Literature DB >> 16863935

Geometric accuracy of a real-time target tracking system with dynamic multileaf collimator tracking system.

Paul J Keall1, Herbert Cattell, Damodar Pokhrel, Sonja Dieterich, Kenneth H Wong, Martin J Murphy, S Sastry Vedam, Krishni Wijesooriya, Radhe Mohan.   

Abstract

PURPOSE: Dynamically compensating for target motion during radiotherapy will increase treatment accuracy. A laboratory system for real-time target tracking with a dynamic MLC has been developed. In this study, the geometric accuracy limits of this DMLC target tracking system were evaluated. METHODS AND MATERIALS: A motion simulator was programmed to follow patient-derived tumor motion paths, parallel to the leaf motion direction. A target attached to the simulator was optically tracked, and the leaf positions adjusted to continually align the DMLC beam aperture to the target. Analysis of the tracking accuracy was based on video images of the target and beam alignment. The system response time was determined and the tracking error measured. Response time-corrected tracking accuracy was also calculated to investigate the accuracy limits of an improved system.
RESULTS: The response time of the system is 160 +/- 2 ms. The geometric precision for tracking patient motion is 0.6 to 1.1 mm (1 sigma) for the 3 patient datasets tested, with tracking errors relative to the original patient motion of 35, 40, and 100%.
CONCLUSIONS: A DMLC target tracking system has been developed that can account for detected motion parallel to the leaf motion direction. The tracking error has a negligible systematic component. Reducing the response time will further increase the overall system accuracy.

Entities:  

Mesh:

Year:  2006        PMID: 16863935     DOI: 10.1016/j.ijrobp.2006.04.038

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  39 in total

1.  Verification of MLC based real-time tumor tracking using an electronic portal imaging device.

Authors:  Sarah Han-Oh; Byong Yong Yi; Fritz Lerma; Barry L Berman; Minzhi Gui; Cedric Yu
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  The impact of leaf width and plan complexity on DMLC tracking of prostate intensity modulated arc therapy.

Authors:  Tobias Pommer; Marianne Falk; Per Rugaard Poulsen; Paul J Keall; Ricky T O'Brien; Per Munck af Rosenschöld
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

3.  Use of dMLC for implementation of dynamic respiratory-gated radiation therapy.

Authors:  Eric W Pepin; Huanmei Wu; Hiroki Shirato
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

4.  Experimental investigation of a moving averaging algorithm for motion perpendicular to the leaf travel direction in dynamic MLC target tracking.

Authors:  Jai-Woong Yoon; Amit Sawant; Yelin Suh; Byung-Chul Cho; Tae-Suk Suh; Paul Keall
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

5.  Registration of clinical volumes to beams-eye-view images for real-time tracking.

Authors:  Jonathan H Bryant; Joerg Rottmann; John H Lewis; Pankaj Mishra; Paul J Keall; Ross I Berbeco
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

Review 6.  Are three doses of stereotactic ablative radiotherapy (SABR) more effective than 30 doses of conventional radiotherapy?

Authors:  Anna O Simeonova; Katharina Fleckenstein; Hansjörg Wertz; Anian Frauenfeld; Judit Boda-Heggemann; Frank Lohr; Frederik Wenz
Journal:  Transl Lung Cancer Res       Date:  2012-03

7.  Interobserver variability of patient positioning using four different CT datasets for image registration in lung stereotactic body radiotherapy.

Authors:  Markus Oechsner; Barbara Chizzali; Michal Devecka; Stefan Münch; Stephanie Elisabeth Combs; Jan Jakob Wilkens; Marciana Nona Duma
Journal:  Strahlenther Onkol       Date:  2017-07-19       Impact factor: 3.621

8.  Determination of patient-specific internal gross tumor volumes for lung cancer using four-dimensional computed tomography.

Authors:  Muthuveni Ezhil; Sastry Vedam; Peter Balter; Bum Choi; Dragan Mirkovic; George Starkschall; Joe Y Chang
Journal:  Radiat Oncol       Date:  2009-01-27       Impact factor: 3.481

9.  Experimental investigation of a general real-time 3D target localization method using sequential kV imaging combined with respiratory monitoring.

Authors:  Byungchul Cho; Per Poulsen; Dan Ruan; Amit Sawant; Paul J Keall
Journal:  Phys Med Biol       Date:  2012-10-24       Impact factor: 3.609

10.  Management of three-dimensional intrafraction motion through real-time DMLC tracking.

Authors:  Amit Sawant; Raghu Venkat; Vikram Srivastava; David Carlson; Sergey Povzner; Herb Cattell; Paul Keall
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

View more