A comparative study of MR imaging profile of titanium pedicle screws.

A S Malik1, O Boyko, N Aktar, W F Young.   

Abstract

OBJECTIVE: We compared the MR imaging profile of three different types of titanium pedicle screw implants in common usage in a human cadaveric model. We additionally compared the change in temperature during imaging among three constructs.
MATERIAL AND METHODS: Titanium-based lumbar pedicle screw/rod constructs from three manufacturers were implanted sequentially in a human cadaveric spine. MR imaging was then performed using both conventional spin-echo sequences and advanced imaging pulse sequences. Changes in tissue temperature were also measured during imaging to assess differences among the various implants. MR images were compared in a blinded fashion by two neuroradiologists.
RESULTS: No significant differences in imaging profile were noted between the three types of titanium implants with regards to their MR artifact profile. Fast spin-echo sequences led to a decrease in perceptible MR artifacts. Moreover, there were no significant differences in temperature increase among the three manufacturers (mean increase 0.5 degrees C) during imaging.
CONCLUSION: Slight differences in the percentage of titanium among the three pedicle screw systems does not appear to result in artifact differences during MR imaging. Therefore, with regard to imaging profile considerations, the three systems studied should be considered interchangeable.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11350287     DOI: 10.1080/028418501127346846

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.701


  7 in total

1.  The interobserver-validated relevance of intervertebral spacer materials in MRI artifacting.

Authors:  T Ernstberger; G Heidrich; T Bruening; S Krefft; G Buchhorn; H M Klinger
Journal:  Eur Spine J       Date:  2006-02-07       Impact factor: 3.134

2.  Implant detectibility of intervertebral disc spacers in post fusion MRI: evaluation of the MRI scan quality by using a scoring system--an in vitro study.

Authors:  Thorsten Ernstberger; Gabert Heidrich; Wolfgang Schultz; Eckhardt Grabbe
Journal:  Neuroradiology       Date:  2006-11-04       Impact factor: 2.804

3.  A standardized evaluation of artefacts from metallic compounds during fast MR imaging.

Authors:  Shumei Murakami; Rinus G Verdonschot; Miyoshi Kataoka; Naoya Kakimoto; Hiroaki Shimamoto; Sven Kreiborg
Journal:  Dentomaxillofac Radiol       Date:  2016-08-19       Impact factor: 2.419

4.  Safety and Feasibility of Topical Application of Limonene as a Massage Oil to the Breast.

Authors:  Jessica A Miller; Patricia A Thompson; Iman A Hakim; Ana Maria Lopez; Cynthia A Thomson; Wade Chew; Chiu-Hsieh Hsu; H-H Sherry Chow
Journal:  J Cancer Ther       Date:  2012-10

5.  Artificial cervical vertebra and intervertebral complex replacement through the anterior approach in animal model: a biomechanical and in vivo evaluation of a successful goat model.

Authors:  Jie Qin; Xijing He; Dong Wang; Peng Qi; Lei Guo; Sihua Huang; Xuan Cai; Haopeng Li; Rui Wang
Journal:  PLoS One       Date:  2012-12-27       Impact factor: 3.240

6.  Artifacts in spine magnetic resonance imaging due to different intervertebral test spacers: an in vitro evaluation of magnesium versus titanium and carbon-fiber-reinforced polymers as biomaterials.

Authors:  Thorsten Ernstberger; Gottfried Buchhorn; Gabert Heidrich
Journal:  Neuroradiology       Date:  2009-05-26       Impact factor: 2.804

7.  Magnetic resonance imaging evaluation of intervertebral test spacers: an experimental comparison of magnesium versus titanium and carbon fiber reinforced polymers as biomaterials.

Authors:  T Ernstberger; G Buchhorn; G Heidrich
Journal:  Ir J Med Sci       Date:  2009-08-20       Impact factor: 1.568

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.