Significance of initial aortic aneurysm pressure sensor readings varies with aortic endograft design.

Gautam Shrikhande1, Sikandar Khan, Hafiz Hussain, Alicia Menezes, James McKinsey, Harry Bush.   

Abstract

BACKGROUND: The differences in implantable pressure sensor aneurysm sac readings were compared following endovascular aneurysm repair (EVAR) among three different stent grafts.
METHODS: From January 2006 to March 2009, 51 aortic stent grafts were implanted along with the Endosure sensor. Grafts used were Zenith, Talent, and Excluder. In the present retrospective study, pulse ratios were measured intraoperatively before and after aneurysm sac exclusion and in follow-up (within 30 days). Analysis of variance was used to determine significance.
RESULTS: The average aneurysm size that was repaired was 5.75 cm (range: 4-8.5 cm); 41 patients (80.4%) being male with an average age of 76.3 years (range: 58-90 years). Thirty-four grafts were Zenith, 9 were Talent, and 8 were Excluder. The average pre-aneurysm exclusion pulse ratios for the Zenith, Talent, and Excluder were 1.00, 1.08, and 0.95 (p = 0.18), respectively. The average post-aneurysm exclusion pulse ratios were 0.34, 0.67, and 0.35, respectively (p = 0.003). Pulse ratios at the time of follow-up (within 30 days) were 0.17, 0.22, and 0.11, respectively (p = 0.44). Nine of 51 (17.6%) patients had a reduction of pulse ratio of less than 30% after endograft implantation. Five of those 9 patients (55.5%) had the Talent endograft, although there was no angiographic evidence of endoleak.
CONCLUSIONS: The Talent graft has significantly higher pulse ratios following endograft implantation despite having no angiographic evidence of endoleak when compared to the Zenith and Excluder grafts. However, the pulse ratios decreased to levels similar to Zenith and Excluder within 30 days of endograft implantation. This suggests that the Talent endograft may have increased porosity initially. If pulse ratios do not appropriately decrease immediately after device implantation, further angiographic imaging may not be necessary if no obvious endoleak is seen.

Entities:  

Mesh:

Year:  2010        PMID: 20838822     DOI: 10.1007/s00268-010-0789-5

Source DB:  PubMed          Journal:  World J Surg        ISSN: 0364-2313            Impact factor:   3.352


  16 in total

1.  Chronic intraaneurysmal pressure measurement: an experimental method for evaluating the effectiveness of endovascular aortic aneurysm exclusion.

Authors:  L A Sanchez; P L Faries; M L Marin; T Ohki; R E Parsons; B Marty; D Soeiro; S Olivieri; F J Veith
Journal:  J Vasc Surg       Date:  1997-08       Impact factor: 4.268

2.  The development of endotension is associated with increased transmission of pressure and serous components in porous expanded polytetrafluoroethylene stent-grafts: characterization using a canine model.

Authors:  Susan M Trocciola; Rajeev Dayal; Rabih A Chaer; Stephanie C Lin; Brian DeRubertis; Evan J Ryer; Robert L Hynececk; Matthew J Pierce; Martin Prince; Juan Badimon; Michael L Marin; Valentin Fuster; K Craig Kent; Peter L Faries
Journal:  J Vasc Surg       Date:  2006-01       Impact factor: 4.268

3.  Quality control during endovascular aneurysm repair: monitoring aneurysmal sac pressure and superficial femoral artery flow velocity.

Authors:  G D Treharne; I M Loftus; M M Thompson; N Lennard; J Smith; G Fishwick; P R Bell
Journal:  J Endovasc Surg       Date:  1999-08

4.  Endotension distribution and the role of thrombus following endovascular AAA exclusion.

Authors:  John P Pacanowski; Scott L Stevens; Michael B Freeman; Robert S Dieter; Lance A Klosterman; Stacy S Kirkpatrick; John W Ragsdale; S Elizabeth Davis; Mitchell H Goldman
Journal:  J Endovasc Ther       Date:  2002-10       Impact factor: 3.487

5.  Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms.

Authors:  R A Baum; J P Carpenter; C Cope; M A Golden; O C Velazquez; D G Neschis; M E Mitchell; C F Barker; R M Fairman
Journal:  J Vasc Surg       Date:  2001-01       Impact factor: 4.268

6.  Endotension: an explanation for continued AAA growth after successful endoluminal repair.

Authors:  G H White; J May; P Petrasek; R Waugh; M Stephen; J Harris
Journal:  J Endovasc Surg       Date:  1999-11

7.  Intra-aneurysm pressure measurements in successfully excluded abdominal aortic aneurysm after endovascular repair.

Authors:  Björn Sonesson; Nuno Dias; Martin Malina; PerAke Olofsson; Dennis Griffin; Bengt Lindblad; Krassi Ivancev
Journal:  J Vasc Surg       Date:  2003-04       Impact factor: 4.268

8.  Initial results of wireless pressure sensing for endovascular aneurysm repair: the APEX Trial--Acute Pressure Measurement to Confirm Aneurysm Sac EXclusion.

Authors:  Takao Ohki; Kenneth Ouriel; Pierre Galvagni Silveira; Barry Katzen; Rodney White; Frank Criado; Edward Diethrich
Journal:  J Vasc Surg       Date:  2007-02       Impact factor: 4.268

9.  First experience in human beings with a permanently implantable intrasac pressure transducer for monitoring endovascular repair of abdominal aortic aneurysms.

Authors:  Sharif H Ellozy; Alfio Carroccio; Robert A Lookstein; Michael E Minor; Claudie M Sheahan; Jacob Juta; Andrew Cha; Rolando Valenzuela; Michael D Addis; Tikva S Jacobs; Victoria J Teodorescu; Michael L Marin
Journal:  J Vasc Surg       Date:  2004-09       Impact factor: 4.268

10.  Reduction of aneurysm pressure and wall stress after endovascular repair of abdominal aortic aneurysm in a canine model.

Authors:  W A Marston; E Criado; C A Baird; B A Keagy
Journal:  Ann Vasc Surg       Date:  1996-03       Impact factor: 1.466

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.