Real-time target position estimation using stereoscopic kilovoltage/megavoltage imaging and external respiratory monitoring for dynamic multileaf collimator tracking.

Byungchul Cho1, Per Rugaard Poulsen, Amit Sawant, Dan Ruan, Paul J Keall.   

Abstract

PURPOSE: To develop a real-time target position estimation method using stereoscopic kilovoltage (kV)/megavoltage (MV) imaging and external respiratory monitoring, and to investigate the performance of a dynamic multileaf collimator tracking system using this method. METHODS AND MATERIALS: The real-time three-dimensional internal target position estimation was established by creating a time-varying correlation model that connected the external respiratory signals with the internal target motion measured intermittently using kV/MV imaging. The method was integrated into a dynamic multileaf collimator tracking system. Tracking experiments were performed for 10 thoracic/abdominal traces. A three-dimensional motion platform carrying a gold marker and a separate one-dimensional motion platform were used to reproduce the target and external respiratory motion, respectively. The target positions were detected by kV (1 Hz) and MV (5.2 Hz) imaging, and external respiratory motion was captured by an optical system (30 Hz). The beam-target alignment error was quantified as the positional difference between the target and circular beam center on the MV images acquired during tracking. The correlation model error was quantified by comparing a model estimate and measured target positions.
RESULTS: The root-mean-square errors in the beam-target alignment that had ranged from 3.1 to 7.6 mm without tracking were reduced to <1.5 mm with tracking, except during the model building period (6 s). The root-mean-square error in the correlation model was submillimeters in all directions.
CONCLUSION: A novel real-time target position estimation method was developed and integrated into a dynamic multileaf collimator tracking system and demonstrated an average submillimeter geometric accuracy after initializing the internal/external correlation model. The method used hardware tools available on linear accelerators and therefore shows promise for clinical implementation.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20615623      PMCID: PMC2953601          DOI: 10.1016/j.ijrobp.2010.02.052

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  15 in total

1.  A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup.

Authors:  Christoph Bert; Katherine G Metheany; Karen Doppke; George T Y Chen
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

2.  Determination of maximum leaf velocity and acceleration of a dynamic multileaf collimator: implications for 4D radiotherapy.

Authors:  K Wijesooriya; C Bartee; J V Siebers; S S Vedam; P J Keall
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

3.  Geometric accuracy of a real-time target tracking system with dynamic multileaf collimator tracking system.

Authors:  Paul J Keall; Herbert Cattell; Damodar Pokhrel; Sonja Dieterich; Kenneth H Wong; Martin J Murphy; S Sastry Vedam; Krishni Wijesooriya; Radhe Mohan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-08-01       Impact factor: 7.038

4.  Respiratory gated beam delivery cannot facilitate margin reduction, unless combined with respiratory correlated image guidance.

Authors:  Stine S Korreman; Trine Juhler-Nøttrup; Arthur L Boyer
Journal:  Radiother Oncol       Date:  2007-11-26       Impact factor: 6.280

5.  Combined kV and MV imaging for real-time tracking of implanted fiducial markers.

Authors:  R D Wiersma; Weihua Mao; L Xing
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

6.  Evaluation of an infrared camera and X-ray system using implanted fiducials in patients with lung tumors for gated radiation therapy.

Authors:  Twyla R Willoughby; Alan R Forbes; Daniel Buchholz; Katja M Langen; Thomas H Wagner; Omar A Zeidan; Patrick A Kupelian; Sanford L Meeks
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-08-02       Impact factor: 7.038

7.  Management of three-dimensional intrafraction motion through real-time DMLC tracking.

Authors:  Amit Sawant; Raghu Venkat; Vikram Srivastava; David Carlson; Sergey Povzner; Herb Cattell; Paul Keall
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

8.  A monoscopic method for real-time tumour tracking using combined occasional x-ray imaging and continuous respiratory monitoring.

Authors:  Byungchul Cho; Yelin Suh; Sonja Dieterich; Paul J Keall
Journal:  Phys Med Biol       Date:  2008-05-06       Impact factor: 3.609

9.  Inference of hysteretic respiratory tumor motion from external surrogates: a state augmentation approach.

Authors:  D Ruan; J A Fessler; J M Balter; R I Berbeco; S Nishioka; H Shirato
Journal:  Phys Med Biol       Date:  2008-05-06       Impact factor: 3.609

10.  Respiration tracking in radiosurgery.

Authors:  Achim Schweikard; Hiroya Shiomi; John Adler
Journal:  Med Phys       Date:  2004-10       Impact factor: 4.071

View more
  13 in total

1.  Feasibility of low-dose single-view 3D fiducial tracking concurrent with external beam delivery.

Authors:  Michael A Speidel; Brian P Wilfley; Annie Hsu; Dimitre Hristov
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

Review 2.  Review of Real-Time 3-Dimensional Image Guided Radiation Therapy on Standard-Equipped Cancer Radiation Therapy Systems: Are We at the Tipping Point for the Era of Real-Time Radiation Therapy?

Authors:  Paul J Keall; Doan Trang Nguyen; Ricky O'Brien; Pengpeng Zhang; Laura Happersett; Jenny Bertholet; Per R Poulsen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2018-04-14       Impact factor: 7.038

3.  Experimental investigation of a general real-time 3D target localization method using sequential kV imaging combined with respiratory monitoring.

Authors:  Byungchul Cho; Per Poulsen; Dan Ruan; Amit Sawant; Paul J Keall
Journal:  Phys Med Biol       Date:  2012-10-24       Impact factor: 3.609

4.  Toward the development of intrafraction tumor deformation tracking using a dynamic multi-leaf collimator.

Authors:  Yuanyuan Ge; Ricky T O'Brien; Chun-Chien Shieh; Jeremy T Booth; Paul J Keall
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

5.  Four-dimensional planning for motion synchronized dose delivery in lung stereotactic body radiation therapy.

Authors:  Hidenobu Tachibana; Amit Sawant
Journal:  Radiother Oncol       Date:  2016-04-30       Impact factor: 6.280

6.  Motion management during IMAT treatment of mobile lung tumors--a comparison of MLC tracking and gated delivery.

Authors:  Marianne Falk; Tobias Pommer; Paul Keall; Stine Korreman; Gitte Persson; Per Poulsen; Per Munck af Rosenschöld
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

7.  Technical Note: 3D localization of lung tumors on cone beam CT projections via a convolutional recurrent neural network.

Authors:  Chuang Wang; Margie Hunt; Lei Zhang; Andreas Rimner; Ellen Yorke; Michael Lovelock; Xiang Li; Tianfang Li; Gig Mageras; Pengpeng Zhang
Journal:  Med Phys       Date:  2020-01-28       Impact factor: 4.071

8.  Design and validation of a MV/kV imaging-based markerless tracking system for assessing real-time lung tumor motion.

Authors:  Pengpeng Zhang; Margie Hunt; Arina B Telles; Hai Pham; Michael Lovelock; Ellen Yorke; Guang Li; Laura Happersett; Andreas Rimner; Gig Mageras
Journal:  Med Phys       Date:  2018-11-13       Impact factor: 4.071

9.  Monitoring tumor motion by real time 2D/3D registration during radiotherapy.

Authors:  Christelle Gendrin; Hugo Furtado; Christoph Weber; Christoph Bloch; Michael Figl; Supriyanto Ardjo Pawiro; Helmar Bergmann; Markus Stock; Gabor Fichtinger; Dietmar Georg; Wolfgang Birkfellner
Journal:  Radiother Oncol       Date:  2011-08-30       Impact factor: 6.280

10.  Investigation of the change in marker geometry during respiration motion: a preliminary study for dynamic-multi-leaf real-time tumor tracking.

Authors:  Rie Yamazaki; Seiko Nishioka; Hiroyuki Date; Hiroki Shirato; Takao Koike; Takeshi Nishioka
Journal:  Radiat Oncol       Date:  2012-12-18       Impact factor: 3.481