| Literature DB >> 16887031 |
José A Peñagarícano1, Yulong Yan, Chengyu Shi, Mark E Linskey, Vaneerat Ratanatharathorn.
Abstract
BACKGROUND: Helical Tomotherapy (HT) integrates linear accelerator and computerized tomography (CT) technology to deliver IMRT. Targets are localized (i.e. outlined as gross tumor volume [GTV] and planning target volume [PTV]) on the planning kVCT study while daily MVCT is used for correction of patient's set-up and assessment of inter-fraction anatomy changes. Based on dosimetric comparisons, this study aims to find dosimetric equivalency between single fraction HT and Gamma Knife stereotactic radiosurgery (GKSRS) for the treatment of single brain metastasis.Entities:
Mesh:
Year: 2006 PMID: 16887031 PMCID: PMC1557668 DOI: 10.1186/1748-717X-1-26
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Represents the Tomotherapy dose distribution (in Gy) for one of the five presented patients.
Figure 2Represents the Gamma Knife dose distribution (in percent of the prescribed dose) for one of the five presented patients.
Conformality Index (CI) of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery Plans in Patients with Single Brain Metastasis.
| TOMOTHERAPY | GAMMA KNIFE | |
| Patient #1 | 0.577 | 0.597 |
| Patient #2 | 0.562 | 0.618 |
| Patient #3 | 0.603 | 0.593 |
| Patient #4 | 0.644 | 0.696 |
| Patient #5 | 0.547 | 0.507 |
CI = (PTV volume within the prescribed iso-dose line)2/[(tumor volume)*(volume of the prescribed dose)]
Coverage of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery Plans in Patients with Single Brain Metastasis.
| TOMOTHERAPY | GAMMA KNIFE | |
| Patient #1 | 0.983 | 1.000 |
| Patient #2 | 0.944 | 0.990 |
| Patient #3 | 0.978 | 0.970 |
| Patient #4 | 0.962 | 0.962 |
| Patient #5 | 0.997 | 0.997 |
Coverage = (PTV volume within prescription iso-dose line)/(PTV volume)
Beam-on Treatment (minutes) Time of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery Plans in Patients with Single Brain Metastasis.
| TOMOTHERAPY | GAMMA KNIFE | |
| Patient #1 | 43.00 | 50.77 |
| Patient #2 | 34.00 | 21.00 |
| Patient #3 | 30.00 | 14.16 |
| Patient #4 | 36.00 | 36.70 |
| Patient #5 | 49.00 | 21.00 |
Isodose volumes (cubic centimeter) of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery Plans in Patients with Single Brain Metastasis.
| Isodose Line | Patient #1 | Patient #2 | Patient #3 | Patient #4 | Patient #5 | |||||
| Tomo Vol | GK Vol | Tomo Vol | GK Vol | Tomo Vol | GK Vol | Tomo Vol | GK Vol | Tomo Vol | GK Vol | |
| 40 Gy | 0.009 | 0.009 | 0.001 | 0.001 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 |
| 35 Gy | 0.082 | 0.136 | 0.039 | 0.191 | 0.000 | 0.000 | 0.012 | 0.009 | 0.136 | 0.021 |
| 30 Gy | 0.208 | 0.332 | 0.159 | 0.320 | 0.028 | 0.016 | 0.096 | 0.138 | 0.756 | 0.119 |
| 25 Gy | 0.430 | 0.540 | 0.371 | 0.431 | 0.195 | 0.067 | 0.277 | 0.492 | 1.685 | 0.689 |
| 20 Gy | 0.748 | 0.905 | 0.709 | 0.589 | 0.476 | 0.225 | 0.584 | 0.938 | 3.093 | 2.418 |
| 15 Gy | 1.315 | 1.417 | 1.306 | 0.877 | 0.961 | 0.704 | 1.099 | 1.665 | 5.441 | 4.750 |
| 10 Gy | 2.851 | 2.678 | 2.865 | 1.545 | 2.111 | 1.344 | 2.363 | 2.997 | 11.986 | 9.339 |
| 5 Gy | 11.463 | 7.643 | 11.394 | 4.230 | 8.739 | 3.774 | 9.196 | 8.309 | 47.574 | 22.885 |
Tomo Vol = Tomotherapy Volume.
GK Vol = Gamma Knife Volume.
Minimum dose (Gy) to 100% of the PTV volume.
| TOMOTHERAPY | GAMMA KNIFE | |
| Patient #1 | 19.3 | 22.8 |
| Patient #2 | 16.4 | 17.6 |
| Patient #3 | 12.6 | 14.1 |
| Patient #4 | 17 | 19.1 |
| Patient #5 | 15.4 | 15.2 |