Literature DB >> 20012921

Dose reduction in multislice CT by means of bismuth shields: results of in vivo measurements and computed evaluation.

P Catuzzo1, S Aimonetto, G Fanelli, P Marchisio, T Meloni, L Mistretta, M Pasquino, E Richetta, S Tofani.   

Abstract

PURPOSE: We investigated the amount of patient dose reduction in the thyroid, lens of the eye and the breast when using bismuth protections in multislice computed tomography (CT) exams as well as their influence on the quality of diagnostic images.
MATERIALS AND METHODS: The radiation dose was measured by using thermoluminescence dosimeters. The study was conducted on the two CT scanners installed in our radiology department (64 and eight slices). The shield effects on the CT image were evaluated by measuring the signal-to-noise ratio in a phantom and in vivo, and by verifying the presence of artefacts on patients' images. The obtained organ-dose reduction factors were used to evaluate the effects of shielding on the effective dose.
RESULTS: The shielding attenuation ranged from 30% to 60% depending on the CT scan protocols and organs. The difference between shielded and unshielded signal-to-noise ratio was statistically significant but within the standard requirements for quality assurance. Results were in agreement with the radiologists' perception of image quality. The use of the shields allowed up to 38% reduction of effective dose.
CONCLUSIONS: Use of bismuth shields significantly decreases both organ and effective radiation dose, with a consequent reduction in health risk for the patient, quantified in 1.4 fewer cases of radiation-induced tumours every 5 years in our centre (12,100 exams/year), in agreement with the risk factors proposed by Publication 60 of the International Commission on Radiological Protection (ICRP). The relative inexpensiveness of these protections, their easy application and their substantial lack of influence on image quality suggest their massive introduction into routine clinical practice.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20012921     DOI: 10.1007/s11547-009-0469-4

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  9 in total

1.  Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the users' needs?

Authors:  D J McLaughlin; R B Mooney
Journal:  Clin Radiol       Date:  2004-05       Impact factor: 2.350

2.  Patient dose reduction in CT.

Authors:  M A Lewis; S Edyvean
Journal:  Br J Radiol       Date:  2005-10       Impact factor: 3.039

Review 3.  The evolution and state-of-the-art principles of multislice computed tomography.

Authors:  Gerhard Kohl
Journal:  Proc Am Thorac Soc       Date:  2005

4.  Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality.

Authors:  J Geleijns; M Salvadó Artells; W J H Veldkamp; M López Tortosa; A Calzado Cantera
Journal:  Eur Radiol       Date:  2006-04-08       Impact factor: 5.315

Review 5.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

6.  Computed tomography and radiation exposure.

Authors:  Maurice Tubiana
Journal:  N Engl J Med       Date:  2008-02-21       Impact factor: 91.245

7.  The breast: in-plane x-ray protection during diagnostic thoracic CT--shielding with bismuth radioprotective garments.

Authors:  K D Hopper; S H King; M E Lobell; T R TenHave; J S Weaver
Journal:  Radiology       Date:  1997-12       Impact factor: 11.105

8.  Radioprotection to the eye during CT scanning.

Authors:  K D Hopper; J D Neuman; S H King; A R Kunselman
Journal:  AJNR Am J Neuroradiol       Date:  2001 Jun-Jul       Impact factor: 3.825

Review 9.  Strategies for CT radiation dose optimization.

Authors:  Mannudeep K Kalra; Michael M Maher; Thomas L Toth; Leena M Hamberg; Michael A Blake; Jo-Anne Shepard; Sanjay Saini
Journal:  Radiology       Date:  2004-01-22       Impact factor: 11.105

  9 in total
  12 in total

1.  Dose reduction of cone beam CT scanning for the entire oral and maxillofacial regions with thyroid collars.

Authors:  X M Qu; G Li; G C H Sanderink; Z Y Zhang; X C Ma
Journal:  Dentomaxillofac Radiol       Date:  2012-07       Impact factor: 2.419

2.  Effect of bismuth breast shielding on radiation dose and image quality in coronary CT angiography.

Authors:  Andrew J Einstein; Carl D Elliston; Daniel W Groves; Bin Cheng; Steven D Wolff; Gregory D N Pearson; M Robert Peters; Lynne L Johnson; Sabahat Bokhari; Gary W Johnson; Ketan Bhatia; Theodore Pozniakoff; David J Brenner
Journal:  J Nucl Cardiol       Date:  2011-11-09       Impact factor: 5.952

3.  Comparison of diagnostic performance between single- and multiphasic contrast-enhanced abdominopelvic computed tomography in patients admitted to the emergency department with abdominal pain: potential radiation dose reduction.

Authors:  Shin Hye Hwang; Je Sung You; Mi Kyong Song; Jin-Young Choi; Myeong-Jin Kim; Yong Eun Chung
Journal:  Eur Radiol       Date:  2014-11-26       Impact factor: 5.315

Review 4.  Quality assurance of imaging techniques used in the clinical management of osteoporosis.

Authors:  G Guglielmi; J Damilakis; G Solomou; A Bazzocchi
Journal:  Radiol Med       Date:  2012-10-22       Impact factor: 3.469

5.  [Current strategies for dosage reduction in computed tomography].

Authors:  M S May; W Wuest; M M Lell; M Uder; W A Kalender; B Schmidt
Journal:  Radiologe       Date:  2012-10       Impact factor: 0.635

6.  Usefulness of a lead shielding device for reducing the radiation dose to tissues outside the primary beams during CT.

Authors:  Jae-Joon Chung; Eun-Suk Cho; Sung Min Kang; Jeong-Sik Yu; Dae Jung Kim; Joo Hee Kim
Journal:  Radiol Med       Date:  2014-06-06       Impact factor: 3.469

7.  Shielding effect of radiation dose reduction fiber during the use of C-arm fluoroscopy: a phantom study.

Authors:  Hyemi Cha; Kisung Lee; Moon Seok Park; Kyoung Min Lee; Kyeyoung Cho; Ki Hyuk Sung
Journal:  J Radiat Res       Date:  2020-09-08       Impact factor: 2.724

8.  Optimised low-dose multidetector CT protocol for children with cranial deformity.

Authors:  Jose Luis Vazquez; Miguel Angel Pombar; Jose Manuel Pumar; Victor Miguel del Campo
Journal:  Eur Radiol       Date:  2013-03-16       Impact factor: 5.315

9.  Lens Dose Reduction by Patient Posture Modification During Neck CT.

Authors:  Elizabeth G Mosher; John A Butman; Les R Folio; Nadia M Biassou; Choonsik Lee
Journal:  AJR Am J Roentgenol       Date:  2018-03-16       Impact factor: 3.959

10.  Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans.

Authors:  Rafel Tappouni; Bradley Mathers
Journal:  ISRN Radiol       Date:  2012-11-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.