Spatial resolution measurements by Radia diagnostic software with SEDENTEXCT image quality phantom in cone beam CT for dental use.

Hiroshi Watanabe1, Yoshikazu Nomura1, Ami Kuribayashi1, Tohru Kurabayashi1.   

Abstract

OBJECTIVES: We aimed to employ the Radia diagnostic software with the safety and efficacy of a new emerging dental X-ray modality (SEDENTEXCT) image quality (IQ) phantom in CT, and to evaluate its validity.
METHODS: The SEDENTEXCT IQ phantom and Radia diagnostic software were employed. The phantom was scanned using one medical full-body CT and two dentomaxillofacial cone beam CTs. The obtained images were imported to the Radia software, and the spatial resolution outputs were evaluated. The oversampling method was employed using our original wire phantom as a reference. The resultant modulation transfer function (MTF) curves were compared. The null hypothesis was that MTF curves generated using both methods would be in agreement. One-way analysis of variance tests were applied to the f50 and f10 values from the MTF curves. The f10 values were subjectively confirmed by observing the line pair modules.
RESULTS: The Radia software reported the MTF curves on the xy-plane of the CT scans, but could not return f50 and f10 values on the z-axis. The null hypothesis concerning the reported MTF curves on the xy-plane was rejected. There were significant differences between the results of the Radia software and our reference method, except for f10 values in CS9300. These findings were consistent with our line pair observations.
CONCLUSIONS: We evaluated the validity of the Radia software with the SEDENTEXCT IQ phantom. The data provided were semi-automatic, albeit with problems and statistically different from our reference. We hope the manufacturer will overcome these limitations.

Mesh:

Year:  2017        PMID: 29161903      PMCID: PMC6047630          DOI: 10.1259/dmfr.20170307

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  8 in total

Review 1.  Dental cone beam CT: A review.

Authors:  Timo Kiljunen; Touko Kaasalainen; Anni Suomalainen; Mika Kortesniemi
Journal:  Phys Med       Date:  2015-10-23       Impact factor: 2.685

2.  Evaluation of the spatial resolution characteristics of a cone-beam breast CT scanner.

Authors:  Alexander L C Kwan; John M Boone; Kai Yang; Shih-Ying Huang
Journal:  Med Phys       Date:  2007-01       Impact factor: 4.071

3.  Location dependency of the spatial resolution of cone beam computed tomography for dental use.

Authors:  Yoshihiro Ozaki; Hiroshi Watanabe; Yoshikazu Nomura; Eiichi Honda; Yasunori Sumi; Tohru Kurabayashi
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2013-11

4.  Modulation transfer function evaluation of cone beam computed tomography for dental use with the oversampling method.

Authors:  H Watanabe; E Honda; T Kurabayashi
Journal:  Dentomaxillofac Radiol       Date:  2010-01       Impact factor: 2.419

5.  A comparative study for spatial resolution and subjective image characteristics of a multi-slice CT and a cone-beam CT for dental use.

Authors:  Hiroshi Watanabe; Eiichi Honda; Akemi Tetsumura; Tohru Kurabayashi
Journal:  Eur J Radiol       Date:  2009-10-09       Impact factor: 3.528

Review 6.  Quality assurance phantoms for cone beam computed tomography: a systematic literature review.

Authors:  Marcus V L de Oliveira; Ann Wenzel; Paulo S F Campos; Rubens Spin-Neto
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

Review 7.  Spatial resolution in CBCT machines for dental/maxillofacial applications-what do we know today?

Authors:  D Brüllmann; R K W Schulze
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

8.  Development and applicability of a quality control phantom for dental cone-beam CT.

Authors:  Ruben Pauwels; Harry Stamatakis; Giorgos Manousaridis; Adrian Walker; Koen Michielsen; Hilde Bosmans; Ria Bogaerts; Reinhilde Jacobs; Keith Horner; Kostas Tsiklakis
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

  8 in total
  2 in total

1.  [Design and optimization of a cone-beam CT system for extremity imaging].

Authors:  Kun Ma; Mingqiang Li; Xi Tao; Dong Zeng; Yongbo Wang; Zhaoying Bian; Ziquan Wei; Gaofeng Chen; Qianjin Feng; Jianhua Ma; Jing Huang
Journal:  Nan Fang Yi Ke Da Xue Xue Bao       Date:  2018-11-30

2.  Optimization of exposure parameters and relationship between subjective and technical image quality in cone-beam computed tomography.

Authors:  Ha-Na Park; Chang-Ki Min; Kyoung-A Kim; Kwang-Joon Koh
Journal:  Imaging Sci Dent       Date:  2019-06-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.