Magnetic resonance imaging evaluation of intervertebral test spacers: an experimental comparison of magnesium versus titanium and carbon fiber reinforced polymers as biomaterials.

T Ernstberger1, G Buchhorn, G Heidrich.   

Abstract

INTRODUCTION: Intervertebral spacers are made of different materials, which can affect the postfusion magnetic resonance imaging (MRI) scans. Susceptibility artifacts, especially for metallic implants, can decrease the image quality. This study aimed to determine whether magnesium as a lightweight and biocompatible metal is suitable as a biomaterial for spinal implants based on its MRI artifacting behavior.
MATERIALS AND METHODS: To compare artifacting behaviors, we implanted into one porcine cadaveric spine different test spacers made of magnesium, titanium, and CFRP. All test spacers were scanned using two T1-TSE MRI sequences. The artifact dimensions were traced on all scans and statistically analyzed.
RESULTS: The total artifact volume and median artifact area of the titanium spacers were statistically significantly larger than magnesium spacers (P < 0.001), while magnesium and CFRP spacers produced almost identical artifacting behaviors (P > 0.05).
CONCLUSION: Our results suggest that spinal implants made with magnesium alloys will behave more like CFRP devices in MRI scans.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19693644      PMCID: PMC3128752          DOI: 10.1007/s11845-009-0394-5

Source DB:  PubMed          Journal:  Ir J Med Sci        ISSN: 0021-1265            Impact factor:   1.568


Introduction

Spinal fusion devices such as implantable interbody spacers are well-established and routinely used by spine surgeons to keep adjacent vertebrae spaced apart while bone ingrowth and fusion take place. Such spacers also provide weight-bearing support between adjacent vertebrae. The principal state-of-the-art spinal implants are made from titanium alloys and carbon fiber-reinforced polymers (CFRP). These biomaterials have enjoyed clinical success and rapid widespread use by improving patient outcomes. However, these materials have clinical and radiological limitations. Titanium is an excellently bioinert material that exhibits high biocompatibility. Titanium spacers produce good bone ingrowth without bone grafting. However, in magnetic resonance imaging (MRI) studies, titanium-based implants tend to cause distortion of the magnetic field which may obscure normal regional anatomy [1]. These properties pose difficulties in the postoperative MRI follow-up and evaluation of the fusion process due to the artifacting of its causes [5, 6]. The other principal material used for spacers consists of CFRP. Spacers made of this non-metallic biomaterial are not associated with the postoperative diagnostic problems of titanium because carbon produces a very low rate of artifact reactions and its radiolucency properties allow easier evaluation of the fusion process by MRI [3]. Carbon’s modulus of elasticity affords good load-bearing with sufficient hardness. But unlike titanium, carbon spacers undergo poor osteointegration because a soft tissue interface develops around the material surface that prevents direct ingrowth of bone. As a result, carbon spacers have to be filled with bone allografts to achieve long-term stability [1, 3]. CFRP implants have, therefore, been reviewed very critically in the literature [13]. Surgeons, over a century ago, recognized the potential of the lightweight metal magnesium as a biocompatible, osteoconductive, degradable implant material [7]. In 1907, Lambotte [7] was the first to introduce magnesium-based orthopedic devices; using a pure magnesium plate; he secured a bone fracture of the lower leg with gold-plated nails. A half a century later, magnesium-based metals were reported to have osteoconductive bioactivity and produce a more rapid formation of hard callus when used to support fractures in humans [16, 21]. The large amounts of evidence supporting the clinical advantages of magnesium have been summarized in a recent review paper [14]. None of the studies to date have yet investigated the diagnostic behavior of magnesium in MRI. This situation motivated us to determine whether magnesium is a suitable biomaterial for spinal implants by studying its MRI artifacting behavior.

Materials and methods

To evaluate the behavior of spacers made with a magnesium alloy, we compared their artifacting in diagnostic MRI scans with that of spacers made of a conventional titanium alloy and of CFRP. We consecutively implanted three spacers made of each of the three biomaterials dimensioned in small, medium, and large sizes in one cadaveric spine of a Gottingen mini pig (Figs. 1a–c, 2). The three spacers in group I were made of a magnesiumaluminiummanganese alloy (MgAlMn50), the three in group II of a titanium–aluminum–vanadium alloy (TiAl6V4), and those in group III of a carbon fiber-reinforced polymer (CFRP).
Fig. 1

Cylindrical test implants. a Magnesium (implant group I), b titanium (implant group II), c CFRP (implant group III)

Fig. 2

Cadavaric porcine spine model with an implanted medium titanium test cylinder

Cylindrical test implants. a Magnesium (implant group I), b titanium (implant group II), c CFRP (implant group III) Cadavaric porcine spine model with an implanted medium titanium test cylinder Table 1 presents the implant characteristics. A cylinder was chosen as for spacer shape because cylinders have demonstrated lowest rate of MRI artifacting behavior [4]. The spacer sizes—small, medium, and large—were dimensioned the same for each group (height in cm × base area in cm2); and their implant volume (IV) in cm3 and cross sectional area (CSA) in cm² was calculated for each size (Table 1). The spacer sizes were dimensioned as listed after a Newman–Keuls multiple comparison analysis showed that the selected sizes would produce significantly different artifacting behaviors (P < 0.001). Thus, a total of nine individual spacers were implanted, scanned by MRI, and evaluated for their artifacting behavior on the scans.
Table 1

Spacer dimensions

Sizes for all groupsDimensionsHeight × base area (cm × cm2)Cross sectional area (CSA, cm2)Implant volume (IV, cm3)
Small1.5 × 0.781.51.2
Medium2.0 × 1.132.42.3
Large2.5 × 1.543.83.5
Spacer dimensions

Spacer implantation

For each serial MRI study, the cylindrical implant was placed exactly between two adjacent vertebrae of the cadaveric porcine spine. The spine with implant was then completely packed in a soft-tissue mass and placed in a plastic container [4]. To create comparable trial conditions, markings were drawn on the container wall to demarcate the vertebrae and implant positions. These demarcations were used to define the median artifact area (MAA). The container with the spine implanted with each spacer was examined by serial MRI (Fig. 3).
Fig. 3

Median MRI artifact range depicted in a selection of three large test implants

Median MRI artifact range depicted in a selection of three large test implants

Magnetic resonance imaging

Magnetic resonance imaging was performed with a 1.5 T MRI (Magnetom Symphony, Siemens AG Medical Solutions, Erlangen, Germany). The T1w-TSE sequences were used to acquire a slice thickness of 3 mm (Fig. 3a–c) which included a first sequence (TR 600; TE 14; flip angle 15; band width 150), and a second sequence (TR 2,260, TE 14, flip angle 15, band width 150). We selected a matrix of 512 × 512 pixels combined with a field of view (FOV) of 500 mm. The T1w-TSE sequence has been established to produce best imaging results for implants and the least amount of intrinsic artifacting [3, 5, 6, 8–10, 12, 20]. Using a current version of DICOM reader software, one author (TE) measured the artifact area on the scan of each of the nine implants six times, i.e., a total of 54 individual tracings were recorded and analyzed. The measurements started with the slice with the first artifacting reaction and ended with the last slice exhibiting an artifact reaction. Corresponding to the respective implant’s CSA, the middle slice of all slices exhibiting artifact reactions was defined as the MAA for each implant. To calculate the total artifact volume (TAV) for each spacer, all artifact areas measured for that spacer were added and multiplied by the slice thickness of 3 mm according to the multisection slice technique described by Debatin et al. [2]. The ratio of CSA to MAA and the ratio of IV to TAV were calculated and presented in tables (Table 1).

Statistical analysis

Newman–Keuls multiple comparisons were used to calculate intragroup differences in TAV and MAA (Table 2). T test correlations were performed to determine any intergroup differences regarding the implant materials (Table 2). A P value of <0.05 indicated a significant difference between the means of any two groups.
Table 2

Intragroup comparison of target variables

Spacer materialSizeMAAa cm2 Mean ± SDRatioCSA:MAATAVa cm2 Mean ± SDRatioIV:TAV
Group I MgAlMn50 (n = 3)Small1.91 ± 0.041:1.31.83 ± 0.091:1.5
Medium3.26 ± 0.061:1.44.17 ± 0.091:1.8
Large4.06 ± 0.071:1.25.08 ± 0.151:1.3
Group II TiAl6V4 (n = 3)Small3.26 ± 0.041:2.25.71 ± 0.091:4.8
Medium4.61 ± 0.231:1.99.32 ± 0.101:4.1
Large5.54 ± 0.041:1.610.84 ± 0.131:2.9
Group III CFRP (n = 3)Small1.89 ± 0.071:1.31.81 ± 0.071:1.5
Medium3.18 ± 0.061:1.34.09 ± 0.111:1.7
Large4.06 ± 0.131:1.25.08 ± 0.131:1.3

CSA cross sectional area, MAA median artifact area, IV implant volume, TAV total artifact volume, SD standard deviation

aNewman–Keuls multiple comparison analysis P < 0.001

Intragroup comparison of target variables CSA cross sectional area, MAA median artifact area, IV implant volume, TAV total artifact volume, SD standard deviation aNewman–Keuls multiple comparison analysis P < 0.001

Results

Table 1 presents the spacer dimensions. Table 2 shows the intragroup comparisons of target variables. Table 3 lists the results of the intergroup t test correlations between TAV and MAA in relation to spacer material. Mean artifacting behavior increased with spacer size. When magnesium was compared with titanium, there were significant differences in both MAA and TAV. When magnesium was compared with carbon, the differences were not significant. In fact, magnesium produces an artifacting behavior very similar to that of CFRP.
Table 3

Intergroup comparisons of artifacting behavior by t test correlation

Spacer materialSize P valuea
MAATAV
Group I versus group IISmall≤0.001≤0.001
Medium≤0.001≤0.001
Large≤0.001≤0.001
Group I versus group IIISmall0.590.61
Medium0.090.26
Large1.00.96

MAA median artifact area, TAV total artifact volume

aSignificance level P < 0.05

Intergroup comparisons of artifacting behavior by t test correlation MAA median artifact area, TAV total artifact volume aSignificance level P < 0.05

Discussion

Spinal surgeons have not stopped searching for the optimum spacer material that combines high biocompatibility with artifact-free MRI imaging behavior in the implant environment. This study conducted to determine whether cylindrical spacers made of the biomaterial magnesium are suitable as spinal implants by comparing their MRI artifacting with that of identically dimensioned spacers made of a titanium alloy and a carbon fiber-reinforced polymer. In radiological spinal diagnostics, MRI is highly effective for clarifying postfusion questions regarding osseus and soft-tissue structures in relation to implant position. A comparative in vitro study shows that MRI has a higher sensitivity than CT in detecting osseus changes in the implant’s direct surroundings [19]. Moreover, MRI is well suited to demonstrate myelopathies, inflammatory and infectious processes, and any neurodegenerative changes. The MRI imaging behavior of spinal implants is obviously well documented in the literature [8, 10–12, 15, 17, 18, 20]. However, the aims of the published studies differed in that most focused on determining sequence-related artifact size. In a phantom study by Rudisch et al. [11], the relevance of metallic artifacts and implant-related characteristics, such as implant material and position, was demonstrated in addition to effects caused by the selected MRI sequence. In materials with a higher magnetizability like titanium alloys, implant shape additionally has an effect on the range of MRI artifacts [4]. The results of this comparative study showed that implant material and volume both affected the MRI artifacting behavior of our cylindrical test spacers. It was also noted that the smaller the implant size, the smaller was the range of susceptibility artifacts produced. The ratios calculated in Table 2 prove that the magnesium metal alloy exhibited behavior artifacting that was more like a non-metal. Our results confirm previous findings that MRI artifacting caused by solid implants is influenced by implant material, volume, and shape [4]. Judging from its non-metal-like MRI artifacting behavior alone, magnesium would appear to be a more suitable biomaterial for spinal implants than titanium. Given its osseoconductive potential as a metal [7], implant alloys made with magnesium would combine the advantages to the two principal spacer materials currently used, but without their limitations, at least in terms of MRI artifacting. Hence, magnesium alloys may show promise as spinal implants.
  18 in total

Review 1.  The postoperative spine.

Authors:  C B Henk; W Brodner; S Grampp; M Breitenseher; M Thurnher; G H Mostbeck; H Imhof
Journal:  Top Magn Reson Imaging       Date:  1999-08

2.  [Osseontegration of trial implants of carbon fiber reinforced plastics].

Authors:  U Schreiner; M Schwarz; G Scheller; H Schroeder-Boersch; L Jani
Journal:  Z Orthop Ihre Grenzgeb       Date:  2000 Nov-Dec

3.  [Artefacts and ferromagnetism dependent on different metal alloys in magnetic resonance imaging. An experimental study].

Authors:  M Thomsen; U Schneider; S J Breusch; J Hansmann; M Freund
Journal:  Orthopade       Date:  2001-08       Impact factor: 1.087

Review 4.  Review article: MRI of the postoperative lumbar spine.

Authors:  J W M Van Goethem; P M Parizel; J R Jinkins
Journal:  Neuroradiology       Date:  2002-08-10       Impact factor: 2.804

5.  [Influence of sequence type on the extent of the susceptibility artifact in MRI--a shoulder specimen study after suture anchor repair].

Authors:  T Herold; W C Caro; G Heers; L Perlick; J Grifka; S Feuerbach; W Nitz; M Lenhart
Journal:  Rofo       Date:  2004-09

6.  Magnetic resonance imaging--cardiac ejection fraction measurements. Phantom study comparing four different methods.

Authors:  J F Debatin; S N Nadel; H D Sostman; C E Spritzer; A J Evans; T M Grist
Journal:  Invest Radiol       Date:  1992-03       Impact factor: 6.016

7.  A carbon fiber implant to aid interbody lumbar fusion. Two-year clinical results in the first 26 patients.

Authors:  J W Brantigan; A D Steffee
Journal:  Spine (Phila Pa 1976)       Date:  1993-10-15       Impact factor: 3.468

8.  Postimplantation MRI with cylindric and cubic intervertebral test implants: evaluation of implant shape, material, and volume in MRI artifacting--an in vitro study.

Authors:  Thorsten Ernstberger; Gabert Heidrich; Gottfried Buchhorn
Journal:  Spine J       Date:  2006-11-20       Impact factor: 4.166

9.  Magnetic resonance imaging evaluation of the spine with metal implants. General safety and superior imaging with titanium.

Authors:  R Rupp; N A Ebraheim; E R Savolaine; W T Jackson
Journal:  Spine (Phila Pa 1976)       Date:  1993-03-01       Impact factor: 3.468

10.  A comparative study of MR imaging profile of titanium pedicle screws.

Authors:  A S Malik; O Boyko; N Aktar; W F Young
Journal:  Acta Radiol       Date:  2001-05       Impact factor: 1.701

View more
  4 in total

1.  Positron emission tomography-computed tomography in the assessment of viability of femoral head in acetabular fractures.

Authors:  Pavan Belehalli; Malhar Kumar; Bangalore Prakash; Lokesh Veerappa
Journal:  Int Orthop       Date:  2014-01-16       Impact factor: 3.075

2.  MR-compatible antibiotic interlocked nail fabrication for the management of long bone infections: first case report of a new technique.

Authors:  Cyril Mauffrey; George W Chaus; Nathan Butler; Heather Young
Journal:  Patient Saf Surg       Date:  2014-03-17

3.  Fabrication and Characterization of Carbon Fiber-Reinforced Nano-Hydroxyapatite/Polyamide46 Biocomposite for Bone Substitute.

Authors:  Zhennan Deng; Hongjuan Han; Jingyuan Yang; Yuanyuan Li; Shengnan Du; Jianfeng Ma
Journal:  Med Sci Monit       Date:  2017-05-24

4.  Biodegradable magnesium Herbert screw - image quality and artifacts with radiography, CT and MRI.

Authors:  Lena Sonnow; Sören Könneker; Peter M Vogt; Frank Wacker; Christian von Falck
Journal:  BMC Med Imaging       Date:  2017-02-14       Impact factor: 1.930

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.