Photon counting multienergy x-ray imaging: effect of the characteristic x rays on detector performance.

Polad M Shikhaliev1, Shannon G Fritz, John W Chapman.   

Abstract

PURPOSE: The purpose of this work was to investigate the effect of characteristic x rays on the performance of photon counting detectors for multienergy x-ray imaging. X-ray and CT systems with photon counting detectors have compelling advantages compared to energy integrating detectors, and cadmium zinc telluride (CZT) detector is the detector of choice. However, current CZT detectors exhibit several limitations that hamper their practical applications. These limitations include hole trapping, high leakage current, and charge sharing between detector pixels. Charge sharing occurs due to the diffusion of charge when it drifts toward the pixel electrodes. It also occurs due to nonlocal reabsorption of characteristic and scattered x rays created in the detector volume. Hole trapping, leakage current, and charge diffusion may potentially have technical solutions. Characteristic x-ray escape and scatter, however, are fundamental in nature and cannot be easily addressed. The x-ray scatter in the CZT material is small at photon energies used in x-ray imaging. Therefore, the remaining major factor is characteristic x ray.
METHODS: Monte Carlo simulations were used for this study. An experimental photon counting multienergy x-ray imaging system was used to compare simulations to experimental results. An x-ray spectrum at 120 kVp tube voltage was used. The x-ray energy range was split into five subregions (energy bins) and Monte Carlo simulations were performed at average x-ray energies corresponding to these energy bins. The detector pixel size was changed within the 0.1-1 mm range, which covered all possible applications including radiography and CT imaging. The pixel shapes included square and strip pixels. For strip pixels, tilted angle irradiation of the CZT detector was also investigated.
RESULTS: The characteristic x rays escaped the pixels in approximately 70% of all x-ray interactions for the smallest pixel size of 0.1 mm. The escape fraction decreased to 20% for the largest pixel size of 1 mm. All escape fractions, for all pixel sizes, at five energies, for square and strip pixels, and at three tilt angles were calculated and presented in tables. Simulated and measured spectra at 120 kVp were compared.
CONCLUSIONS: Characteristic x-ray escape deteriorates energy and spatial resolution, particularly for small pixel sizes. Correction methods should be developed based on the results of the simulations and experimental study.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19994521     DOI: 10.1118/1.3245875

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Photon counting performance of amorphous selenium and its dependence on detector structure.

Authors:  Jann Stavro; Amir H Goldan; Wei Zhao
Journal:  J Med Imaging (Bellingham)       Date:  2018-10-30

Review 2.  Photon-counting Detector CT: System Design and Clinical Applications of an Emerging Technology.

Authors:  Shuai Leng; Michael Bruesewitz; Shengzhen Tao; Kishore Rajendran; Ahmed F Halaweish; Norbert G Campeau; Joel G Fletcher; Cynthia H McCollough
Journal:  Radiographics       Date:  2019 May-Jun       Impact factor: 5.333

3.  Dose-efficient ultrahigh-resolution scan mode using a photon counting detector computed tomography system.

Authors:  Shuai Leng; Zhicong Yu; Ahmed Halaweish; Steffen Kappler; Katharina Hahn; Andre Henning; Zhoubo Li; John Lane; David L Levin; Steven Jorgensen; Erik Ritman; Cynthia McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2016-12-22

4.  Dual-source multienergy CT with triple or quadruple x-ray beams.

Authors:  Lifeng Yu; Liqiang Ren; Zhoubo Li; Shuai Leng; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2018-07-24

5.  Radiation dose efficiency of multi-energy photon-counting-detector CT for dual-contrast imaging.

Authors:  Liqiang Ren; Kishore Rajendran; Cynthia H McCollough; Lifeng Yu
Journal:  Phys Med Biol       Date:  2019-12-13       Impact factor: 3.609

6.  A framework for performance characterization of energy-resolving photon-counting detectors.

Authors:  Mats Persson; Paurakh L Rajbhandary; Norbert J Pelc
Journal:  Med Phys       Date:  2018-10-12       Impact factor: 4.071

7.  Breast composition measurement with a cadmium-zinc-telluride based spectral computed tomography system.

Authors:  Huanjun Ding; Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

8.  A cascaded model of spectral distortions due to spectral response effects and pulse pileup effects in a photon-counting x-ray detector for CT.

Authors:  Jochen Cammin; Jennifer Xu; William C Barber; Jan S Iwanczyk; Neal E Hartsough; Katsuyuki Taguchi
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

9.  Optimization of K-edge imaging for vulnerable plaques using gold nanoparticles and energy resolved photon counting detectors: a simulation study.

Authors:  Yahya Alivov; Pavlo Baturin; Huy Q Le; Justin Ducote; Sabee Molloi
Journal:  Phys Med Biol       Date:  2013-12-13       Impact factor: 3.609

10.  Resolution characterization of a silicon-based, photon-counting computed tomography prototype capable of patient scanning.

Authors:  Joakim da Silva; Fredrik Grönberg; Björn Cederström; Mats Persson; Martin Sjölin; Zlatan Alagic; Robert Bujila; Mats Danielsson
Journal:  J Med Imaging (Bellingham)       Date:  2019-10-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.