OBJECTIVE: We wanted to evaluate the performance of prospective electrocardiogram (ECG)-gated axial scans for assessing coronary stents as compared with retrospective ECG-gated helical scans.Entities:
Keywords: Computed tomography (CT), angiography; Experimental studies; Multi-slice CT (MSCT); Stent
Mesh:
Substances:
Year: 2009 PMID: 19412509 PMCID: PMC2672176 DOI: 10.3348/kjr.2009.10.3.217
Source DB: PubMed Journal: Korean J Radiol ISSN: 1229-6929 Impact factor: 3.500
Commercial Name, Manufacturer, Material, Length, Strut Diameter and Weight Per Unit Length of Examined Stents
Fig. 1Vascular models of stenosis.
A. Stent was implanted into tube and contrast material diluted to 396 HU was filled in stented tubes as "patent" model (upper model). In stenotic models, smaller tube filled with contrast material diluted to 396 HU was fixed in stented tube that was filled with contrast material diluted to 79 HU (lower model).
B. Vascular model was fixed in polypropylene cylinder filled with salad oil in parallel with central axis of cylinder. Polypropylene cylinder was fixed in water-filled spherical styrol container, with their central axes overlapping.
Fig. 2Longitudinal and cross-sectional reformations of patent models. Window setting was 1,400 HU width and 500 HU center. FOV = field of view
Fig. 3Longitudinal and cross-sectional reformations of stenotic models. Window setting was 1,400 HU width and 500 HU center. FOV = field of view
Fig. 4Magnified views of cross-sectional reformations of patent and stenotic models. A. Bx-Velocity, B. Express2, C. Micro Driver. Window setting was 1,400 HU width and 500 HU center. FOV = field of view
Visual Evaluation of In-stent Stenosis for Vascular Models
Note.-FOV = field of view, ECG = electrocardiogram
Each stent was evaluated as patent, possibly stenotic or definitely stenotic based on visibility of luminal low attenuation area corresponding to neointima.
Diameter Measurement of Stent Lumen
Note.-FOV = field of view, ECG = electrocardiogram
Measurement error of stent lumen is given as mean ± standard deviation (mm). Numbers in parentheses are artificial lumen narrowing.
CT Attenuation Inside Stents of Vascular Model
Note.-FOV = field of view, ECG = electrocardiogram
CT attenuation of stent lumen is given as mean ± standard deviation (HU).
Fig. 5CT attenuation profile on cross section of stenotic models with using axial scan with 18-cm field of view, and axial and helical scans with 9-cm field of view. A. Bx-Velocity, B. Express2, C. Micro Driver. Profile curve passes through center of vascular model. FOV = field of view