OBJECTIVES: The purpose of this study was to evaluate the temperature change during low-speed drilling using infrared thermography. MATERIAL AND METHODS: Pig ribs were used to provide cortical bone of a similar quality to human mandible. Heat production by three implant drill systems (two conventional drilling systems and one low-speed drilling system) was evaluated by measuring the bone temperature using infrared thermography. Each system had two different bur sizes. The drill systems used were twist drill (2.0 mm/2.5 mm), which establishes the direction of the implant, and finally a 3.0 mm-pilot drill. Thermal images were recorded using the IRI1001 system (Infrared Integrated Systems Ltd.). Baseline temperature was 31±1ºC. Measurements were repeated 10 times, and a static load of 10 kg was applied while drilling. Data were analyzed using descriptive statistics. Statistical analysis was conducted with two-way ANOVA. RESULTS AND CONCLUSIONS: Mean values (n=10 drill sequences) for maximum recorded temperature (Max TºC), change in temperature (ΔTºC) from baseline were as follows. The changes in temperature (ΔTºC) were 1.57ºC and 2.46ºC for the lowest and the highest values, respectively. Drilling at 50 rpm without irrigation did not produce overheating. There was no significant difference in heat production between the 3 implant drill systems (p>0.05). No implant drill system produced heat exceeding 47ºC, which is the critical temperature for bone necrosis during low-speed drilling. Low-speed drilling without irrigation could be used during implant site preparation.
OBJECTIVES: The purpose of this study was to evaluate the temperature change during low-speed drilling using infrared thermography. MATERIAL AND METHODS:Pig ribs were used to provide cortical bone of a similar quality to human mandible. Heat production by three implant drill systems (two conventional drilling systems and one low-speed drilling system) was evaluated by measuring the bone temperature using infrared thermography. Each system had two different bur sizes. The drill systems used were twist drill (2.0 mm/2.5 mm), which establishes the direction of the implant, and finally a 3.0 mm-pilot drill. Thermal images were recorded using the IRI1001 system (Infrared Integrated Systems Ltd.). Baseline temperature was 31±1ºC. Measurements were repeated 10 times, and a static load of 10 kg was applied while drilling. Data were analyzed using descriptive statistics. Statistical analysis was conducted with two-way ANOVA. RESULTS AND CONCLUSIONS: Mean values (n=10 drill sequences) for maximum recorded temperature (Max TºC), change in temperature (ΔTºC) from baseline were as follows. The changes in temperature (ΔTºC) were 1.57ºC and 2.46ºC for the lowest and the highest values, respectively. Drilling at 50 rpm without irrigation did not produce overheating. There was no significant difference in heat production between the 3 implant drill systems (p>0.05). No implant drill system produced heat exceeding 47ºC, which is the critical temperature for bone necrosis during low-speed drilling. Low-speed drilling without irrigation could be used during implant site preparation.
Drilling procedures during dental implant site preparation may cause not only mechanical
damage to the bone involved but a temperature increase in the bone adjacent to the
implant site. Significant temperature increases can result in heat-induced bone
injury[2]. Thermonecrosis has been
reported in the literatures[13-17]. Eriksson and Albrektsson[14] (1983) conducted an experiment on the
rabbit tibia to evaluate the effects of heat production on bone regeneration. They found
that heating the implants in the rabbit tibia to a temperature of 50ºC for 1 min
was enough to cause 30% of the bone to be resorbed. This was a slow-developing process
that extended over a period of 4 weeks. They reported that setting the temperature
between 47 and 50ºC for 1 min reduced significantly bone formation around the
implants, while no significant effects were observed after heating up to 47ºC for
1 min[13]. Eriksson and Adell[12] (1986) measured temperature change
during drilling with a thermocouple, indicating that drilling using the osseointegration
technique does not impair bone regeneration due to excessive heat production[4,12]. Watanabe, et al.[26]
(1992) measured heat distribution in bone using thermography and found that the maximum
temperature was generated when bone was cut using drills. Based on the result that
maximum temperature during drilling was higher without irrigation than with irrigation,
they indicated that bone temperature should not exceed the critical temperature without
irrigation. Similar results have been reported by Yacker, et al.[27] (1996). Benington, et al.[7] (1996) have reported that temperature
rises up to 130.1ºC without irrigation after monitoring changes in bone
temperature during the sequence of drilling for implant site preparation by the
osseointegration technique using infrared thermography. However, most studies have been
performed to investigate temperature change at a conventional drilling speed, which is
over 1,000 rpm until now. This study evaluated the safety of low-speed drilling at 50
rpm without irrigation due to temperature changes.
MATERIAL AND METHODS
Pig ribs with a 2.0 mm-thick cortical bone were obtained and sliced into 10 cm-long
sections. The main reason for the selection of pig ribs was their superior homogeneity
in the thickness of the cortical bone. The cow or pig mandible was considered first.
However, the irregular thickness of their cortical bone specimen was a definite
disadvantage. Although pig ribs are weaker than the human mandible, they are excellent
in homogeneity of the thickness of cortical bone. They were embedded in a resin block,
and immobilized to the table. In an attempt to duplicate clinical conditions, the lower
half of the specimen was soaked in a water bath at 37±1ºC, and the upper
half of the specimen at a room temperature of 25ºC. Initially, the temperature of
the bone was set at 31±1ºC (Figure
1).
Figure 1
Pig rib immobilized to resin block
Pig rib immobilized to resin blockFor applying a static load, it was pressed 10 times with the same force as in the
operation. About 10 kg of static load was applied. Bicon implant drill system (Bicon,
Boston, MA, USA) was selected as the low-speed drill system in this study.
Brånemark drill system (Nobelbiocare, Sweden) and Osstem drill system (Osstem,
Seoul, Korea) were used as conventional speed drill systems. Table 1 shows the drilling systems, drill diameter size, irrigation
and drill speed used in this study. The measurements were repeated 10 times and 10 kg of
static load was applied while drilling.
Table 1
Implant drilling systems tested in the study
Implant System
Drill size
Irrigation
Drill speed (rpm)
Tested
Recommended
Bränemark
2.0 mm, 3.0 mm
No
50
1,200
Osstem
2.0 mm, 3.0 mm
No
50
1,200
Bicon
2.0 mm, 3.0 mm
No
50
50
Implant drilling systems tested in the studyTo ensure the consistency in the drilling procedure, one operator used a conventional
dental handpiece (W&H, Bürmoos, Austria). The manufacturer's instructions
were followed during implant site preparation with low-speed drilling protocol. No
irrigation was performed since infrared radiation cannot transmit through water.
Consequently, low-speed drilling at 50 rpm with high torque was performed because
drilling without irrigation may cause bone necrosis. The temperature was recorded using
a digital thermometer IRI1001 system (InfraRed Integrated Systems Ltd., UK) (Figure 2).
Figure 2
Infrared thermography measured the temperature changes
Infrared thermography measured the temperature changesTo prevent distraction from bur wear, brand-new burs were used. However, since the
experiment was repeated 10 times, bur wear might have been a variable as experiments
progressed. Brånemark drill system was used only once according to the
manufacturer's instructions. The schematic diagram of the experimental implant drill
systems is shown in Figure 3.
Figure 3
Schematic diagram of the experimental system
Schematic diagram of the experimental systemData were analyzed using descriptive statistics. Statistical analysis was conducted with
two-way ANOVA (α=0.05).
RESULTS
Drilling at 50 rpm without irrigation did not produce overheating at the tip of the
drills; temperatures changes were 1.57ºC and 2.46ºC for the lowest and the
highest values, respectively.The drill systems used were a 2.0 mm twist drill, which establishes the direction of the
implant, and a 3.0 mm pilot drill, which progressively increases the diameter of the
site.As expected, the conventional systems did not increase the temperature of pig cortical
bone up to the critical point of 47ºC, which was reported by Eriksson, et
al.[4] (2007). The low-speed
drilling showed low temperature change as well. The results are summarized in Table 2.
Table 2
Mean baseline temperature, maxiamum temperature, and temperature difference
according to implant drill systems and drill diameter
System
Diameter (mm)
Mean Baseline Temp. (ºC)
Mean Max. Temp. (ºC)
Mean T (ºC)
Bränemark
2
30.99
32.56
1.57
3
30.59
32.58
1.99
Osstem
2
31.74
33.53
1.79
3
31.2
33.56
2.46
Bicon
2.5
30.61
32.28
1.67
3
30.45
32.17
1.72
Mean baseline temperature, maxiamum temperature, and temperature difference
according to implant drill systems and drill diameterNo implant drill systems produced heat exceeding the critical temperature. There were no
significant differences in heat production among the 3 systems. Temperature became
higher as drill diameter increased. The implant drill system and the drill diameter were
defined as independent factors. There was no significant difference among the 3 implant
drill systems (p>0.05). There was statistically significant difference in heat
production according to drill diameter (p<0.05). The larger the drill diameter was,
the higher the bone temperature because the greater bone volume to be cut created more
friction during the drilling procedure. Figure 4
shows the relationship between drill diameter and bone temperature.
Figure 4
Relationship between drill diameter and bone temperature
Relationship between drill diameter and bone temperature
DISCUSSION
The purpose of this study was to investigate the heat produced by 3 individual implant
drill systems after repeated drilling including low-speed drilling technique without
irrigation.The concept of low-speed drilling has been suggested as an alternative to the
conventional procedure. This technique has pros, including easy manipulation of drilling
path, and cons, including longer drilling time. The strength of low-speed drilling can
be easily controlled to maintain the path of drilling. Conventional drilling can change
its drilling path on its own when the drill encounters a dense cortical bone. Low-speed
drilling can inform the operator that the path has changed so that the operator can
correct it. Ericsson and Albrektsson[14]
(1983) demonstrated that heating bone to 47ºC for 1 min reduced the amount of
bone that grew into the implant. They established that the threshold level for bone
survival during implant site preparation is 47ºC, keeping drilling time below 1
min. Bachus, et al.[5] (2000) have
demonstrated that through the application of a strong force to the drill, both the
maximum cortical temperature above 50ºC and its duration can be effectively
reduced and subsequently decrease the potential for thermal necrosis in the neighboring
cortical bone.Multiple factors have been implicated in the production of heat during osteotomy
preparation. These factors include drill speed[9,24,25], cortical thickness[3,23,27], drill sharpness[1,11], drill force[1,18], drill depth[11], drill
design[11,18], drill diameter[11], graduated compared with 1-step drilling[11,27], and irrigation[6,15,22]. Bur wear has been considered as well. Chacon, et al.[10] (2006) have concluded that bur wear and
bur geometry play a major role in heat production. From a heat generation standpoint,
Misir, et al.[23] (1972) conclude that
preparing an implant site with using surgical drill guides generates heat more than
classical implant site preparation regardless of the irrigation type.Bone temperature can be measured by two methods. One is the method using a thermocouple
embedded in bone, and the other is infrared thermography. Thermography detects the
radiant infrared value of a subject, discriminates the distribution of temperature as a
visible image, and expresses it in color on a monitor. Thermography has a few
advantages: its ease of use and real time measurement. Although thermography is limited
in accuracy with an error of up to ±1ºC, this error is not large enough to
affect the conclusion. Infrared thermography was used in this study because the method
using a thermocouple required too much effort to set up the instrument, and also because
it was unclear whether this method was more accurate.Implant site preparation techniques currently in use involve drilling at speeds of 1,000
to 1,500 rpm to prepare potential recipient sites. The mechanical and thermal damage to
the tissue surrounding the implant during drilling could have a destructive effect on
the initial state of the cavity housing the implant[19,20]. In this study, gradual
drilling caused less friction, and thus less trauma to the bone. Compared to
conventional drilling, low-speed drilling took much longer duration of drilling time. A
single drill with the same size (2 mm and 3 mm) was used for each implant drill system
except for the Brånemark system. Different Brånemark system burs with the
same size were used only once according to the manufacturer's protocol, as previously
reported by Chacon, et al.[9] (1996).
Those authors showed that temperature increases when drills were used multiple
times.Clinicians sometimes need a small amount of bone graft material during implant surgery.
Some use alloplastic material and others take autogenous bone from the patient's donor
site. Others have used bone was collected with new devices during surgery for bone
graft. This is particularly important because clinically autogenous bone has been
reported to be the ideal defect filler. Bone collected during low-speed drilling may be
easier to manipulate than bone collected by other methods, such as bone trap and bone
collector. The effects of drill speed on heat production have been studied to improve
irrigation delivery systems[19,20]. The fact that the drill is used at low
speeds helps improving the quality of bone obtained, as shown by light microscopic and
ultrastructural analysis[4], and the
biological conditions of bone tissue in the new-alveolar sockets. However, this method
has a disadvantage in that friction may possibly traumatize the collected bone as well
as the surrounding bone because it cannot be irrigated during drilling. To prevent bone
necrosis, clinicians should drill at a low speed between 20 and 100 rpm without
irrigation. As the drill completes considerably more rotations at high speed than at low
speed, yet advances the same distance, high-speed drilling shreds the tissue to a
greater extent. The effects of drill speed on heat production have been studied to
improve irrigation delivery systems[25].
Several authors[8,21] have compared bone particles obtained from different
commercial bone traps and examined differences in the percentage of bone to coagulum.
Bone trap design also affects the mass and the nature of the collected tissue[28]. The presence of microorganisms found
during the collection process can be very high due to saliva retention with some bone
traps[8]. However, the displaced
tissue retained in the drill bit during low-speed drilling was very easy to collect,
reducing saliva contamination.In the present study, low-speed drilling without irrigation did not increase bone
temperature. The small number of data included in this study limits the extrapolation of
results. A randomized clinical trial comparing low-speed drilling and conventional speed
drilling under controlled clinical conditions should be conducted to provide further
evidence of the effect of low-speed drilling protocols. Animal studies may also be
helpful in gathering histological data on bone regeneration from collected bone during
low-speed drilling.
CONCLUSIONS
Within the limitations of this study, it is suggested that low-speed drilling at 50 rpm
without irrigation may not significantly increase bone temperature, and that there may
be a direct relationship between bur diameter and bone temperature. This may explain why
we start drilling with the smallest-sized bur and increase the bur size progressively
during implant surgery.
Authors: Malcolm Peter Jackson Young; Helen Victoria Worthington; Richard Ernest Lloyd; David Bernard Drucker; Philip Sloan; Duncan Howard Carter Journal: Clin Oral Implants Res Date: 2002-06 Impact factor: 5.977
Authors: Ian C Benington; Paul A Biagioni; James Briggs; Sean Sheridan; Philip-John Lamey Journal: Clin Oral Implants Res Date: 2002-06 Impact factor: 5.977
Authors: R A Delgado-Ruiz; E Velasco Ortega; G E Romanos; S Gerhke; I Newen; J L Calvo-Guirado Journal: Clin Oral Investig Date: 2017-04-22 Impact factor: 3.573
Authors: B Bullon; E F Bueno; M Herrero; A Fernandez-Palacin; J V Rios; P Bullon; F J Gil Journal: J Mater Sci Mater Med Date: 2015-01-29 Impact factor: 3.896
Authors: Juan Carlos Bernabeu-Mira; David Soto-Peñaloza; Miguel Peñarrocha-Diago; Fabio Camacho-Alonso; Rebeca Rivas-Ballester; David Peñarrocha-Oltra Journal: Clin Oral Investig Date: 2021-04-24 Impact factor: 3.573
Authors: Gabriela Giro; Nick Tovar; Charles Marin; Estevam A Bonfante; Ryo Jimbo; Marcelo Suzuki; Malvin N Janal; Paulo G Coelho Journal: Int J Biomater Date: 2013-01-30