Quality of glucose measurement with blood glucose meters at the point-of-care: relevance of interfering factors.

Lutz Heinemann1.   

Abstract

AIM: A good understanding of the relevance of interfering factors having an impact on blood glucose (BG) measurement is needed to obtain the required quality. This depends on the application in which meters designed for self-monitoring of BG (SMBG) are used.
METHODS: By means of a literature search all publications (from January 1, 1980 to August 10, 2009) were identified that report about the influence of potentially interfering substances/factors on the measurement quality of BG meters.
RESULTS: Certain substances (e.g., maltose) can have a profound and misleading impact on the BG measurement result when the enzymatic reaction embedded on the given test strips cross-reacts. Also, a number of other drugs (e.g., acetaminophen) and factors (like temperature and altitude) affect the reliability of BG measurement massively. However, the susceptibility of the BG meter (depending on the enzyme technology of the test strips) differs significantly.
CONCLUSIONS: In daily practice the factors that have a relevant impact on the reliability of BG measurements with modern BG meters are rarely met. Clearly this also depends on the intended use (SMBG in patient hands vs. point-of-care testing in hospitals). To avoid misleading measurement results requires adequate training of all people involved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20879962     DOI: 10.1089/dia.2010.0076

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  18 in total

Review 1.  Blood glucose measurements in critically ill patients.

Authors:  Tom Van Herpe; Dieter Mesotten
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  Accuracy of point-of-care blood glucose measurements in critically ill patients in shock.

Authors:  Carlo Jan Pati-An Garingarao; Myrna Buenaluz-Sedurante; Cecilia Alegado Jimeno
Journal:  J Diabetes Sci Technol       Date:  2014-06-08

3.  Analysis of a Unique Postmarket Surveillance Dataset That a Glucose Test-Strip Demonstrates no Evidence of Interference and Robust Clinical Accuracy Irrespective of the Prescription Medication Status of a Large Cohort of Patients With Diabetes.

Authors:  Mike Grady; Hilary Cameron; Stuart Phillips; Gillian Smith; Steven Setford
Journal:  J Diabetes Sci Technol       Date:  2019-09-03

4.  A Comparison Study Between Point-of-Care Testing Systems and Central Laboratory for Determining Blood Glucose in Venous Blood.

Authors:  Huiping Wei; Fang Lan; Qitian He; Haiwei Li; Fuyong Zhang; Xue Qin; Shan Li
Journal:  J Clin Lab Anal       Date:  2016-08-25       Impact factor: 2.352

5.  Hematocrit compensation in electrochemical blood glucose monitoring systems.

Authors:  Maria Teodorczyk; Marco Cardosi; Steven Setford
Journal:  J Diabetes Sci Technol       Date:  2012-05-01

6.  Consensus report: the current role of self-monitoring of blood glucose in non-insulin-treated type 2 diabetes.

Authors:  David C Klonoff; Lawrence Blonde; George Cembrowski; Antonio Roberto Chacra; Guillaume Charpentier; Stephen Colagiuri; George Dailey; Robert A Gabbay; Lutz Heinemann; David Kerr; Antonio Nicolucci; William Polonsky; Oliver Schnell; Robert Vigersky; Jean-François Yale
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

7.  Comparative performance assessment of point-of-care testing devices for measuring glucose and ketones at the patient bedside.

Authors:  Ferruccio Ceriotti; Ewa Kaczmarek; Elena Guerra; Fabrizio Mastrantonio; Fausto Lucarelli; Francesco Valgimigli; Andrea Mosca
Journal:  J Diabetes Sci Technol       Date:  2014-12-17

8.  Performance of a new meter designed for assisted monitoring of blood glucose and point-of-care testing.

Authors:  Sandra Macrury; Aparna Srinivasan; John J Mahoney
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

Review 9.  Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge.

Authors:  Michael Erbach; Guido Freckmann; Rolf Hinzmann; Bernhard Kulzer; Ralph Ziegler; Lutz Heinemann; Oliver Schnell
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

10.  Evaluation of CareSens POCT Devices for Glucose Testing in the Routine Hospital Setting.

Authors:  Kagan Huysal; Yasemin U Budak; Hakan Demirci; Mehmet Ali Önelge
Journal:  J Clin Diagn Res       Date:  2015-10-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.