Literature DB >> 16414397

The development of endotension is associated with increased transmission of pressure and serous components in porous expanded polytetrafluoroethylene stent-grafts: characterization using a canine model.

Susan M Trocciola1, Rajeev Dayal, Rabih A Chaer, Stephanie C Lin, Brian DeRubertis, Evan J Ryer, Robert L Hynececk, Matthew J Pierce, Martin Prince, Juan Badimon, Michael L Marin, Valentin Fuster, K Craig Kent, Peter L Faries.   

Abstract

OBJECTIVE: This study used a canine model of abdominal aortic aneurysms (AAAs) to compare intra-aneurysmal pressure and thrombus formation after exclusion with Dacron and expanded polytetrafluoroethylene (ePTFE) stent-grafts.
METHODS: Prosthetic AAAs with implanted strain-gauge pressure transducers were treated by stent-graft exclusion using Food and Drug Administration-approved devices in 10 mongrel dogs: five Dacron (AneuRx) and five ePTFE (original Excluder). Intra-aneurysmal pressure was measured over 4 weeks after AAA exclusion and indexed to the systemic pressure, represented as a percentage of a simultaneously obtained systemic pressure (value = 1.0). Magnetic resonance imaging (MRI) of the intra-aneurysmal thrombus was performed at 1, 2, and 4 weeks after exclusion and expressed as a signal-to-noise ratio (S:N) to control for background signal intensity. Comparisons of pressures and S:N between the two stent-grafts was analyzed with the Student's t test. Intra-aneurysmal thrombus was characterized histologically.
RESULTS: In animals excluded with both Dacron and ePTFE stent-grafts, the intra-aneurysmal pressure was nonpulsatile and reduced to <30% of systemic pressure. Significantly greater pressure transmission was observed after AAA exclusion using ePTFE compared with Dacron stent grafts (systolic pressure: ePTFE, 0.28 +/- 0.12 vs Dacron, 0.11 +/- 0.02, P < .001; mean pressure: ePTFE, 0.16 +/- 0.08 vs Dacron, 0.06 +/- 0.02, P < .001). MRI confirmed the absence of perfusion in all aneurysms. The T1-weighted signal intensity remained persistently elevated (S:N at 1 week, 2.7 +/- 0.4 vs 2 weeks, 4.0 +/- 0.2 vs 4 weeks, 5.4 +/- 1.3) in ePTFE-treated intra-aneurysmal thrombus, suggesting an absence of thrombus organization. In contrast, progressive evolution of T1 signal intensity in aneurysms excluded by Dacron stent-grafts was consistent with maturation from intact red blood cells (S:N at 1 week, 3.3 +/- 0.4) to methemoglobin (S:N at 2 weeks, 6.1 +/- 0.8), and then hemosiderin and ferritin (S:N at 4 weeks, 2.4 +/- 0.5). Histologically, ePTFE-excluded aneurysms contained poorly organized thrombus with red blood cell fragments and haphazardly arranged fibrin deposition indicative of active remodeling and continued influx of transudated serum. In aneurysms excluded by Dacron stent-grafts, dense, mature collagenous connective tissue and organized fibrin were present, indicative of greater thrombus organization.
CONCLUSIONS: Stent-graft treatment reduces intra-aneurysmal pressure to <30% of systemic pressure when no endoleak is present; however, significantly greater pressure is present in aneurysms treated with porous ePTFE stent-grafts than Dacron grafts. Histologic and MRI imaging analysis suggest that active transudation of serous blood components may be contributing to this increased intra-aneurysmal pressure.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16414397     DOI: 10.1016/j.jvs.2005.09.023

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  4 in total

1.  Significance of initial aortic aneurysm pressure sensor readings varies with aortic endograft design.

Authors:  Gautam Shrikhande; Sikandar Khan; Hafiz Hussain; Alicia Menezes; James McKinsey; Harry Bush
Journal:  World J Surg       Date:  2010-12       Impact factor: 3.352

Review 2.  Endotension: twenty years of a controversial term.

Authors:  Álvaro Torres-Blanco; Manuel Miralles-Hernández
Journal:  CVIR Endovasc       Date:  2021-06-05

3.  An in Vitro Twist Fatigue Test of Fabric Stent-Grafts Supported by Z-Stents vs. Ringed Stents.

Authors:  Jing Lin; Robert Guidoin; Jia Du; Lu Wang; Graeham Douglas; Danjie Zhu; Mark Nutley; Lygia Perron; Ze Zhang; Yvan Douville
Journal:  Materials (Basel)       Date:  2016-02-16       Impact factor: 3.623

Review 4.  Techniques and future perspectives for the prevention and treatment of endoleaks after endovascular repair of abdominal aortic aneurysms.

Authors:  Gianluigi Orgera; Marcello Andrea Tipaldi; Florindo Laurino; Pierleone Lucatelli; Alberto Rebonato; Ioannis Paraskevopoulos; Michele Rossi; Miltiadis Krokidis
Journal:  Insights Imaging       Date:  2019-09-23
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.