Lena Sonnow1, Sören Könneker2, Peter M Vogt2, Frank Wacker3, Christian von Falck3.
Abstract
BACKGROUND: Magnesium alloys have recently been rediscovered as biodegradable implants in musculoskeletal surgery. This study is an ex-vivo trial to evaluate the imaging characteristics of magnesium implants in different imaging modalities as compared to conventional metallic implants.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28196474 PMCID: PMC5310087 DOI: 10.1186/s12880-017-0187-7
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
MRI scanning parameters
| Sequence | TR (ms) | TE (ms) | Flip Angle (°) | Band-width (Hz/pix) | Slice thickness (mm) | Slice spacing (mm) | FOV (mm) |
|---|---|---|---|---|---|---|---|
| Avanto (1.5 T): | |||||||
| PDw TSE (FS) | 2800 | 38 | 160 | 85 | 2.0 | 2.2 | 140 |
| T1w TSE | 571 | 13 | 180 | 130 | 2.0 | 2.2 | 140 |
| Skyra (3 T): | |||||||
| PDw TSE (FS) | 2200 | 31 | 150 | 200 | 2.0 | 2.2 | 200 |
| T1w TSE | 869 | 11 | 150 | 257 | 2.0 | 2.2 | 200 |
| PDw TSE WARP (FS) | 2200 | 31 | 150 | 200 | 2.0 | 2.2 | 200 |
Artifact size of the solely scanned screws
| Imaging modality | Mg: mean artifact size [mm] | Ti: mean artifact size [mm] | Mean difference in artifact size [mm] | T-value |
|
|---|---|---|---|---|---|
| DX | 4.5 ± 0.3 | 5.2 ± 0.2 | 0.7 ± 0.2 | 4.1 | <0.005 |
| MDCT | 5.7 ± 0.4 | 11.9 ± 0.5 | 6.2 ± 0.2 | 27.0 | <0.001 |
| FPCT | 5.5 ± 0.2 | 11.4 ± 0.5 | 5.9 ± 0.2 | 31.8 | <0.001 |
| 3 T MRI | |||||
| PDw TSE | 5.2 ± 0.7 | 7.9 ± 0.5 | 2.6 ± 0.3 | 9.1 | <0.001 |
| T1w TSE | 8.4 ± 0.7 | 12.9 ± 1.0 | 4.5 ± 0.4 | 10.6 | <0.001 |
| PDw TSE WARP | 5.0 ± 0.7 | 6.2 ± 0.5 | 1.1 ± 0.3 | 3.6 | <0.005 |
Fig. 1Mg-screw (top) and Ti-screw (bottom) in FPCT. The screw axis aligns to the z-axis of the scanner (0° position)
Artifact size of the screws implanted in chicken tibia in different screw positions (parallel (0°) and perpendicular (90°) to the longitudinal scanner axis) and different modalities
| Imaging modality | Mg: mean artifact size [mm] | Ti: mean artifact size [mm] | Mean difference in artifact size [mm] | T-value |
|
|---|---|---|---|---|---|
| DX | 0a | 3.9 ± 0.3 | 3.9 ± 0.2 | 21.0 | <0.001 |
| MDCT | |||||
| 0° | 0a | 14.1 ± 1.2 | 14.1 ± 1.2 | 45.5 | <0.001 |
| 90° | 0a | 19.7 ± 1.4 | 19.7 ± 1.4 | 54.5 | <0.001 |
| FPCT | |||||
| 0° | 5.7 ± 0.3 | 14.1 ± 0.7 | 8.5 ± 0.2 | 45.1 | <0.001 |
| 90° | 6.9 ± 0.4 | 27.0 ± 2.4 | 20.1 ± 0.6 | 32.0 | <0.001 |
| 1.5 T | |||||
| PDw TSE (FS) | 4.6 ± 0.5 | 6.2 ± 0.2 | 1.6 ± 0.1 | 12.0 | <0.001 |
| 90° | 7.8 ± 0.3 | 12.1 ± 0.1 | 4.3 ± 0.1 | 53.7 | <0.001 |
| 1.5 T | |||||
| T1w TSE | |||||
| 0° | 3.5 ± 0.4 | 5.3 ± 0.2 | 1.8 ± 0.1 | 15.6 | <0.001 |
| 90° | 4.7 ± 0.2 | 9.0 ± 0.2 | 4.3 ± 0.1 | 58.9 | <0.001 |
| 3 T | |||||
| PDw TSE (FS) | |||||
| 0° | 4.7 ± 0.4 | 8.6 ± 0.2 | 3.9 ± 0.1 | 33.8 | <0.001 |
| 90° | 8.4 ± 0.2 | 15.4 ± 0.3 | 7.0 ± 0.1 | 75.2 | <0.001 |
| 3 T | |||||
| T1w TSE | |||||
| 0° | 4.5 ± 0.3 | 8.2 ± 0.2 | 3.7 ± 0.1 | 39.7 | <0.001 |
| 90° | 6.7 ± 0.2 | 10.3 ± 0.4 | 3.6 ± 0.1 | 31.2 | <0.001 |
| 3 T | |||||
| PDw TSE WARP (FS) | |||||
| 0° | 4.7 ± 0.3 | 8.5 ± 0.2 | 3.8 ± 0.1 | 40.8 | <0.001 |
| 90° | 8.3 ± 0.2 | 13.4 ± 0.5 | 5.1 ± 0.1 | 36.7 | <0.001 |
aIn DX and MDCT the artifacts around the magnesium screw in chicken tibia were of such a small magnitude that they were innumerable and had to be considered zero
Fig. 2Artifacts (mm) in different modalities with a screw position at 0°
Fig. 3Artifacts (mm) in different modalities with a screw position at 90°
Fig. 4Mg-screw (top) and Ti-screw (bottom) in MDCT in the 0° and 90° degree positions
Fig. 5Mg-screw (top) and Ti-screw (bottom) in FPCT in the 0° and 90° degree positions
Fig. 6Artifact appearance in 3 T MRI PDw TSE (FS) in two positions of the screw (0°/90°)
Fig. 7Magnesium and titanium screws in two positions (0°/90°) with a WARP application