Effect of bismuth breast shielding on radiation dose and image quality in coronary CT angiography.

Andrew J Einstein1, Carl D Elliston, Daniel W Groves, Bin Cheng, Steven D Wolff, Gregory D N Pearson, M Robert Peters, Lynne L Johnson, Sabahat Bokhari, Gary W Johnson, Ketan Bhatia, Theodore Pozniakoff, David J Brenner.   

Abstract

BACKGROUND: Coronary computed tomographic angiography (CCTA) is associated with high radiation dose to the female breasts. Bismuth breast shielding offers the potential to significantly reduce dose to the breasts and nearby organs, but the magnitude of this reduction and its impact on image quality and radiation dose have not been evaluated.
METHODS: Radiation doses from CCTA to critical organs were determined using metal-oxide-semiconductor field-effect transistors positioned in a customized anthropomorphic whole-body dosimetry verification phantom. Image noise and signal were measured in regions of interest (ROIs) including the coronary arteries.
RESULTS: With bismuth shielding, breast radiation dose was reduced 46%-57% depending on breast size and scanning technique, with more moderate dose reduction to the heart, lungs, and esophagus. However, shielding significantly decreased image signal (by 14.6 HU) and contrast (by 28.4 HU), modestly but significantly increased image noise in ROIs in locations of coronary arteries, and decreased contrast-to-noise ratio by 20.9%.
CONCLUSIONS: While bismuth breast shielding can significantly decrease radiation dose to critical organs, it is associated with an increase in image noise, decrease in contrast-to-noise, and changes tissue attenuation characteristics in the location of the coronary arteries.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22068687      PMCID: PMC3266996          DOI: 10.1007/s12350-011-9473-x

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  31 in total

1.  Female breast radiation exposure during thorax multidetector computed tomography and the effectiveness of bismuth breast shield to reduce breast radiation dose.

Authors:  Mehmet Halit Yilmaz; Sait Albayram; Doğan Yaşar; Harun Ozer; Ibrahim Adaletli; Doğan Selçuk; Canan Akman; Ayça Altuğ
Journal:  J Comput Assist Tomogr       Date:  2007 Jan-Feb       Impact factor: 1.826

2.  Pediatric chest MDCT using tube current modulation: effect on radiation dose with breast shielding.

Authors:  Courtney Coursey; Donald P Frush; Terry Yoshizumi; Greta Toncheva; Giao Nguyen; S Bruce Greenberg
Journal:  AJR Am J Roentgenol       Date:  2008-01       Impact factor: 3.959

3.  Slice wars vs dose wars in multiple-row detector CT.

Authors:  Mahadevappa Mahesh; James M Hevezi
Journal:  J Am Coll Radiol       Date:  2009-03       Impact factor: 5.532

4.  Coronary calcium scoring with MDCT: the radiation dose to the breast and the effectiveness of bismuth breast shield.

Authors:  Mehmet Halit Yilmaz; Doğan Yaşar; Sait Albayram; Ibrahim Adaletli; Harun Ozer; Mustafa Ozbayrak; Ismail Mihmanli; Canan Akman
Journal:  Eur J Radiol       Date:  2006-09-07       Impact factor: 3.528

5.  In-plane shielding for CT: effect of off-centering, automatic exposure control and shield-to-surface distance.

Authors:  Mannudeep K Kalra; Pragya Dang; Sarabjeet Singh; Sanjay Saini; Jo-Anne O Shepard
Journal:  Korean J Radiol       Date:  2009-03-03       Impact factor: 3.500

6.  Radiation dose savings for adult pulmonary embolus 64-MDCT using bismuth breast shields, lower peak kilovoltage, and automatic tube current modulation.

Authors:  Lynne M Hurwitz; Terry T Yoshizumi; Philip C Goodman; Rendon C Nelson; Greta Toncheva; Giao B Nguyen; Carolyn Lowry; Colin Anderson-Evans
Journal:  AJR Am J Roentgenol       Date:  2009-01       Impact factor: 3.959

7.  Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study.

Authors:  W Bob Meijboom; Matthijs F L Meijs; Joanne D Schuijf; Maarten J Cramer; Nico R Mollet; Carlos A G van Mieghem; Koen Nieman; Jacob M van Werkhoven; Gabija Pundziute; Annick C Weustink; Alexander M de Vos; Francesca Pugliese; Benno Rensing; J Wouter Jukema; Jeroen J Bax; Mathias Prokop; Pieter A Doevendans; Myriam G M Hunink; Gabriel P Krestin; Pim J de Feyter
Journal:  J Am Coll Cardiol       Date:  2008-12-16       Impact factor: 24.094

8.  Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations.

Authors:  Sabrina V Vollmar; Willi A Kalender
Journal:  Eur Radiol       Date:  2008-04-15       Impact factor: 5.315

9.  Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography.

Authors:  Andrew J Einstein; Milena J Henzlova; Sanjay Rajagopalan
Journal:  JAMA       Date:  2007-07-18       Impact factor: 56.272

10.  Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial.

Authors:  Matthew J Budoff; David Dowe; James G Jollis; Michael Gitter; John Sutherland; Edward Halamert; Markus Scherer; Raye Bellinger; Arthur Martin; Robert Benton; Augustin Delago; James K Min
Journal:  J Am Coll Cardiol       Date:  2008-11-18       Impact factor: 24.094

View more
  7 in total

1.  Prevalence of Protective Shielding Utilization for Radiation Dose Reduction in Adult Patients Undergoing Body Scanning Using Computed Tomography.

Authors:  Shoaib Safiullah; Roshan Patel; Brittany Uribe; Kyle Spradling; Chandana Lall; Lishi Zhang; Zhamshid Okhunov; Ralph V Clayman; Jaime Landman
Journal:  J Endourol       Date:  2017-08-14       Impact factor: 2.942

2.  Scattered Dose Calculations and Measurements in a Life-Like Mouse Phantom.

Authors:  David Welch; Leah Turner; Michael Speiser; Gerhard Randers-Pehrson; David J Brenner
Journal:  Radiat Res       Date:  2017-01-31       Impact factor: 2.841

3.  Sample size requirements for estimating effective dose from computed tomography using solid-state metal-oxide-semiconductor field-effect transistor dosimetry.

Authors:  Sigal Trattner; Bin Cheng; Radoslaw L Pieniazek; Udo Hoffmann; Pamela S Douglas; Andrew J Einstein
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

Review 4.  Recent advances in cardiac computed tomography dose reduction strategies: a review of scientific evidence and technical developments.

Authors:  Sandeep S Hedgire; Vinit Baliyan; Brian B Ghoshhajra; Mannudeep K Kalra
Journal:  J Med Imaging (Bellingham)       Date:  2017-08-24

5.  Construction of mouse phantoms from segmented CT scan data for radiation dosimetry studies.

Authors:  D Welch; A D Harken; G Randers-Pehrson; D J Brenner
Journal:  Phys Med Biol       Date:  2015-04-10       Impact factor: 3.609

6.  Bismuth-Silicon and Bismuth-Polyurethane Composite Shields for Breast Protection in Chest Computed Tomography Examinations.

Authors:  Parinaz Mehnati; Mehran Arash; Parisa Akhlaghi
Journal:  J Med Phys       Date:  2018 Jan-Mar

7.  Bismuth shield affecting CT image quality and radiation dose in adjacent and distant zones relative to shielding surface: A phantom study.

Authors:  Ying-Lan Liao; Nan-Ku Lai; Yeu-Sheng Tyan; Hui-Yu Tsai
Journal:  Biomed J       Date:  2019-11-04       Impact factor: 4.910

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.