Experimental validation of Monte Carlo (MANTIS) simulated x-ray response of columnar CsI scintillator screens.

Melanie Freed1, Stuart Miller, Katherine Tang, Aldo Badano.   

Abstract

PURPOSE: MANTIS is a Monte Carlo code developed for the detailed simulation of columnar CsI scintillator screens in x-ray imaging systems. Validation of this code is needed to provide a reliable and valuable tool for system optimization and accurate reconstructions for a variety of x-ray applications. Whereas previous validation efforts have focused on matching of summary statistics, in this work the authors examine the complete point response function (PRF) of the detector system in addition to relative light output values.
METHODS: Relative light output values and high-resolution PRFs have been experimentally measured with a custom setup. A corresponding set of simulated light output values and PRFs have also been produced, where detailed knowledge of the experimental setup and CsI:Tl screen structures are accounted for in the simulations. Four different screens were investigated with different thicknesses, column tilt angles, and substrate types. A quantitative comparison between the experimental and simulated PRFs was performed for four different incidence angles (0 degrees, 15 degrees, 30 degrees, and 45 degrees) and two different x-ray spectra (40 and 70 kVp). The figure of merit (FOM) used measures the normalized differences between the simulated and experimental data averaged over a region of interest.
RESULTS: Experimental relative light output values ranged from 1.456 to 1.650 and were in approximate agreement for aluminum substrates, but poor agreement for graphite substrates. The FOMs for all screen types, incidence angles, and energies ranged from 0.1929 to 0.4775. To put these FOMs in context, the same FOM was computed for 2D symmetric Gaussians fit to the same experimental data. These FOMs ranged from 0.2068 to 0.8029. Our analysis demonstrates that MANTIS reproduces experimental PRFs with higher accuracy than a symmetric 2D Gaussian fit to the experimental data in the majority of cases. Examination of the spatial distribution of differences between the PRFs shows that the main reason for errors between MANTIS and the experimental data is that MANTIS-generated PRFs are sharper than the experimental PRFs.
CONCLUSIONS: The experimental validation of MANTIS performed in this study demonstrates that MANTIS is able to reliably predict experimental PRFs, especially for thinner screens, and can reproduce the highly asymmetric shape seen in the experimental data. As a result, optimizations and reconstructions carried out using MANTIS should yield results indicative of actual detector performance. Better characterization of screen properties is necessary to reconcile the simulated light output values with experimental data.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19994503      PMCID: PMC2773454          DOI: 10.1118/1.3233683

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  A fast, angle-dependent, analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems.

Authors:  Melanie Freed; Subok Park; Aldo Badano
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  Segmented crystalline scintillators: empirical and theoretical investigation of a high quantum efficiency EPID based on an initial engineering prototype CsI(TI) detector.

Authors:  Amit Sawant; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Yi Wang; Yixin Li; Hong Du; Louis Perna
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

3.  Resolution at oblique incidence angles of a flat panel imager for breast tomosynthesis.

Authors:  James G Mainprize; Aili K Bloomquist; Michael P Kempston; Martin J Yaffe
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

4.  MANTIS: combined x-ray, electron and optical Monte Carlo simulations of indirect radiation imaging systems.

Authors:  Aldo Badano; Josep Sempau
Journal:  Phys Med Biol       Date:  2006-03-01       Impact factor: 3.609

5.  Anisotropic imaging performance in breast tomosynthesis.

Authors:  Aldo Badano; Iacovos S Kyprianou; Robert J Jennings; Josep Sempau
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

6.  Effect of oblique X-ray incidence in flat-panel computed tomography of the breast.

Authors:  Aldo Badano; Iacovos S Kyprianou; Melanie Freed; Robert J Jennings; Josep Sempau
Journal:  IEEE Trans Med Imaging       Date:  2008-12-09       Impact factor: 10.048

7.  Monte Carlo and Lambertian light guide models of the light output from scintillation crystals at megavoltage energies.

Authors:  Philip M Evans; M Amin Mosleh-Shirazi; Emma J Harris; Joao Seco
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

8.  X-ray imaging performance of structured cesium iodide scintillators.

Authors:  Wei Zhao; Goran Ristic; J A Rowlands
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

9.  Modeling scintillator-photodiodes as detectors for megavoltage CT.

Authors:  T T Monajemi; S Steciw; B G Fallone; S Rathee
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

Review 10.  Solid-state, flat-panel, digital radiography detectors and their physical imaging characteristics.

Authors:  A R Cowen; S M Kengyelics; A G Davies
Journal:  Clin Radiol       Date:  2008-01-31       Impact factor: 2.350

View more
  6 in total

1.  Development and characterization of a dynamic lesion phantom for the quantitative evaluation of dynamic contrast-enhanced MRI.

Authors:  Melanie Freed; Jacco A de Zwart; Prasanna Hariharan; Matthew R Myers; Aldo Badano
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

2.  A fast, angle-dependent, analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems.

Authors:  Melanie Freed; Subok Park; Aldo Badano
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

3.  A comparative analysis of OTF, NPS, and DQE in energy integrating and photon counting digital x-ray detectors.

Authors:  Raymond J Acciavatti; Andrew D A Maidment
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

4.  Modeling and evaluation of a high-resolution CMOS detector for cone-beam CT of the extremities.

Authors:  Qian Cao; Alejandro Sisniega; Michael Brehler; J Webster Stayman; John Yorkston; Jeffrey H Siewerdsen; Wojciech Zbijewski
Journal:  Med Phys       Date:  2017-11-27       Impact factor: 4.071

5.  Deriving depth-dependent light escape efficiency and optical Swank factor from measured pulse height spectra of scintillators.

Authors:  Adrian Howansky; Boyu Peng; Anthony R Lubinsky; Wei Zhao
Journal:  Med Phys       Date:  2017-02-13       Impact factor: 4.071

6.  Modelling the transport of optical photons in scintillation detectors for diagnostic and radiotherapy imaging.

Authors:  Emilie Roncali; Mohammad Amin Mosleh-Shirazi; Aldo Badano
Journal:  Phys Med Biol       Date:  2017-10-04       Impact factor: 3.609

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.