| Literature DB >> 20920373 |
Kambiz Shahgaldi1, Aristomenis Manouras, Anna Abrahamsson, Petri Gudmundsson, Lars-Ake Brodin, Reidar Winter.
Abstract
BACKGROUND: Three dimensional echocardiography (3DE) approaches the accuracy of cardiac magnetic resonance in measuring left ventricular (LV) volumes and ejection fraction (EF). The multibeat modality in comparison to single-beat (SB) requires breath-hold technique and regular heart rhythm which could limit the use of this technique in patients with atrial fibrillation (AF) due to stitching artifact. The study aimed to investigate whether SB full volume 3DE acquisition reduces inter- and intraobserver variability in assessment of LV volumes and EF in comparison to four-beat (4B) ECG-gated full volume 3DE recording in patients with AF.Entities:
Mesh:
Year: 2010 PMID: 20920373 PMCID: PMC2972240 DOI: 10.1186/1476-7120-8-45
Source DB: PubMed Journal: Cardiovasc Ultrasound ISSN: 1476-7120 Impact factor: 2.062
Clinical characteristics of the study population.
| Group A | Group B | |
|---|---|---|
| Age (years) | 53 ± 17 | 65 ± 12 |
| Males (%) | 67 | 52 |
| Heart rate (bpm) | 67 ± 10 | 97 ± 27 |
| Suspected heart failure | 60 | 29 |
| Murmur/Valvular heart disease | 15 | 23 |
| Suspicion of Left ventricular thrombus | 5 | 8 |
| Routine control of transplanted heart patients | 10 | 0 |
| Unspecific exclusion of cardiac pathology | 0 | 20 |
| Other | 10 | 20 |
Figure 1(A) Four-beat ECG-gated full volume three-dimensional loop presented in 9-slice of the left ventricular (from apex to base). Note the stitching artifacts due to irregular heart rate in all 9-slices in an atrial fibrillation patient (marked with arrows). (B) Single-beat three dimensional echocardiography presented in 9-slice. No stitching artifacts can be detected in a patient having atrial fibrillation.
Figure 2Quad view presentation of left ventricular using 4D auto LVQ software for measurement of left ventricular volumes and ejection fraction with three-dimensional echocardiography. Volume time-plot and quantitative analysis and three-dimensional model are presented in the right panel.
Mean ± SD values for left ventricular end-diastolic volume (ml), end-systolic volume (ml) and ejection fraction (%) for the different imaging strategies
| EDV (ml) | ESV (ml) | EF (%) | HR (bpm) | |
|---|---|---|---|---|
| Four-beat full volume | 102 ± 25 | 47 ± 19 | 54 ± 10 | 67 ± 10 |
| Single-beat full volume | 99 ± 25 | 47 ± 20 | 53 ± 10 | 67 ± 11 |
| Four-beat full volume | 87 ± 27*** | 52 ± 22*** | 41 ± 11* | 97 ± 27 |
| Single-beat full volume | 119 ± 34 | 64 ± 27 | 47 ± 12 | 97 ± 26 |
EDV: end-diastolic volume, ESV: end-systolic volume, EF: ejection fraction, HR: heart rate,
*** p < 0,0001 vs. end-diastolic, end-systolic volume and *p < 0.05 vs. ejection fraction by single beat.
Mean differences (ml) and limits of agreement (mean ± SD) in left ventricular volumes and ejection fraction measured by single beat and four-beat ECG-gated three-dimensional echocardiography in patients with atrial fibrillation
| Group B | Group B | Group B | |
|---|---|---|---|
| EDV 4B - EDV SB | ESV 4B - ESV SB | EF 4B - EF SB | |
| Mean difference ± SD | -32.5 ± 15 | -12.2 ± 10.5 | -7 ± 10 |
| Limits of agreement | -62.5 to -2.5 | -33.2 to 8.8 | -27 to 13 |
EDV: end-diastolic volume, ESV: end-systolic volume, EF: ejection fraction,
4B: four-beat, SB: single-beat.
Figure 3A-C. Bland-Altman plot of differences between end-diastolic volume (A), end-systolic volume (B) and ejection fraction (C) determined by four-beat and and single-beat three-dimensional echocardiography in atrial fibrillation patients.
Intra- and interobserver reproducibilities for left ventricular volumes and ejection fraction by single- and four-beat three-dimensional echocardiography.
| EDV (ml) | ESV (ml) | EF (%) | |
|---|---|---|---|
| Intraobserver variability (%) | |||
| Four-beat full volume | 4.4 | 5.1 | 3.4 |
| Single-beat full volume | 3.9 | 4.6 | 4.1 |
| Interobserver variability (%) | |||
| Four-beat full volume | 7.3 | 8.5 | 3.9 |
| Single-beat full volume | 6.4 | 6.5 | 4.5 |
| Intraobserver variability (%) | |||
| Four-beat full volume | 9* | 11.4 | 8.3** |
| Single-beat full volume | 4.5 | 2.9 | 4.8 |
| Interobserver variability (%) | |||
| Four-beat full volume | 10.4* | 15.2* | 17.9** |
| Single-beat full volume | 7.6 | 7.2 | 5.6 |
EDV: end-diastolic volume, ESV: end-systolic volume and EF: ejection fraction, *p < 0.05 and **p < 0.001.