Literature DB >> 16413899

Heat production by 3 implant drill systems after repeated drilling and sterilization.

Guillermo E Chacon1, Daniel L Bower, Peter E Larsen, Edwin A McGlumphy, F Michael Beck.   

Abstract

PURPOSE: The goal was to measure heat generated in bone by 3 implant drill systems after repeated drilling and sterilization.
MATERIALS AND METHODS: Temperature was measured with thermocouple technology in vitro using the bovine femoral cortical bone model. Intermittent drilling was accomplished at a constant 2.4-kg load and drill speed of 2,500 rpm. External irrigation at 40 mL/min with normal saline was used. Three implant drill systems-system A (triple twist drills with a relief angle), system B (triple twist drills without a relief angle), and system C (double twist drills with a relief angle)-were evaluated and heat was measured at the final drill in the drilling sequence (4.0 mm or 4.2 mm) at a depth of 15 mm. Thermocouples were placed 0.5 mm from the osteotomy at a depth of 15 mm. Heat measurements were recorded out to 25 uses.
RESULTS: Results showed temperature increased with multiple uses. System A and C drills had temperature measurements that were below 47 degrees C, even after 25 uses. System B drills had temperatures that exceeded 47 degrees C from the initial use. Light microscopy showed little drill wear even after 25 uses.
CONCLUSIONS: Drill geometry plays a major role in heat production and may explain the increased temperature readings seen in system B. These drills lack relief angles and have the smallest clearance angles of the 3 systems. It also has fewer drills in its drilling sequence compared with systems A and C. This study shows that temperatures increase when drills are used multiple times. Systems A and C had acceptable temperature measurements out to 25 uses. System B drills showed significantly higher heat production with little visual signs of wear.

Entities:  

Mesh:

Year:  2006        PMID: 16413899     DOI: 10.1016/j.joms.2005.10.011

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  31 in total

Review 1.  A new concept for navigated laser surgery.

Authors:  Sebastian Stopp; Herbert Deppe; Tim Lueth
Journal:  Lasers Med Sci       Date:  2007-07-28       Impact factor: 3.161

2.  Bone treatment laser-navigated surgery.

Authors:  Bettina Hohlweg-Majert; Herbert Deppe; Marc C Metzger; Sebstian Stopp; Klaus-Dietrich Wolff; Tim C Lueth
Journal:  Lasers Med Sci       Date:  2009-03-11       Impact factor: 3.161

3.  Study of temperature variation in cortical bone during osteotomies with trephine drills.

Authors:  Sergio Alexandre Gehrke; Marcelo Khoury Pazetto; Sérgio de Oliveira; Stefano Corbella; Silvio Taschieri; Fábio E C Mardegan
Journal:  Clin Oral Investig       Date:  2014-01-08       Impact factor: 3.573

Review 4.  Heat generated by dental implant drills during osteotomy-a review: heat generated by dental implant drills.

Authors:  Sunil Kumar Mishra; Ramesh Chowdhary
Journal:  J Indian Prosthodont Soc       Date:  2014-02-18

5.  Identification of possible factors influencing temperatures elevation during implant site preparation with piezoelectric technique.

Authors:  Luca Lamazza; Domenica Laurito; Marco Lollobrigida; Orlando Brugnoletti; Girolamo Garreffa; Alberto De Biase
Journal:  Ann Stomatol (Roma)       Date:  2015-02-09

6.  Slow drilling speeds for single-drill implant bed preparation. Experimental in vitro study.

Authors:  R A Delgado-Ruiz; E Velasco Ortega; G E Romanos; S Gerhke; I Newen; J L Calvo-Guirado
Journal:  Clin Oral Investig       Date:  2017-04-22       Impact factor: 3.573

7.  Effect of irrigation and stainless steel drills on dental implant bed heat generation.

Authors:  B Bullon; E F Bueno; M Herrero; A Fernandez-Palacin; J V Rios; P Bullon; F J Gil
Journal:  J Mater Sci Mater Med       Date:  2015-01-29       Impact factor: 3.896

Review 8.  Surgical Drill Bit Design and Thermomechanical Damage in Bone Drilling: A Review.

Authors:  Mohd Faizal Ali Akhbar; Akmal Wani Sulong
Journal:  Ann Biomed Eng       Date:  2020-08-28       Impact factor: 3.934

9.  Evaluation of accuracy in implant site preparation performed in single- or multi-step drilling procedures.

Authors:  Nadine Marheineke; Uta Scherer; Martin Rücker; Constantin von See; Björn Rahlf; Nils-Claudius Gellrich; Marcus Stoetzer
Journal:  Clin Oral Investig       Date:  2017-12-17       Impact factor: 3.573

10.  Temperature changes during cortical bone drilling with a newly designed step drill and an internally cooled drill.

Authors:  Goran Augustin; Slavko Davila; Toma Udilljak; Tomislav Staroveski; Danko Brezak; Slaven Babic
Journal:  Int Orthop       Date:  2012-07       Impact factor: 3.075

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.