Quantitative imaging in breast tomosynthesis and CT: comparison of detection and estimation task performance.

Samuel Richard1, Ehsan Samei.   

Abstract

PURPOSE: This work investigates a framework for modeling volumetric breast imaging to compare detection and estimation task performance and optimize quantitative breast imaging.
METHODS: Volumetric reconstructions of a breast phantom, which incorporated electronic, quantum, and anatomical noise with embedded spherical lesions, were simulated over a range of acquisition angles varying from 4 degrees to 204 degrees with a constant total acquisition dose of 1.5 mGy. A maximum likelihood estimator was derived in terms of the noise power spectrum, which yielded figures of merit for quantitative imaging performance in terms of accuracy and precision. These metrics were computed for estimation of lesion area, volume, and location. Estimation task performance was optimized as a function of acquisition angle and compared to the performance of a more conventional lesion detection task.
RESULTS: Results revealed tradeoffs between electronic, quantum, and anatomical noise. The detection of a 4 mm sphere was optimal at an acquisition angle of 84 degrees, where reconstructed images using a smaller acquisition angle exhibited increased anatomical noise and reconstructed images using a larger acquisition angle exhibited increased quantum and electronic noise. For all estimation tasks, accuracy was found to be fairly constant as a function acquisition angle indicating adequate system calibration, whereas a more significant dependence on acquisition angle was observed for precision performance. Precision for the 2D area estimation task was optimal at approximately 104 degrees, while precision of the 3D volume estimation task was optimal at larger angles (approximately 124 degrees). Precision for the localization task showed orientation dependence where localization was significantly inferior in the depth direction. Overall, precision for localization was optimal at larger angles (i.e., > 125 degrees) compared to the size estimation tasks. Results suggested that for quantitative imaging tasks, the acquisition angle should be larger than currently used in conventional breast tomosynthesis for lesion detection.
CONCLUSIONS: Analysis of quantitative imaging performance using Fourier-based metrics highlights the difference between estimation and detection task in volumetric breast imaging and provides a meaningful framework for optimizing the performance of breast imaging systems for quantitative imaging applications.

Mesh:

Year:  2010        PMID: 20632574     DOI: 10.1118/1.3429025

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  On the orientation of mammographic structure.

Authors:  I Reiser; S Lee; R M Nishikawa
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

2.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

3.  Beyond noise power in 3D computed tomography: the local NPS and off-diagonal elements of the Fourier domain covariance matrix.

Authors:  Angel R Pineda; Daniel J Tward; Antonio Gonzalez; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

4.  A statistical, task-based evaluation method for three-dimensional x-ray breast imaging systems using variable-background phantoms.

Authors:  Subok Park; Robert Jennings; Haimo Liu; Aldo Badano; Kyle Myers
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

5.  The effect of angular dose distribution on the detection of microcalcifications in digital breast tomosynthesis.

Authors:  Yue-Houng Hu; Wei Zhao
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

6.  The effect of amorphous selenium detector thickness on dual-energy digital breast imaging.

Authors:  Yue-Houng Hu; Wei Zhao
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

Review 7.  A review of breast tomosynthesis. Part II. Image reconstruction, processing and analysis, and advanced applications.

Authors:  Ioannis Sechopoulos
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

8.  Correlation of free-response and receiver-operating-characteristic area-under-the-curve estimates: results from independently conducted FROC∕ROC studies in mammography.

Authors:  Federica Zanca; Stephen L Hillis; Filip Claus; Chantal Van Ongeval; Valerie Celis; Veerle Provoost; Hong-Jun Yoon; Hilde Bosmans
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

9.  Task-based strategy for optimized contrast enhanced breast imaging: analysis of six imaging techniques for mammography and tomosynthesis.

Authors:  Lynda C Ikejimba; Nooshin Kiarashi; Sujata V Ghate; Ehsan Samei; Joseph Y Lo
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

10.  Investigating simulation-based metrics for characterizing linear iterative reconstruction in digital breast tomosynthesis.

Authors:  Sean D Rose; Adrian A Sanchez; Emil Y Sidky; Xiaochuan Pan
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.