Postimplantation MRI with cylindric and cubic intervertebral test implants: evaluation of implant shape, material, and volume in MRI artifacting--an in vitro study.

Thorsten Ernstberger1, Gabert Heidrich, Gottfried Buchhorn.   

Abstract

BACKGROUND CONTEXT: Interbody spacers for anterior spine fusion are made of different materials, which can affect the postfusion magnetic resonance imaging (MRI) scans. Susceptibility artifacts specially for metallic implants can decrease the image quality.
PURPOSE: This study focused on the influence of determined implant parameters like shape, implant volume, and implant material in MRI artifacting with regard to solid geometrical titanium and carbon test spacers. STUDY DESIGN/
SETTING: A comparative study on an in vitro spine model. The study was performed at a university research facility.
METHODS: In this study we evaluated the postimplantation MRI scans of three determined cuboids and six cylinders, which were made of titanium alloy and carbon fiber reinforced polymer. MRI scans were carried out by using T1 turbo spin echo (TSE) sequences. The total artifact volume (TAV) as well as median artifact area (MAA) were calculated. Additionally the implant volume (IV)/TAV and cross-sectional area (CSA)/MAA relation were determined. Statistical analyses were calculated with the t test and Newman-Keuls test for multiple comparisons.
RESULTS: Considering all test implants with an increasing implant size, the TAV and the MAA became significantly larger (p<.001) with simultaneous reduction of the IV/TAV and CSA/MAA relation. In contrast to the carbon implant group, for titanium cylinders with an implant volume equivalent to the cuboids significant differences in MRI artifacting (p<.05) could be demonstrated.
CONCLUSIONS: Susceptibility artifacts were clearly affected by the implant material, shape, and implant volume. Independent of the implant material, with regard to a more advantageous IV/TAV and CSA/MAA relation, for larger implants the artifact rate was more limited to the implant's direct surroundings. On the basis of a high magnetizability for titanium implants with an equal IV, the range of implant-related susceptibility artifacts was influenced by the material as well as implant shape in favor of cylindric implants. According to the very low MRI artifact rate of carbon implants, the implant shape did not have any significant effect on the artifact behavior.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17482121     DOI: 10.1016/j.spinee.2006.03.016

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  7 in total

1.  [Effect of oral alloy post and core on magnetic resonance imaging].

Authors:  Yin Fang; Li Xin; Ma Qingyun; Wang Lu; Song Ye
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2014-12

2.  Metal-induced artifacts in computed tomography and magnetic resonance imaging: comparison of a biodegradable magnesium alloy versus titanium and stainless steel controls.

Authors:  Lukas Filli; Roger Luechinger; Thomas Frauenfelder; Stefan Beck; Roman Guggenberger; Nadja Farshad-Amacker; Gustav Andreisek
Journal:  Skeletal Radiol       Date:  2014-11-23       Impact factor: 2.199

3.  Implant detectibility of intervertebral disc spacers in post fusion MRI: evaluation of the MRI scan quality by using a scoring system--an in vitro study.

Authors:  Thorsten Ernstberger; Gabert Heidrich; Wolfgang Schultz; Eckhardt Grabbe
Journal:  Neuroradiology       Date:  2006-11-04       Impact factor: 2.804

4.  Susceptibility artifacts induced by crowns of different materials with prepared teeth and titanium implants in magnetic resonance imaging.

Authors:  Xiaomeng Gao; Qianbing Wan; Qingping Gao
Journal:  Sci Rep       Date:  2022-01-10       Impact factor: 4.379

5.  Artifacts in spine magnetic resonance imaging due to different intervertebral test spacers: an in vitro evaluation of magnesium versus titanium and carbon-fiber-reinforced polymers as biomaterials.

Authors:  Thorsten Ernstberger; Gottfried Buchhorn; Gabert Heidrich
Journal:  Neuroradiology       Date:  2009-05-26       Impact factor: 2.804

6.  Magnetic resonance imaging evaluation of intervertebral test spacers: an experimental comparison of magnesium versus titanium and carbon fiber reinforced polymers as biomaterials.

Authors:  T Ernstberger; G Buchhorn; G Heidrich
Journal:  Ir J Med Sci       Date:  2009-08-20       Impact factor: 1.568

7.  Influence of orthodontic appliance-derived artifacts on 3-T MRI movies.

Authors:  Erika Ozawa; Ei-Ichi Honda; Kulthida Nunthayanon Parakonthun; Hiroko Ohmori; Kazuo Shimazaki; Tohru Kurabayashi; Takashi Ono
Journal:  Prog Orthod       Date:  2018-02-19       Impact factor: 2.750

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.