-24       Impact factor: 3.573

Review 7.  Heat Development During Medical Drilling: Influencing Factors and Examination Methods - Overview and First Results.

Authors:  Ole Jung; Carolin Lindner; Sven Pantermehl; Mike Barbeck
Journal:  In Vivo       Date:  2021 Nov-Dec       Impact factor: 2.155

8.  The effect of simplifying dental implant drilling sequence on osseointegration: an experimental study in dogs.

Authors:  Gabriela Giro; Nick Tovar; Charles Marin; Estevam A Bonfante; Ryo Jimbo; Marcelo Suzuki; Malvin N Janal; Paulo G Coelho
Journal:  Int J Biomater       Date:  2013-01-30

9.  Geometrical design characteristics of orthodontic mini-implants predicting maximum insertion torque.

Authors:  Višnja Katić; Ervin Kamenar; David Blažević; Stjepan Spalj
Journal:  Korean J Orthod       Date:  2014-07-24       Impact factor: 1.372

10.  The Effect of Saline Coolant on Temperature Levels during Decortication with a Midas Rex: An in Vitro Model Using Sheep Cervical Vertebrae.

Authors:  Asher Livingston; Tian Wang; Chris Christou; Matthew H Pelletier; William R Walsh
Journal:  Front Surg       Date:  2015-07-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.