Evaluation of CareSens POCT Devices for Glucose Testing in the Routine Hospital Setting.

Kagan Huysal1, Yasemin U Budak1, Hakan Demirci2, Mehmet Ali Önelge1.   

Abstract

INTRODUCTION: Glucose meters are used routinely in hospital wards to manage blood glucose levels in patients requiring frequent monitoring of blood glucose.
OBJECTIVE: Our institution has 50 POC instruments utilized by diverse population of all ages and medical conditions. The primary objective of our study was to investigate whether all these CareSens glucose meters (I-sense Inc, Seoul, South Korea) results in hospitalized patients during routine clinical care jointly satisfy the specified quality specifications, as defined by Clinical and Laboratory Standards Institute (CLSI) guideline POCT12-A3.
MATERIALS AND METHODS: The records of hospitalized patients who underwent simultaneous measures of glucose levels with both glucose meters and a central laboratory analyser between January and June 2013 were retrospectively analysed. We also performed a prospective evaluation of the accuracy of the CareSens glucose Strip.
RESULTS: Glucose concentrations measured in 840 patients ranged from 1.66 to 31.72 mmol/L The Bland-Altman difference plot between the auto analyser and all the 50 CareSens glucosemeters revealed a mean bias of -2.2%, with analytical biases for the two methods varying from -31.1% to 26.8%. Eighty four percent of the glucose meter's glucose values were within ± 12.5% for values 5.54 mmol/L of the comparative laboratory glucose values and 93% of the results were within 20% of the reference for glucose >4.2 mmol/L and 65% of the results were within 0.8 mmol/L for glucose <4.2 mmol/L.
CONCLUSION: CareSens glucose meter readings in hospital settings, especially in hypoglycaemic patients, should be confirmed by central laboratory analysers whenever possible.

Entities:  

Keywords:  Autoanalysis; Blood glucose; Methods; Qualitycontrol

Year:  2015        PMID: 26557509      PMCID: PMC4625228          DOI: 10.7860/JCDR/2015/13357.6664

Source DB:  PubMed          Journal:  J Clin Diagn Res        ISSN: 0973-709X


  16 in total

Review 1.  Quality of glucose measurement with blood glucose meters at the point-of-care: relevance of interfering factors.

Authors:  Lutz Heinemann
Journal:  Diabetes Technol Ther       Date:  2010-09-30       Impact factor: 6.118

Review 2.  A glucose meter evaluation co-designed with both health professional and consumer input.

Authors:  Harmony Thompson; Huan Chan; Florence J Logan; Helen F Heenan; Lynne Taylor; Chris Murray; Christopher M Florkowski; Christopher M A Frampton; Helen Lunt
Journal:  N Z Med J       Date:  2013-11-22

3.  Point of care estimation of blood glucose in neonates.

Authors:  Sushma Nangia
Journal:  Indian Pediatr       Date:  2012-08       Impact factor: 1.411

4.  A comparison of blood glucose meters in Australia.

Authors:  Matthew Cohen; Erin Boyle; Carol Delaney; Jonathan Shaw
Journal:  Diabetes Res Clin Pract       Date:  2005-07-11       Impact factor: 5.602

5.  Approved IFCC recommendation on reporting results for blood glucose (abbreviated).

Authors:  Paul D'Orazio; Robert W Burnett; Niels Fogh-Andersen; Ellis Jacobs; Katsuhiko Kuwa; Wolf R Külpmann; Lasse Larsson; Andrzej Lewenstam; Anton H J Maas; Gerhard Mager; Jerzy W Naskalski; Anthony O Okorodudu
Journal:  Clin Chem       Date:  2005-09       Impact factor: 8.327

6.  Comparison of blood glucose meters in a New Zealand diabetes centre.

Authors:  Christopher Florkowski; Christopher Budgen; Deborah Kendall; Helen Lunt; M Peter Moore
Journal:  Ann Clin Biochem       Date:  2009-05-19       Impact factor: 2.057

Review 7.  Blood glucose measurement in the intensive care unit: what is the best method?

Authors:  Huong T Le; Neil S Harris; Abby J Estilong; Arvid Olson; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

8.  Reliability of point-of-care capillary blood glucose measurements in the critical value range.

Authors:  Ron B Schifman; Tan T Nguyen; Susan T Page
Journal:  Arch Pathol Lab Med       Date:  2014-07       Impact factor: 5.534

Review 9.  A review of standards and statistics used to describe blood glucose monitor performance.

Authors:  Jan S Krouwer; George S Cembrowski
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

Review 10.  Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review.

Authors:  Shigeaki Inoue; Moritoki Egi; Joji Kotani; Kiyoshi Morita
Journal:  Crit Care       Date:  2013-03-18       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.