BACKGROUND: The aim of this study was to investigate the effect of applanation tonometry on the accuracy of biometric measurements and the reproducibility of keratometric measurements. PATIENTS AND METHODS: In this clinical prospective study a total of 16 patients were enrolled who presented for cataract surgery on 30 eyes. Biometry measurements were made using an IOLMaster optical biometer before and 10 min after standard Goldmann applanation tonometry. Local anesthesia was achieved by administering one drop of fluorescein-stained oxybuprocaine 0.4%.
RESULTS: The mean absolute difference of intraocular lens power for emmetropia was 0.31 ± 0.33 D (range -1.56 to +1.12 D) and the mean standard deviation for R1 and R2 (n=28 eyes) was 0.014 ± 0.013 mm before and 0.024 ± 0.021 mm after tonometry. This difference was statistically significant (t-test, p<0.01). Moderate to severe irregular corneal surfaces were noted in some cases after tonometry.
CONCLUSIONS: Local anesthesia and applanation tonometry reduced the accuracy of IOL power calculations with differences exceeding ± 1 D and the reliability of consecutive keratometry measurements decreased significantly. Therefore, it is advisable to perform biometry measurements at the very beginning of the preoperative examination.

Entities:  

Mesh:

Year:  2011        PMID: 20922529     DOI: 10.1007/s00347-010-2240-6

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  24 in total

Review 1.  Optical coherence biometry.

Authors:  Wolfgang Haigis
Journal:  Dev Ophthalmol       Date:  2002

2.  Iris colour and the influence of local anaesthetics on pre-corneal tear film stability.

Authors:  S Patel; S Laidlaw; L Mathewson; L McCallum; C Nicholson
Journal:  Acta Ophthalmol (Copenh)       Date:  1991-06

3.  A new optical low coherence reflectometry device for ocular biometry in cataract patients.

Authors:  P J Buckhurst; J S Wolffsohn; S Shah; S A Naroo; L N Davies; E J Berrow
Journal:  Br J Ophthalmol       Date:  2009-04-19       Impact factor: 4.638

4.  Accuracy of a new partial coherence interferometry analyser for biometric measurements.

Authors:  M P Holzer; M Mamusa; G U Auffarth
Journal:  Br J Ophthalmol       Date:  2009-03-15       Impact factor: 4.638

5.  Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations.

Authors:  J T Holladay
Journal:  J Cataract Refract Surg       Date:  1997-11       Impact factor: 3.351

6.  Comparison of the SRK/T formula and other theoretical and regression formulas.

Authors:  D R Sanders; J A Retzlaff; M C Kraff; H V Gimbel; M G Raanan
Journal:  J Cataract Refract Surg       Date:  1990-05       Impact factor: 3.351

7.  Effect of fluorescein instillation on the pre-corneal tear film stability.

Authors:  L S Mengher; A J Bron; S R Tonge; D J Gilbert
Journal:  Curr Eye Res       Date:  1985-01       Impact factor: 2.424

8.  Effect of keratometer and axial length measurement errors on primary implant power calculations.

Authors:  J R McEwan; R K Massengill; S D Friedel
Journal:  J Cataract Refract Surg       Date:  1990-01       Impact factor: 3.351

9.  Partial coherence interferometry: a novel approach to biometry in cataract surgery.

Authors:  W Drexler; O Findl; R Menapace; G Rainer; C Vass; C K Hitzenberger; A F Fercher
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

10.  Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis.

Authors:  W Haigis; B Lege; N Miller; B Schneider
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2000-09       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.