Evaluation of the peripheral dose in stereotactic radiotherapy and radiosurgery treatments.

Erika Di Betta1, Laura Fariselli, Achille Bergantin, Federica Locatelli, Antonella Del Vecchio, Sara Broggi, Maria Luisa Fumagalli.   

Abstract

PURPOSE: The main purpose of this work was to compare peripheral doses absorbed during stereotactic treatment of a brain lesion delivered using different devices. These data were used to estimate the risk of stochastic effects.
METHODS: Treatment plans were created for an anthropomorphic phantom and delivered using a LINAC with stereotactic cones and a multileaf collimator, a CyberKnife system (before and after a supplemental shielding was applied), a TomoTherapy system, and a Gamma Knife unit. For each treatment, 5 Gy were prescribed to the target. Measurements were performed with thermoluminescent dosimeters inserted roughly in the position of the thyroid, sternum, upper lung, lower lung, and gonads.
RESULTS: Mean doses ranged from of 4.1 (Gamma Knife) to 62.8 mGy (LINAC with cones) in the thyroid, from 2.3 (TomoTherapy) to 30 mGy (preshielding CyberKnife) in the sternum, from 1.7 (TomoTherapy) to 20 mGy (preshielding CyberKnife) in the upper part of the lungs, from 0.98 (Gamma Knife) to 15 mGy (preshielding CyberKnife) in the lower part of the lungs, and between 0.3 (Gamma Knife) and 10 mGy (preshielding CyberKnife) in the gonads.
CONCLUSIONS: The peripheral dose absorbed in the sites of interest with a 5 Gy fraction is low. Although the risk of adverse side effects calculated for 20 Gy delivered in 5 Gy fractions is negligible, in the interest of optimum patient radioprotection, further studies are needed to determine the weight of each contributor to the peripheral dose.

Entities:  

Mesh:

Year:  2010        PMID: 20831066      PMCID: PMC2902541          DOI: 10.1118/1.3447724

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  X-ray leakage during radiotherapy.

Authors:  S C Lillicrap; H M Morgan; J T Shakeshaft
Journal:  Br J Radiol       Date:  2000-07       Impact factor: 3.039

2.  Peripheral doses in CyberKnife radiosurgery.

Authors:  Paula L Petti; Cynthia F Chuang; Vernon Smith; David A Larson
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

3.  Use of peripheral dose data from uniform dynamic multileaf collimation fields to estimate out-of-field organ dose in patients treated employing sliding window intensity-modulated radiotherapy.

Authors:  Shamurailatpam Dayananda Sharma; Ritu Raj Upreti; Deepak Dattatray Deshpande
Journal:  Phys Med Biol       Date:  2006-05-24       Impact factor: 3.609

4.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

5.  Peripheral dose in ocular treatments with CyberKnife and Gamma Knife radiosurgery compared to proton radiotherapy.

Authors:  A Zytkovicz; I Daftari; T L Phillips; C F Chuang; L Verhey; P L Petti
Journal:  Phys Med Biol       Date:  2007-09-17       Impact factor: 3.609

6.  Peripheral dose measurement for CyberKnife radiosurgery with upgraded linac shielding.

Authors:  Cynthia F Chuang; David A Larson; Andrea Zytkovicz; Vernon Smith; Paula L Petti
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

7.  Cancer induction by radiotherapy: dose dependence and spatial relationship to irradiated volume.

Authors:  Wolfgang Dörr; Thomas Herrmann
Journal:  J Radiol Prot       Date:  2002-09       Impact factor: 1.394

8.  Dosimetric comparison of helical tomotherapy and Gamma Knife stereotactic radiosurgery for single brain metastasis.

Authors:  José A Peñagarícano; Yulong Yan; Chengyu Shi; Mark E Linskey; Vaneerat Ratanatharathorn
Journal:  Radiat Oncol       Date:  2006-08-03       Impact factor: 3.481

9.  Peripheral dose distributions for a linear accelerator equipped with a secondary multileaf collimator and universal wedge.

Authors:  Sasa Mutic; Jacqueline Esthappan; Eric E Klein
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

10.  Out-of-field dosimetry measurements for a helical tomotherapy system.

Authors:  Chester Ramsey; Rebecca Seibert; Stephen L Mahan; Dharmin Desai; Daniel Chase
Journal:  J Appl Clin Med Phys       Date:  2006-08-24       Impact factor: 2.102

View more
  4 in total

1.  Adaptive fractionated stereotactic Gamma Knife radiotherapy of meningioma using integrated stereotactic cone-beam-CT and adaptive re-planning (a-gkFSRT).

Authors:  F Stieler; F Wenz; Y Abo-Madyan; B Schweizer; M Polednik; C Herskind; F A Giordano; S Mai
Journal:  Strahlenther Onkol       Date:  2016-07-05       Impact factor: 3.621

2.  Peripheral doses in patients undergoing Cyberknife treatment for intracranial lesions. A single centre experience.

Authors:  Vassiliki Vlachopoulou; Christos Antypas; Harry Delis; Argyrios Tzouras; Nikolaos Salvaras; Dimitrios Kardamakis; George Panayiotakis
Journal:  Radiat Oncol       Date:  2011-11-14       Impact factor: 3.481

3.  Dosimetric comparison of three intensity-modulated radiation therapies for left breast cancer after breast-conserving surgery.

Authors:  Huai-Wen Zhang; Bo Hu; Chen Xie; Yun-Lai Wang
Journal:  J Appl Clin Med Phys       Date:  2018-03-10       Impact factor: 2.102

4.  Evaluating dose distributions of normal organs for patients undergoing VMAT therapy of nasopharyngeal carcinoma using Rando phantom and TLD-100H.

Authors:  Yun-Chih Chen; Hung-Chih Lin; Wei-Hou Lai; Hung-Chang Hung; Hsien-Chun Tseng; Chien-Yi Chen
Journal:  Technol Health Care       Date:  2022       Impact factor: 1.205

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.