Performance of a new meter designed for assisted monitoring of blood glucose and point-of-care testing.

Sandra Macrury1, Aparna Srinivasan, John J Mahoney.   

Abstract

BACKGROUND: Blood glucose (BG) meters used for assisted monitoring of blood glucose (AMBG) require different attributes compared with meters designed for home use. These include safety considerations (i.e., minimized risk of blood-borne pathogen transmission), capability for testing multiple blood sample types, and enhanced performance specifications. The OneTouch® Verio™Pro+ BG meter is designed to incorporate all of these attributes.
METHODS: Meter accuracy was assessed in clinical studies with arterial, venous, and capillary blood samples with a hematocrit range of 22.9-59.8%. The effect of interferents, including anticoagulants, on accuracy was evaluated. The meter disinfection protocol was validated, and instructions for use and user acceptance of the system were assessed.
RESULTS: A total of 97% (549/566) of BG measures from all blood sample types and 95.5% (191/200) of arterial blood samples were within ±12 mg/dl or 12.5% of reference measurements. The system was unaffected by 4 anticoagulants and 57 of 59 endogenous and exogenous compounds; it was affected by 2 compounds: pralidoxime iodide and xylose. Bleach wipes were sufficient to disinfect the meter. Users felt that the meter's quality control (QC) prompts would help them to comply with regulatory requirements.
CONCLUSION: The meter provided accurate measurements of different blood samples over a wide hematocrit range and was not affected by 57 physiologic and therapeutic compounds. The QC prompts and specific infection-mitigating design further aid to make this meter system practical for AMBG in care facilities.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23566997      PMCID: PMC3737640          DOI: 10.1177/193229681300700215

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  21 in total

1.  Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer.

Authors:  Z Tang; X Du; R F Louie; G J Kost
Journal:  Am J Clin Pathol       Date:  2000-01       Impact factor: 2.493

2.  Inaccurate glucose determination by fingerstick in a patient with peripheral arterial disease.

Authors:  Tawee Tanvetyanon; Michael D Walkenstein; Antonio Marra
Journal:  Ann Intern Med       Date:  2002-11-05       Impact factor: 25.391

3.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design.

Authors:  R F Louie; Z Tang; D V Sutton; J H Lee; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-02       Impact factor: 5.534

Review 4.  Out of the cold: management of hypothermia and frostbite.

Authors:  Jay Biem; Niels Koehncke; Dale Classen; James Dosman
Journal:  CMAJ       Date:  2003-02-04       Impact factor: 8.262

5.  Accuracy of fingerstick glucose values in shock patients.

Authors:  H F Sylvain; M E Pokorny; S M English; N H Benson; T W Whitley; C J Ferenczy; J G Harrison
Journal:  Am J Crit Care       Date:  1995-01       Impact factor: 2.228

6.  Bioavailability of tolazamide from tablets: comparison of in vitro and in vivo results.

Authors:  P G Welling; R B Patel; U R Patel; W R Gillespie; W A Craig; K S Albert
Journal:  J Pharm Sci       Date:  1982-11       Impact factor: 3.534

7.  Kinetic analysis of D-xylose absorption in normal subjects and in patients with chronic renal failure.

Authors:  R M Craig; P Murphy; T P Gibson; A Quintanilla; G C Chao; C Cochrane; A Patterson; A J Atkinson
Journal:  J Lab Clin Med       Date:  1983-03

Review 8.  The pharmacokinetics and pharmacodynamics of levodopa in the treatment of Parkinson's disease.

Authors:  Soo-Peang Khor; Ann Hsu
Journal:  Curr Clin Pharmacol       Date:  2007-09

9.  A cluster of hepatitis B infections associated with incorrect use of a capillary blood sampling device in a nursing home in the Netherlands, 2007.

Authors:  H M Götz; M Schutten; G J Borsboom; B Hendriks; G van Doornum; O de Zwart
Journal:  Euro Surveill       Date:  2008-07-03

10.  Comparison of POCT and central laboratory blood glucose results using arterial, capillary, and venous samples from MICU patients on a tight glycemic protocol.

Authors:  John R Petersen; Donna F Graves; Danyel H Tacker; Anthony O Okorodudu; Amin A Mohammad; Victor J Cardenas
Journal:  Clin Chim Acta       Date:  2008-06-18       Impact factor: 3.786

View more
  2 in total

1.  Several steps forward: a new meter for multiple patient use.

Authors:  Huong T Le; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

2.  Interference Assessment of Various Endogenous and Exogenous Substances on the Performance of the Eversense Long-Term Implantable Continuous Glucose Monitoring System.

Authors:  Carrie Lorenz; Wendolyn Sandoval; Mark Mortellaro
Journal:  Diabetes Technol Ther       Date:  2018-03-30       Impact factor: 6.118

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.