BACKGROUND: Colonoscopy (CSPY) for colorectal cancer screening has several limitations. Colon Capsule Endoscopy (PillCam Colon, CCE) was compared to CSPY under routine screening conditions.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20565828 PMCID: PMC2905323 DOI: 10.1186/1471-230X-10-66
Source DB: PubMed Journal: BMC Gastroenterol ISSN: 1471-230X Impact factor: 3.067
Figure 1Original size of PillCam® Colon Capsule 11 mm × 31 mm
Summary of bowel preparation
| all day | Low-fibre diet |
| all day | liquid diet |
| all day | clear liquids only |
| 18:00-19:30 | 2 Liters of Cololyt® (1 cup every 10-15 minutes) |
| 7:00-8:30 | 2 Liters of Cololyt® (1 cup every 10-15 minutes) |
| 11:00 | 20 mg Domperidone (Motilium®) with a cup of water |
| 11:15 | Ingestion of PillCam® Colon with a cup of water |
| until 13:00 | no food/liquid ingestion |
| 13:00 | Real time view assessment if capsule left stomach, then Booster dose of 45 ml Colophos® + 1 liter of water |
| 17:00 | Real time view assessment if capsule left stomach, then Booster dose of 30 ml Colophos® + 1 liter of water *) |
| 19:30 | 10 mg Bisacodylum rectal suppository *) |
| 22:00 | End of examination |
| Conventional colonoscopy | |
| *) This action was taken only if PillCam® Colon had not been excreted by this time. |
Description of colon cleansing level
| Grade | Description |
|---|---|
| 1 | clean, small amount of feces/dark fluid, good assessability |
| 2 | moderate preparation due to remaining feces/dark fluid, limited assessability but clean enough to preclude significant lesions |
| 3 | poor preparation, enough feces/fluid to strongly limit assessability |
Indications/symptoms (n = 56)
| Screening | 23 (41%) |
| Family history | 4 (7%) |
| Follow-up after polypectomy | 3 (5%) |
| Abdominal pain | 8 (14%) |
| Bleeding frank | 8 (14%) |
| Anemia | 2 (4%) |
| Change in bowel habits | 6 (11%) |
| Diarrhea | 1 (2%) |
| Weight loss | 1 (2%) |
CCE: Sensitivity, Specificity, PPV and NPV for any size polyps and significant size (>5 mm) polyps
| For polyps of any size | For polyps >5 mm | |
|---|---|---|
| Sensitivity % (95% CI) | 79 (61 - 90) | 50 (19 - 81) |
| Specificity % (95% CI) | 54 (35 - 70) | 76 (36 - 86) |
| Positive predictive value % | 63 | 20 |
| Negative predictive value % | 71 | 93 |
Figure 2Images of colon polyps on Colonoscopy (left) vs. CCE (right). A: Polyp smaller or as big as 5 mm, B, C: Polyps bigger than 5 mm