Višnja Katić1, Ervin Kamenar2, David Blažević2, Stjepan Spalj1.
Abstract
OBJECTIVE: To determine the unique contribution of geometrical design characteristics of orthodontic mini-implants on maximum insertion torque while controlling for the influence of cortical bone thickness.Entities:
Keywords: Implant insertion torque; Mini-implant orthodontic anchorage; Mini-implant stability; Orthodontic mini-implant
Year: 2014 PMID: 25133132 PMCID: PMC4130913 DOI: 10.4041/kjod.2014.44.4.177
Source DB: PubMed Journal: Korean J Orthod Impact factor: 1.372
Figure 1Orthodontic mini-implant with handheld screwdriver.
Figure 2Tested mini-implants (from left to right): Dual Top™ (Jeil Medical Corp., Seoul, Korea) 1.4 × 6 mm, and 1.4 × 8 mm; Aarhus (MEDICON eG, Tuttlingen, Germany) 1.5 × 6 mm, and 1.5 × 8 mm; Dual Top™ 1.6 × 6 mm, and 1.6 × 8 mm; Ortho Easy® (FORESTADENT®, Pforzheim, Germany) 1.7 × 6 mm, and 1.7 × 8 mm; Dual Top™ 2.0 × 6 mm, and 2.0 × 8 mm.
Figure 3Geometrical design characteristics of mini-implants (length [a], outer diameter [b], depth [c], pitch [d], and lead angle [e] of the thread).
Figure 4Torque-measuring device with mounted bone sample.
Geometrical design characteristics of orthodontic mini-implants assessed with optical stereomicroscope
Values are presented as mean ± standard deviation.
Thread shape factor means depth/pitch ratio.
Aarhus, Aarhus Anchorage System (MEDICON eG, Tuttlingen, Germany); Ortho Easy, Ortho Easy® (FORESTADENT®, Pforzheim, Germany); Dual Top, Jeil Dual Top™ Anchor System (Jeil Medical Corp., Seoul, Korea).
Descriptive statistics for maximum insertion torque with length and outer diameter of the screw thread evaluated at covariate cortical bone thickness*
Values are presented as mean ± standard error (95% confidence interval).
*1.5190 mm.
Multiple linear regression model for prediction of maximum insertion torque
R = 0.685; R2 = 0.470; adjusted R2 = 0.442; F = 16.662; p < 0.001.
By multiple linear regression analysis.
SE, Standard error; TSF, thread shape factor.
Figure 5Representative curve for manual insertion of orthodontic mini-implants.