| Literature DB >> 26971327 |
Kiyokazu Fukui1, Ayumi Kaneuji2, Tadami Matsumoto2, Kazuhiro Shintani3.
Abstract
There is a rising concern about squeaking in ceramic-on-ceramic total hip arthroplasty (THA). In pin-on-disc testing of a delta-delta coupling, we reproduced squeaking and observed microcracks on worn surfaces. We used a pin-on-disc machine and made discs and pins by cutting delta ceramic to a diameter of 40 mm (D-D). Cross-linked polyethylene was used for a comparison disc (D-P). We performed the same test using another D-D coupling specimen to confirm reproducibility. Squeaking in the D-D specimen was reproduced in wet conditions, though it was not found in the D-P specimen. Fast Fourier transform analysis showed a peak frequency for squeaking of 2794 Hz. The noise occurred at about 6.6 km of sliding distance. Scanning electron microscopy revealed that the worn surface of the pin in D-D at 10.8 km of sliding distance had some microcracks. However, there was no obvious damage to the worn surface of the pin in D-P at the same sliding distance. We confirmed the reproducibility of these findings, obtaining similar results, including squeaking, from another D-D coupling specimen. Our findings show that squeaking may occur in THA using delta ceramic bearings even if implants are placed to avoid extra-articular impingement of the femoral neck. Although the clinical relevance of microcracks is unknown, they may affect long-term outcomes in THA using delta ceramic bearings.Entities:
Keywords: Delta ceramics; Microcracks; Squeaking; Total hip arthroplasty
Mesh:
Year: 2016 PMID: 26971327 DOI: 10.1007/s00590-016-1748-0
Source DB: PubMed Journal: Eur J Orthop Surg Traumatol ISSN: 1633-8065