Literature DB >> 26971327

Squeaking and microcracks in a delta-delta ceramic coupling: pin-on-disc study.

Kiyokazu Fukui1, Ayumi Kaneuji2, Tadami Matsumoto2, Kazuhiro Shintani3.   

Abstract

There is a rising concern about squeaking in ceramic-on-ceramic total hip arthroplasty (THA). In pin-on-disc testing of a delta-delta coupling, we reproduced squeaking and observed microcracks on worn surfaces. We used a pin-on-disc machine and made discs and pins by cutting delta ceramic to a diameter of 40 mm (D-D). Cross-linked polyethylene was used for a comparison disc (D-P). We performed the same test using another D-D coupling specimen to confirm reproducibility. Squeaking in the D-D specimen was reproduced in wet conditions, though it was not found in the D-P specimen. Fast Fourier transform analysis showed a peak frequency for squeaking of 2794 Hz. The noise occurred at about 6.6 km of sliding distance. Scanning electron microscopy revealed that the worn surface of the pin in D-D at 10.8 km of sliding distance had some microcracks. However, there was no obvious damage to the worn surface of the pin in D-P at the same sliding distance. We confirmed the reproducibility of these findings, obtaining similar results, including squeaking, from another D-D coupling specimen. Our findings show that squeaking may occur in THA using delta ceramic bearings even if implants are placed to avoid extra-articular impingement of the femoral neck. Although the clinical relevance of microcracks is unknown, they may affect long-term outcomes in THA using delta ceramic bearings.

Entities:  

Keywords:  Delta ceramics; Microcracks; Squeaking; Total hip arthroplasty

Mesh:

Year:  2016        PMID: 26971327     DOI: 10.1007/s00590-016-1748-0

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  21 in total

1.  Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions.

Authors:  T Stewart; J Tipper; R Streicher; E Ingham; J Fisher
Journal:  J Mater Sci Mater Med       Date:  2001 Oct-Dec       Impact factor: 3.896

2.  Retrieval analysis of squeaking alumina ceramic-on-ceramic bearings.

Authors:  W L Walter; S M Kurtz; C Esposito; W Hozack; K G Holley; J P Garino; M A Tuke
Journal:  J Bone Joint Surg Br       Date:  2011-12

Review 3.  Survival of hard-on-hard bearings in total hip arthroplasty: a systematic review.

Authors:  Michael G Zywiel; Siraj A Sayeed; Aaron J Johnson; Thomas P Schmalzried; Michael A Mont
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

Review 4.  Ceramics in total hip replacement.

Authors:  Didier Hannouche; Moussa Hamadouche; Rémy Nizard; Pascal Bizot; Alain Meunier; Laurent Sedel
Journal:  Clin Orthop Relat Res       Date:  2005-01       Impact factor: 4.176

5.  Influence of prosthetic design on squeaking after ceramic-on-ceramic total hip arthroplasty.

Authors:  Todd V Swanson; David J Peterson; Raghavendran Seethala; Ryan L Bliss; Calvin A Spellmon
Journal:  J Arthroplasty       Date:  2010-09       Impact factor: 4.757

6.  THA with Delta ceramic on ceramic: results of a multicenter investigational device exemption trial.

Authors:  William G Hamilton; James P McAuley; Douglas A Dennis; Jeffrey A Murphy; Thomas J Blumenfeld; Joel Politi
Journal:  Clin Orthop Relat Res       Date:  2009-09-19       Impact factor: 4.176

7.  The squeaking hip: a cause for concern-disagrees.

Authors:  Charlie C Yang; Raymond H Kim; Douglas A Dennis
Journal:  Orthopedics       Date:  2007-09       Impact factor: 1.390

8.  Natural history of squeaking after total hip arthroplasty.

Authors:  Camilo Restrepo; Wadih Y Matar; Javad Parvizi; Richard H Rothman; William J Hozack
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

9.  Three-dimensional modeling of in vitro hip kinematics under micro-separation regime for ceramic on ceramic total hip prosthesis: an analysis of vibration and noise.

Authors:  Elhadi Sariali; Todd Stewart; Zongmin Jin; John Fisher
Journal:  J Biomech       Date:  2009-11-04       Impact factor: 2.712

10.  The incidence of noise generation arising from the large-diameter Delta Motion ceramic total hip bearing.

Authors:  S M McDonnell; G Boyce; J Baré; D Young; A J Shimmin
Journal:  Bone Joint J       Date:  2013-02       Impact factor: 5.082

View more
  2 in total

1.  Preventing ceramic liner fracture after Delta ceramic-on-ceramic total hip arthroplasty.

Authors:  Young-Kyun Lee; Jae-Young Lim; Yong-Chan Ha; Tae-Young Kim; Woon-Hwa Jung; Kyung-Hoi Koo
Journal:  Arch Orthop Trauma Surg       Date:  2020-06-12       Impact factor: 3.067

2.  Arthroscopic evaluation and treatment of a squeaking hip. A case report.

Authors:  Jonathan Bellity; Marc Elkaïm; Didier Hannouche; Rémy Nizard
Journal:  BMC Musculoskelet Disord       Date:  2020-12-03       Impact factor: 2.362

  2 in total