Radiation dose reduction using a CdZnTe-based computed tomography system: comparison to flat-panel detectors.

Q Le Huy1, Justin L Ducote, Sabee Molloi.   

Abstract

PURPOSE: Although x-ray projection mammography has been very effective in early detection of breast cancer, its utility is reduced in the detection of small lesions that are occult or in dense breasts. One drawback is that the inherent superposition of parenchymal structures makes visualization of small lesions difficult. Breast computed tomography using flat-panel detectors has been developed to address this limitation by producing three-dimensional data while at the same time providing more comfort to the patients by eliminating breast compression. Flat panels are charge integrating detectors and therefore lack energy resolution capability. Recent advances in solid state semiconductor x-ray detector materials and associated electronics allow the investigation of x-ray imaging systems that use a photon counting and energy discriminating detector, which is the subject of this article.
METHODS: A small field-of-view computed tomography (CT) system that uses CdZnTe (CZT) photon counting detector was compared to one that uses a flat-panel detector for different imaging tasks in breast imaging. The benefits afforded by the CZT detector in the energy weighting modes were investigated. Two types of energy weighting methods were studied: Projection based and image based. Simulation and phantom studies were performed with a 2.5 cm polymethyl methacrylate (PMMA) cylinder filled with iodine and calcium contrast objects. Simulation was also performed on a 10 cm breast specimen.
RESULTS: The contrast-to-noise ratio improvements as compared to flat-panel detectors were 1.30 and 1.28 (projection based) and 1.35 and 1.25 (image based) for iodine over PMMA and hydroxylapatite over PMMA, respectively. Corresponding simulation values were 1.81 and 1.48 (projection based) and 1.85 and 1.48 (image based). Dose reductions using the CZT detector were 52.05% and 49.45% for iodine and hydroxyapatite imaging, respectively. Image-based weighting was also found to have the least beam hardening effect.
CONCLUSIONS: The results showed that a CT system using an energy resolving detector reduces the dose to the patient while maintaining image quality for various breast imaging tasks.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20384260      PMCID: PMC2842286          DOI: 10.1118/1.3312435

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

1.  Detective quantum efficiency dependence on x-ray energy weighting in mammography.

Authors:  R N Cahn; B Cederström; M Danielsson; A Hall; M Lundqvist; D Nygren
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

2.  A fuzzy c-means (FCM)-based approach for computerized segmentation of breast lesions in dynamic contrast-enhanced MR images.

Authors:  Weijie Chen; Maryellen L Giger; Ulrich Bick
Journal:  Acad Radiol       Date:  2006-01       Impact factor: 3.173

3.  Dual-energy mammography: simulation studies.

Authors:  K Bliznakova; Z Kolitsi; N Pallikarakis
Journal:  Phys Med Biol       Date:  2006-08-22       Impact factor: 3.609

4.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography.

Authors:  J P Schlomka; E Roessl; R Dorscheid; S Dill; G Martens; T Istel; C Bäumer; C Herrmann; R Steadman; G Zeitler; A Livne; R Proksa
Journal:  Phys Med Biol       Date:  2008-07-08       Impact factor: 3.609

5.  Feasibility of real time dual-energy imaging based on a flat panel detector for coronary artery calcium quantification.

Authors:  Tong Xu; Justin L Ducote; Jerry T Wong; Sabee Molloi
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

6.  Tilted angle CZT detector for photon counting/energy weighting x-ray and CT imaging.

Authors:  Polad M Shikhaliev
Journal:  Phys Med Biol       Date:  2006-08-15       Impact factor: 3.609

7.  Dual-energy subtraction mammography.

Authors:  T Asaga; C Masuzawa; A Yoshida; H Matsuura
Journal:  J Digit Imaging       Date:  1995-02       Impact factor: 4.056

8.  A method for selective tissue and bone visualization using dual energy scanned projection radiography.

Authors:  W R Brody; G Butt; A Hall; A Macovski
Journal:  Med Phys       Date:  1981 May-Jun       Impact factor: 4.071

9.  Generalized image combinations in dual KVP digital radiography.

Authors:  L A Lehmann; R E Alvarez; A Macovski; W R Brody; N J Pelc; S J Riederer; A L Hall
Journal:  Med Phys       Date:  1981 Sep-Oct       Impact factor: 4.071

10.  Mammographic density and physical assessment of the breast.

Authors:  C A Swann; D B Kopans; K A McCarthy; G White; D A Hall
Journal:  AJR Am J Roentgenol       Date:  1987-03       Impact factor: 3.959

View more
  20 in total

1.  Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study.

Authors:  H-M Cho; H Ding; B P Ziemer; S Molloi
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  Quantification of breast density with spectral mammography based on a scanned multi-slit photon-counting detector: a feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Phys Med Biol       Date:  2012-07-06       Impact factor: 3.609

3.  Spectral response model for a multibin photon-counting spectral computed tomography detector and its applications.

Authors:  Xuejin Liu; Mats Persson; Hans Bornefalk; Staffan Karlsson; Cheng Xu; Mats Danielsson; Ben Huber
Journal:  J Med Imaging (Bellingham)       Date:  2015-09-11

4.  Evaluation of position-estimation methods applied to CZT-based photon-counting detectors for dedicated breast CT.

Authors:  Andrey Makeev; Martin Clajus; Scott Snyder; Xiaolang Wang; Stephen J Glick
Journal:  J Med Imaging (Bellingham)       Date:  2015-04-28

5.  Perfusion debate.

Authors:  H William Strauss
Journal:  J Nucl Cardiol       Date:  2017-08-02       Impact factor: 5.952

6.  Tight-frame based iterative image reconstruction for spectral breast CT.

Authors:  Bo Zhao; Hao Gao; Huanjun Ding; Sabee Molloi
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

7.  Breast composition measurement with a cadmium-zinc-telluride based spectral computed tomography system.

Authors:  Huanjun Ding; Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

8.  Objective assessment of task performance: a comparison of two FFDM detectors using an anthropomorphic breast phantom.

Authors:  Andrey Makeev; Lynda C Ikejimba; Jesse Salad; Stephen J Glick
Journal:  J Med Imaging (Bellingham)       Date:  2019-10-17

9.  Characteristic performance evaluation of a photon counting Si strip detector for low dose spectral breast CT imaging.

Authors:  Hyo-Min Cho; William C Barber; Huanjun Ding; Jan S Iwanczyk; Sabee Molloi
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

10.  Breast tissue characterization with photon-counting spectral CT imaging: a postmortem breast study.

Authors:  Huanjun Ding; Michael J Klopfer; Justin L Ducote; Fumitaro Masaki; Sabee Molloi
Journal:  Radiology       Date:  2014-05-07       Impact factor: 11.105

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.