Accuracy of a new real-time continuous glucose monitoring algorithm.

D Barry Keenan1, Raymond Cartaya, John J Mastrototaro.   

Abstract

BACKGROUND: Through minimally invasive sensor-based continuous glucose monitoring (CGM), individuals can manage their blood glucose (BG) levels more aggressively, thereby improving their hemoglobin A1c level, while reducing the risk of hypoglycemia. Tighter glycemic control through CGM, however, requires an accurate glucose sensor and calibration algorithm with increased performance at lower BG levels.
METHODS: Sensor and BG measurements for 72 adult and adolescent subjects were obtained during the course of a 26-week multicenter study evaluating the efficacy of the Paradigm REAL-Time (PRT) sensor-augmented pump system (Medtronic Diabetes, Northridge, CA) in an outpatient setting. Subjects in the study arm performed at least four daily finger stick measurements. A retrospective analysis of the data set was performed to evaluate a new calibration algorithm utilized in the Paradigm Veo insulin pump (Medtronic Diabetes) and to compare these results to performance metrics calculated for the PRT.
RESULTS: A total of N = 7193 PRT sensor downloads for 3 days of use, as well as 90,472 temporally and nonuniformly paired data points (sensor and meter values), were evaluated, with 5841 hypoglycemic and 15,851 hyperglycemic events detected through finger stick measurements. The Veo calibration algorithm decreased the overall mean absolute relative difference by greater than 0.25 to 15.89%, with hypoglycemia sensitivity increased from 54.9% in the PRT to 82.3% in the Veo (90.5% with predictive alerts); however, hyperglycemia sensitivity was decreased only marginally from 86% in the PRT to 81.7% in the Veo.
CONCLUSIONS: The Veo calibration algorithm, with sensor error reduced significantly in the 40- to 120-mg/dl range, improves hypoglycemia detection, while retaining accuracy at high glucose levels. 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20167174      PMCID: PMC2825631          DOI: 10.1177/193229681000400114

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  19 in total

1.  A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.

Authors:  J L Parkes; S L Slatin; S Pardo; B H Ginsberg
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

2.  Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring.

Authors:  Dorothee Deiss; Jan Bolinder; Jean-Pierre Riveline; Tadej Battelino; Emanuele Bosi; Nadia Tubiana-Rufi; David Kerr; Moshe Phillip
Journal:  Diabetes Care       Date:  2006-12       Impact factor: 19.112

3.  Evaluating clinical accuracy of systems for self-monitoring of blood glucose.

Authors:  W L Clarke; D Cox; L A Gonder-Frederick; W Carter; S L Pohl
Journal:  Diabetes Care       Date:  1987 Sep-Oct       Impact factor: 19.112

4.  Performance evaluation of the MiniMed continuous glucose monitoring system during patient home use.

Authors:  T M Gross; B W Bode; D Einhorn; D M Kayne; J H Reed; N H White; J J Mastrototaro
Journal:  Diabetes Technol Ther       Date:  2000       Impact factor: 6.118

5.  Quality of diabetes care in U.S. academic medical centers: low rates of medical regimen change.

Authors:  Richard W Grant; John B Buse; James B Meigs
Journal:  Diabetes Care       Date:  2005-02       Impact factor: 19.112

6.  Impaired insulin action in puberty. A contributing factor to poor glycemic control in adolescents with diabetes.

Authors:  S A Amiel; R S Sherwin; D C Simonson; A A Lauritano; W V Tamborlane
Journal:  N Engl J Med       Date:  1986-07-24       Impact factor: 91.245

7.  Implementation of treatment protocols in the Diabetes Control and Complications Trial.

Authors: 
Journal:  Diabetes Care       Date:  1995-03       Impact factor: 19.112

8.  The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus.

Authors:  D M Nathan; S Genuth; J Lachin; P Cleary; O Crofford; M Davis; L Rand; C Siebert
Journal:  N Engl J Med       Date:  1993-09-30       Impact factor: 91.245

9.  Sustained effect of intensive treatment of type 1 diabetes mellitus on development and progression of diabetic nephropathy: the Epidemiology of Diabetes Interventions and Complications (EDIC) study.

Authors: 
Journal:  JAMA       Date:  2003-10-22       Impact factor: 56.272

10.  Glycemic control in early IDDM. The Wisconsin Diabetes Registry.

Authors:  C Allen; D J Zaccaro; M Palta; R Klein; S C Duck; D J D'Alessio
Journal:  Diabetes Care       Date:  1992-08       Impact factor: 19.112

View more
  30 in total

1.  Continuous glucose monitoring considerations for the development of a closed-loop artificial pancreas system.

Authors:  D Barry Keenan; Benyamin Grosman; Harry W Clark; Anirban Roy; Stuart A Weinzimer; Rajiv V Shah; John J Mastrototaro
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

Review 2.  Toward closing the loop: an update on insulin pumps and continuous glucose monitoring systems.

Authors:  Tandy Aye; Jen Block; Bruce Buckingham
Journal:  Endocrinol Metab Clin North Am       Date:  2010-09       Impact factor: 4.741

3.  Evaluation of stochastic adjustment for glucose sensor bias during closed-loop insulin delivery.

Authors:  Craig Kollman; Peter Calhoun; John Lum; Werner Sauer; Roy W Beck
Journal:  Diabetes Technol Ther       Date:  2013-11-15       Impact factor: 6.118

4.  Professional continuous glucose monitoring in subjects with type 1 diabetes: retrospective hypoglycemia detection.

Authors:  Morten Hasselstrøm Jensen; Toke Folke Christensen; Lise Tarnow; Zeinab Mahmoudi; Mette Dencker Johansen; Ole Kristian Hejlesen
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

5.  Autoregressive Modeling of Drift and Random Error to Characterize a Continuous Intravascular Glucose Monitoring Sensor.

Authors:  Tony Zhou; Jennifer L Dickson; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2017-07-14

Review 6.  Hypo- and Hyperglycemic Alarms: Devices and Algorithms.

Authors:  Daniel Howsmon; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2015-04-30

7.  Factory-Calibrated Continuous Glucose Sensors: The Science Behind the Technology.

Authors:  Udo Hoss; Erwin Satrya Budiman
Journal:  Diabetes Technol Ther       Date:  2017-05       Impact factor: 6.118

8.  The accuracy and efficacy of real-time continuous glucose monitoring sensor in Chinese diabetes patients: a multicenter study.

Authors:  Jian Zhou; Xiaofeng Lv; Yiming Mu; Xianling Wang; Jing Li; Xingguang Zhang; Jinxiao Wu; Yuqian Bao; Weiping Jia
Journal:  Diabetes Technol Ther       Date:  2012-08       Impact factor: 6.118

9.  The effect of congestive heart failure on sensor accuracy among hospitalized patients with type 2 diabetes.

Authors:  Kathleen Dungan; Kari Graessle; Colleen Sagrilla
Journal:  Diabetes Technol Ther       Date:  2013-10       Impact factor: 6.118

10.  Characterizing accuracy and precision of glucose sensors and meters.

Authors:  David Rodbard
Journal:  J Diabetes Sci Technol       Date:  2014-07-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.