Accuracy of the SEVEN continuous glucose monitoring system: comparison with frequently sampled venous glucose measurements.

Howard C Zisser1, Timothy S Bailey, Sherwyn Schwartz, Robert E Ratner, Jonathan Wise.   

Abstract

BACKGROUND: The purpose of this study was to compare the accuracy of measurements obtained from the DexCom SEVEN system with Yellow Springs Instrument (YSI) laboratory measurements of venous blood glucose.
METHODS: Seventy-two subjects with insulin-requiring diabetes, aged 18-71, were enrolled in a multicenter, prospective single-arm study. All participants wore the SEVEN continuous glucose monitoring (CGM) system for one, 7-day wear period. Calibration with capillary finger stick measurements was performed 2 hours after sensor insertion and once every 12 hours thereafter. A subset of subjects (28) wore two systems simultaneously to assess precision. All subjects participated in one, 10-hour in-clinic session on day 1, 4, or 7 of the study to compare CGM measurements against a laboratory method (YSI analyzer) using venous measurements taken once every 20 minutes. Carbohydrate consumption and insulin dosing were adjusted in order to obtain a broad range of glucose values.
RESULTS: Comparison of CGM measurements with the laboratory reference method (n = 2318) gave mean and median absolute relative differences (ARDs) of 16.7 and 13.2%, respectively. The percentage was 70.4% in the clinically accurate Clarke error grid A zone and 27.5% in the benign error B zone. Performance of the SEVEN system was consistent over time with mean and median ARD lowest on day 7 as compared to YSI (13.3 and 10.2%, respectively). Average sensor time lag was 5 minutes.
CONCLUSIONS: Measurements of the DexCom SEVEN system were found to be consistent and accurate compared with venous measurements made using a laboratory reference method over 7 days of wear. 2009 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20144429      PMCID: PMC2769895          DOI: 10.1177/193229680900300519

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  29 in total

1.  Continuous glucose monitoring used to adjust diabetes therapy improves glycosylated hemoglobin: a pilot study.

Authors:  B W Bode; T M Gross; K R Thornton; J J Mastrototaro
Journal:  Diabetes Res Clin Pract       Date:  1999-12       Impact factor: 5.602

Review 2.  The economic burden of diabetes and the benefits of improved glycemic control: the potential role of a continuous glucose monitoring system.

Authors:  J S Skyler
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

3.  Continuous subcutaneous glucose monitoring in children with type 1 diabetes.

Authors:  H P Chase; L M Kim; S L Owen; T A MacKenzie; G J Klingensmith; R Murtfeldt; S K Garg
Journal:  Pediatrics       Date:  2001-02       Impact factor: 7.124

4.  Use of the GlucoWatch biographer in children with type 1 diabetes.

Authors:  H Peter Chase; Mary D Roberts; Clare Wightman; Georgeanna Klingensmith; Satish K Garg; Michelle Van Wyhe; Shashi Desai; Wesley Harper; Margarita Lopatin; Miroslaw Bartkowiak; Janet Tamada; Richard C Eastman
Journal:  Pediatrics       Date:  2003-04       Impact factor: 7.124

5.  The Continuous Glucose Monitoring System (CGMS) in type 1 diabetic children is the way to reduce hypoglycemic risk.

Authors:  R Schiaffini; P Ciampalini; A Fierabracci; S Spera; P Borrelli; G F Bottazzo; A Crinò
Journal:  Diabetes Metab Res Rev       Date:  2002 Jul-Aug       Impact factor: 4.876

6.  Continuous subcutaneous glucose monitoring improved metabolic control in pediatric patients with type 1 diabetes: a controlled crossover study.

Authors:  Johnny Ludvigsson; Ragnar Hanas
Journal:  Pediatrics       Date:  2003-05       Impact factor: 7.124

7.  Sustained effect of intensive treatment of type 1 diabetes mellitus on development and progression of diabetic nephropathy: the Epidemiology of Diabetes Interventions and Complications (EDIC) study.

Authors: 
Journal:  JAMA       Date:  2003-10-22       Impact factor: 56.272

8.  The accuracy of the CGMS in children with type 1 diabetes: results of the diabetes research in children network (DirecNet) accuracy study.

Authors: 
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

9.  Improved metabolic control in diabetic adolescents using the continuous glucose monitoring system (CGMS).

Authors:  P Schaepelynck-Bélicar; Ph Vague; G Simonin; V Lassmann-Vague
Journal:  Diabetes Metab       Date:  2003-12       Impact factor: 6.041

10.  Comparison of accuracy and safety of the SEVEN and the Navigator continuous glucose monitoring systems.

Authors:  Satish K Garg; James Smith; Christie Beatson; Benita Lopez-Baca; Mary Voelmle; Peter A Gottlieb
Journal:  Diabetes Technol Ther       Date:  2009-02       Impact factor: 6.118

View more
  36 in total

1.  Continuous glucose monitoring considerations for the development of a closed-loop artificial pancreas system.

Authors:  D Barry Keenan; Benyamin Grosman; Harry W Clark; Anirban Roy; Stuart A Weinzimer; Rajiv V Shah; John J Mastrototaro
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

2.  Use of subcutaneous interstitial fluid glucose to estimate blood glucose: revisiting delay and sensor offset.

Authors:  Kerstin Rebrin; Norman F Sheppard; Garry M Steil
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

3.  Extensive Assessment of Blood Glucose Monitoring During Postprandial Period and Its Impact on Closed-Loop Performance.

Authors:  Lyvia Biagi; Arthur Hirata Bertachi; Ignacio Conget; Carmen Quirós; Marga Giménez; F Javier Ampudia-Blasco; Paolo Rossetti; Jorge Bondia; Josep Vehí
Journal:  J Diabetes Sci Technol       Date:  2017-06-21

4.  Accuracy of a new real-time continuous glucose monitoring algorithm.

Authors:  D Barry Keenan; Raymond Cartaya; John J Mastrototaro
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

5.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

6.  Accuracy and reliability of continuous glucose monitoring systems: a head-to-head comparison.

Authors:  Yoeri M Luijf; Julia K Mader; Werner Doll; Thomas Pieber; Anne Farret; Jerome Place; Eric Renard; Daniela Bruttomesso; Alessio Filippi; Angelo Avogaro; Sabine Arnolds; Carsten Benesch; Lutz Heinemann; J Hans DeVries
Journal:  Diabetes Technol Ther       Date:  2013-05-07       Impact factor: 6.118

7.  Evaluation of a novel continuous glucose measurement device in patients with diabetes mellitus across the glycemic range.

Authors:  Linda Morrow; Marcus Hompesch; Ann M Tideman; Jennifer Matson; Nancy Dunne; Scott Pardo; Joan L Parkes; Holly C Schachner; David A Simmons
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

Review 8.  Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices.

Authors:  Guido Freckmann; Stefan Pleus; Mike Grady; Steven Setford; Brian Levy
Journal:  J Diabetes Sci Technol       Date:  2018-11-19

9.  Redundancy in Glucose Sensing: Enhanced Accuracy and Reliability of an Electrochemical Redundant Sensor for Continuous Glucose Monitoring.

Authors:  Amin Sharifi; Andrea Varsavsky; Johanna Ulloa; Jodie C Horsburgh; Sybil A McAuley; Balasubramanian Krishnamurthy; Alicia J Jenkins; Peter G Colman; Glenn M Ward; Richard J MacIsaac; Rajiv Shah; David N O'Neal
Journal:  J Diabetes Sci Technol       Date:  2016-05-03

10.  Continuous glucose monitoring accuracy results vary between assessment at home and assessment at the clinical research center.

Authors:  Yoeri M Luijf; Angelo Avogaro; Carsten Benesch; Daniela Bruttomesso; Claudio Cobelli; Martin Ellmerer; Lutz Heinemann; Julia K Mader; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2012-09-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.