BACKGROUND: The routine radiation therapy treatment planning does not include secondary radiation and peripheral doses resulting from radiotherapy exposure in patients with nasopharyngeal carcinoma (NPC) undergoing Volumetric Modulated Arc Therapy (VMAT) using an linear accelerator (linac) of Axesse (Elekta 2538).Entities:
Keywords: Axesse linac; ICRP 103; TLD-100H; Rando; Volumetric Modulated Arc Therapy (VMAT); effective dose (E); nasopharyngeal carcinoma (NPC)
Mesh:
Year: 2022 PMID: 35124609 PMCID: PMC9028652 DOI: 10.3233/THC-THC228031
Source DB: PubMed Journal: Technol Health Care ISSN: 0928-7329 Impact factor: 1.205
Figure 1.The NPC treatment of for VMAT (a) axial, (b) sagittal, and (c) coronal of 5 section of Rando views of Pinnacle planning system.
Measurement locations of the 52 points, the weighting factors () of organs or tissues are recommended by ICRP 103 (Eqs (1) and (2)
| Section of rando | Organ/tissue |
| Numbers of TLD |
|---|---|---|---|
| 1 | Skin | 0.01 | 96 |
| 1 | Brain | 0.01 | 15 |
| 3 | Lens | 0.017 | 6 |
| 4,5,24,30 | Bone surface | 0.01 | 39 |
| 6 | Oral mucosa | 0.017 | 3 |
| 6,7 | Salivary glands | 0.01 | 12 |
| 8,31 | Lymph nodes | 0.017 | 3 |
| 8,25,26,27 | Bone marrow | 0.12 | 12 |
| 9 | Esophagus | 0.04 | 6 |
| 10 | Thyroid | 0.04 | 6 |
| 11 | Lungs | 0.12 | 21 |
| 16 | Heart | 0.017 | 3 |
| 18 | Breast | 0.12 | 3 |
| 18 | Stomach/Spleen | 0.12 | 3 |
| 19 | Pancreas | 0.017 | 3 |
| 20 | Liver | 0.04 | 3 |
| 21,29 | Colon | 0.12 | 6 |
| 22 | Kidney | 0.017 | 3 |
| 23 | Small intestine | 0.017 | 3 |
| 28 | Bladder | 0.04 | 3 |
| 32 | Gonads | 0.08 | 3 |
| Total | 1.000 | 252 |
Remainder including: Lens, Oral mucosa, Lymph nodes, Heart, Pancreas, Kidney, Small intestine.
Random and systematic errors coming from the practical evaluation of radiation
| Source | One standard deviation |
|---|---|
| Measurement: | |
| TLD counting statistics | 3% |
| Others: | |
| 5% | |
| VMAT output photon dose | |
| TLD-100H calibration | 3% |
| Inserting positions of phantom organ or tissue | 5% |
| Density effects of PMMA phantom | 1.5% |
| Variation of Harshaw 3500 reader | 3% |
|
| 9.00% |
Figure 2.Equivalent doses (mSv/treatment) of organs of Rando during NPC treatment. was evaluated by inserting many TLDs into each organ. Mean values and errors shown as bars.
Figure 3.Assessing the skin dose () vs. Rando’s section during NPC treatment.