Effect of oblique X-ray incidence in flat-panel computed tomography of the breast.

Aldo Badano1, Iacovos S Kyprianou, Melanie Freed, Robert J Jennings, Josep Sempau.   

Abstract

We quantify the variation in resolution due to anisotropy caused by oblique X-ray incidence in indirect flat-panel detectors for computed tomography breast imaging systems. We consider a geometry and detector type utilized in breast computed tomography (CT) systems currently being developed. Our methods rely on mantis, a combined X-ray, electron, and optical Monte Carlo transport open source code. The physics models are the most accurate available in general-purpose Monte Carlo packages in the diagnostic energy range. We consider maximum-obliquity angles of 10 ( degrees ) and 13 ( degrees ) at the centers of the 30 and 40 cm detector edges, respectively, and 16 ( degrees ) at the corner of the detector. Our results indicate that blur is asymmetric and that the resolution properties vary significantly with the angle (or location) of incidence. Our results suggest that the asymmetry can be as high as a factor of 2.6 between orthogonal directions. Anisotropy maps predicted by mantis provide an understanding of the effect that such variations have on the imaging system and allow more accurate modeling and optimization of breast CT systems. These maps of anisotropy across the detector could lead to improved reconstruction and help motivate physics-based strategies for computer detection of breast lesions.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19272986     DOI: 10.1109/TMI.2008.2010443

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  3 in total

1.  A fast, angle-dependent, analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems.

Authors:  Melanie Freed; Subok Park; Aldo Badano
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  A statistical, task-based evaluation method for three-dimensional x-ray breast imaging systems using variable-background phantoms.

Authors:  Subok Park; Robert Jennings; Haimo Liu; Aldo Badano; Kyle Myers
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

3.  Experimental validation of Monte Carlo (MANTIS) simulated x-ray response of columnar CsI scintillator screens.

Authors:  Melanie Freed; Stuart Miller; Katherine Tang; Aldo Badano
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.