Segmented crystalline scintillators: empirical and theoretical investigation of a high quantum efficiency EPID based on an initial engineering prototype CsI(TI) detector.

Amit Sawant1, Larry E Antonuk, Youcef El-Mohri, Qihua Zhao, Yi Wang, Yixin Li, Hong Du, Louis Perna.   

Abstract

Modern-day radiotherapy relies on highly sophisticated forms of image guidance in order to implement increasingly conformal treatment plans and achieve precise dose delivery. One of the most important goals of such image guidance is to delineate the clinical target volume from surrounding normal tissue during patient setup and dose delivery, thereby avoiding dependence on surrogates such as bony landmarks. In order to achieve this goal, it is necessary to integrate highly efficient imaging technology, capable of resolving soft-tissue contrast at very low doses, within the treatment setup. In this paper we report on the development of one such modality, which comprises a nonoptimized, prototype electronic portal imaging device (EPID) based on a 40 mm thick, segmented crystalline CsI(Tl) detector incorporated into an indirect-detection active matrix flat panel imager (AMFPI). The segmented detector consists of a matrix of 160 x 160 optically isolated, crystalline CsI(Tl) elements spaced at 1016 microm pitch. The detector was coupled to an indirect detection-based active matrix array having a pixel pitch of 508 microm, with each detector element registered to 2 x 2 array pixels. The performance of the prototype imager was evaluated under very low-dose radiotherapy conditions and compared to that of a conventional megavoltage AMFPI based on a Lanex Fast-B phosphor screen. Detailed quantitative measurements were performed in order to determine the x-ray sensitivity, modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE). In addition, images of a contrast-detail phantom and an anthropomorphic head phantom were also acquired. The prototype imager exhibited approximately 22 times higher zero-frequency DQE (approximately 22%) compared to that of the conventional AMFPI (approximately 1%). The measured zero-frequency DQE was found to be lower than theoretical upper limits (approximately 27%) calculated from Monte Carlo simulations, which were based solely on the x-ray energy absorbed in the detector-indicating the presence of optical Swank noise. Moreover, due to the nonoptimized nature of this prototype, the spatial resolution was observed to be significantly lower than theoretical expectations. Nevertheless, due to its high quantum efficiency (approximately 55%), the prototype imager exhibited significantly higher DQE than that of the conventional AMFPI across all spatial frequencies. In addition, the frequency-dependent DQE was observed to be relatively invariant with respect to the amount of incident radiation, indicating x-ray quantum limited behavior. Images of the contrast-detail phantom and the head phantom obtained using the prototype system exhibit good visualization of relatively large, low-contrast features, and appear significantly less noisy compared to similar images from a conventional AMFPI. Finally, Monte Carlo-based theoretical calculations indicate that, with proper optimization, further, significant improvements in the DQE performance of such imagers could be achieved. It is strongly anticipated that the realization of optimized versions of such very high-DQE EPIDs would enable megavoltage projection imaging at very low doses, and tomographic imaging from a "beam's eye view" at clinically acceptable doses.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16696482     DOI: 10.1118/1.2178452

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Performance evaluation of polycrystalline HgI2 photoconductors for radiation therapy imaging.

Authors:  Qihua Zhao; Larry E Antonuk; Youcef El-Mohri; Yi Wang; Hong Du; Amit Sawant; Zhong Su; Jin Yamamoto
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  Monte Carlo investigations of megavoltage cone-beam CT using thick, segmented scintillating detectors for soft tissue visualization.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Amit Sawant; Hong Du
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

3.  High-DQE EPIDs based on thick, segmented BGO and CsI:Tl scintillators: performance evaluation at extremely low dose.

Authors:  Yi Wang; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Louis Perna
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

4.  A Monte Carlo investigation of Swank noise for thick, segmented, crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

5.  Experimental validation of Monte Carlo (MANTIS) simulated x-ray response of columnar CsI scintillator screens.

Authors:  Melanie Freed; Stuart Miller; Katherine Tang; Aldo Badano
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

6.  Metal artifact correction for x-ray computed tomography using kV and selective MV imaging.

Authors:  Meng Wu; Andreas Keil; Dragos Constantin; Josh Star-Lack; Lei Zhu; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

7.  Monte Carlo investigations of the effect of beam divergence on thick, segmented crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Youcef El-Mohri; Larry E Antonuk; Qihua Zhao
Journal:  Phys Med Biol       Date:  2010-06-04       Impact factor: 3.609

8.  Optimization of the design of thick, segmented scintillators for megavoltage cone-beam CT using a novel, hybrid modeling technique.

Authors:  Langechuan Liu; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Hao Jiang
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

9.  Optimization of the performance of segmented scintillators for radiotherapy imaging through novel binning techniques.

Authors:  Youcef El-Mohri; Larry E Antonuk; Richard B Choroszucha; Qihua Zhao; Hao Jiang; Langechuan Liu
Journal:  Phys Med Biol       Date:  2014-02-03       Impact factor: 3.609

10.  A piecewise-focused high DQE detector for MV imaging.

Authors:  Josh Star-Lack; Daniel Shedlock; Dennis Swahn; Dave Humber; Adam Wang; Hayley Hirsh; George Zentai; Daren Sawkey; Isaac Kruger; Mingshan Sun; Eric Abel; Gary Virshup; Mihye Shin; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.