Literature DB >> 16887290

Evaluation of an infrared camera and X-ray system using implanted fiducials in patients with lung tumors for gated radiation therapy.

Twyla R Willoughby1, Alan R Forbes, Daniel Buchholz, Katja M Langen, Thomas H Wagner, Omar A Zeidan, Patrick A Kupelian, Sanford L Meeks.   

Abstract

PURPOSE: To report on the initial clinical use of a commercially available system to deliver gated treatment using implanted fiducials, in-room kV X-rays, and an infrared camera tracking system. METHODS AND MATERIALS: ExacTrac Adaptive Gating from BrainLab is a localization system using infrared cameras and X-rays. Gating signals are the patient's breathing pattern obtained from infrared reflectors on the patient. kV X-rays of an implanted fiducial are synchronized to the breathing pattern. After localization and shift of the patient to isocenter, the breathing pattern is used to gate the radiation. Feasibility tests included localization accuracy, radiation output constancy, and dose distributions with gating. Clinical experience is reported on treatment of patients with small lung lesions.
RESULTS: Localization accuracy of a moving target with gating was 1.7 mm. Dose constancy measurements showed insignificant change in output with gating. Improvements of dose distributions on moving targets improved with gating. Eleven patients with lung lesions were implanted with 20 mmx0.7 mm gold coil (Visicoil). The implanted fiducial was used to localize and treat the patients with gating. Treatment planning and repeat computed tomographic scans showed that the change in center of gross target volume (GTV) to implanted marker averaged 2.47 mm due in part to asymmetric tumor shrinkage.
CONCLUSION: ExacTrac Adaptive Gating has been used to treat lung lesions. Initial system evaluation verified its accuracy and usability. Implanted fiducials are visible in X-rays and did not migrate.

Entities:  

Mesh:

Year:  2006        PMID: 16887290     DOI: 10.1016/j.ijrobp.2006.05.029

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  23 in total

1.  The dosimetric impact of inversely optimized arc radiotherapy plan modulation for real-time dynamic MLC tracking delivery.

Authors:  Marianne Falk; Tobias Larsson; Paul Keall; Byung Chul Cho; Marianne Aznar; Stine Korreman; Per Poulsen; Per Munck Af Rosenschold
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

2.  A multi-GPU real-time dose simulation software framework for lung radiotherapy.

Authors:  A P Santhanam; Y Min; H Neelakkantan; N Papp; S L Meeks; P A Kupelian
Journal:  Int J Comput Assist Radiol Surg       Date:  2012-04-27       Impact factor: 2.924

3.  Fast internal marker tracking algorithm for onboard MV and kV imaging systems.

Authors:  W Mao; R D Wiersma; L Xing
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

4.  Mitigating errors in external respiratory surrogate-based models of tumor position.

Authors:  Kathleen T Malinowski; Thomas J McAvoy; Rohini George; Sonja Dieterich; Warren D D'Souza
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-04-01       Impact factor: 7.038

5.  Evaluation of a template-based algorithm for markerless lung tumour localization on single- and dual-energy kilovoltage images.

Authors:  Alec M Block; Rakesh Patel; Murat Surucu; Matthew M Harkenrider; John C Roeske
Journal:  Br J Radiol       Date:  2016-10-12       Impact factor: 3.039

6.  Real-time target position estimation using stereoscopic kilovoltage/megavoltage imaging and external respiratory monitoring for dynamic multileaf collimator tracking.

Authors:  Byungchul Cho; Per Rugaard Poulsen; Amit Sawant; Dan Ruan; Paul J Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-07-07       Impact factor: 7.038

7.  The correlation of tissue motion within the lung: implications on fiducial based treatments.

Authors:  Ryan L Smith; Deshan Yang; Andrew Lee; Martin L Mayse; Dan A Low; Parag J Parikh
Journal:  Med Phys       Date:  2011-11       Impact factor: 4.071

8.  Real-time dynamic MLC tracking for inversely optimized arc radiotherapy.

Authors:  Marianne Falk; Per Munck af Rosenschöld; Paul Keall; Herbert Cattell; Byung Chul Cho; Per Poulsen; Sergey Povzner; Amit Sawant; Jens Zimmerman; Stine Korreman
Journal:  Radiother Oncol       Date:  2010-01-19       Impact factor: 6.280

9.  Method comparison of automated matching software-assisted cone-beam CT and stereoscopic kilovoltage x-ray positional verification image-guided radiation therapy for head and neck cancer: a prospective analysis.

Authors:  Clifton D Fuller; Todd J Scarbrough; Jan-Jakob Sonke; Coen R N Rasch; Mehee Choi; Joe Y Ting; Samuel J Wang; Niko Papanikolaou; David I Rosenthal
Journal:  Phys Med Biol       Date:  2009-11-24       Impact factor: 3.609

10.  Robust fluoroscopic tracking of fiducial markers: exploiting the spatial constraints.

Authors:  Rui Li; Gregory Sharp
Journal:  Phys Med Biol       Date:  2013-02-26       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.