OBJECTIVES: The aim was to investigate the possibility of evaluating the modulation transfer function (MTF) of cone beam CT (CBCT) for dental use using the oversampling method. METHODS: The CBCT apparatus (3D Accuitomo) with an image intensifier was used with a 100 mum tungsten wire placed inside the scanner at a slight angle to the plane perpendicular to the plane of interest and scanned. 200 contiguous reconstructed images were used to obtain the oversampling line-spread function (LSF). The MTF curve was obtained by computing the Fourier transformation from the oversampled LSF. Line pair tests were also performed using Catphan(R). RESULTS: The oversampling method provided smooth and reproducible MTF curves. The MTF curves revealed that the spatial resolution in the z-axis direction was significantly higher than that in the axial direction. This result was also confirmed by the line pair test. CONCLUSIONS: MTF analysis was performed successfully using the oversampling method. In addition, this study clarified that the 3D Accuitomo had high spatial resolution, especially in the z-axis direction.
OBJECTIVES: The aim was to investigate the possibility of evaluating the modulation transfer function (MTF) of cone beam CT (CBCT) for dental use using the oversampling method. METHODS: The CBCT apparatus (3D Accuitomo) with an image intensifier was used with a 100 mum tungsten wire placed inside the scanner at a slight angle to the plane perpendicular to the plane of interest and scanned. 200 contiguous reconstructed images were used to obtain the oversampling line-spread function (LSF). The MTF curve was obtained by computing the Fourier transformation from the oversampled LSF. Line pair tests were also performed using Catphan(R). RESULTS: The oversampling method provided smooth and reproducible MTF curves. The MTF curves revealed that the spatial resolution in the z-axis direction was significantly higher than that in the axial direction. This result was also confirmed by the line pair test. CONCLUSIONS: MTF analysis was performed successfully using the oversampling method. In addition, this study clarified that the 3D Accuitomo had high spatial resolution, especially in the z-axis direction.
Authors: Miet Loubele; Maria Eugenia Guerrero; Reinhide Jacobs; Paul Suetens; Daniel van Steenberghe Journal: Int J Oral Maxillofac Implants Date: 2007 May-Jun Impact factor: 2.804
Authors: H F-W Sadrozinski; T Geoghegan; E Harvey; R P Johnson; T E Plautz; A Zatserklyaniy; V Bashkirov; R F Hurley; P Piersimoni; R W Schulte; P Karbasi; K E Schubert; B Schultze; V Giacometti Journal: Nucl Instrum Methods Phys Res A Date: 2016-02-07 Impact factor: 1.455
Authors: R Pauwels; L Seynaeve; J C G Henriques; C de Oliveira-Santos; P C Souza; F H Westphalen; I R F Rubira-Bullen; R F Ribeiro-Rotta; M I B Rockenbach; F Haiter-Neto; P Pittayapat; H Bosmans; R Bogaerts; R Jacobs Journal: Dentomaxillofac Radiol Date: 2015-06-19 Impact factor: 2.419
Authors: Muhammad U Ghani; Zhongxing Zhou; Liqiang Ren; Yuhua Li; Bin Zheng; Kai Yang; Hong Liu Journal: Nucl Instrum Methods Phys Res A Date: 2016-01-21 Impact factor: 1.455
Authors: Niral M Sheth; Wojciech Zbijewski; Matthew W Jacobson; Godwin Abiola; Gerhard Kleinszig; Sebastian Vogt; Stefan Soellradl; Jens Bialkowski; William S Anderson; Clifford R Weiss; Greg M Osgood; Jeffrey H Siewerdsen Journal: Med Phys Date: 2018-11-13 Impact factor: 4.071