Flat-panel detector computed tomography for the assessment of coronary artery stents: phantom study in comparison with 16-slice spiral computed tomography.

Andreas H Mahnken1, Tobias Seyfarth, Thomas Flohr, Christopher Herzog, Jochen Stahl, Sven Stanzel, Axel Kuettner, Joachim E Wildberger, Rolf W Günther.   

Abstract

PURPOSE: The evaluation of coronary artery stents is a major limitation of cardiac multislice spiral computed tomography (MSCT). The development of flat-panel detector computed tomography (FPCT) with truly isotropic spatial resolution may overcome this limitation. Thus, we evaluated the use of FPCT in comparison to MSCT for the assessment of coronary artery stents.
MATERIAL AND METHODS: Eight different coronary artery stents with a diameter of 3 mm each were placed in a static chest phantom. The phantom was positioned in the CT gantry at angles of 0 degrees , 45 degrees , and 90 degrees toward the z-axis and examined with the prototype of a FPCT (Siemens, Forchheim, Germany) and a commercially available 16-detector row MSCT (Sensation 16, Siemens). Slice thickness was 0.25 mm with FPCT whereas for MSCT, an effective slice thickness of 1 mm with a reconstruction increment of 0.5 mm was used. Image quality was assessed visually using a 5-point grading scale. Stent diameters were measured and compared using a repeated-measures analysis of variance.
RESULTS: When compared with MSCT, artificial lumen reduction was significantly less using FPCT. On average the visible stent lumen was reduced by 16.1% with FPCT, whereas the mean of the lumen reduction was 47.2% with 16-detector row MSCT. Visible lumen diameter as well as image noise significantly increased using FPCT (P < 0.001). With FPCT delineation of the different stent struts became possible.
CONCLUSION: FPCT proved to be superior when compared with 16-detector row MSCT for the in vitro assessment of coronary artery stents. Improved spatial resolution allows for a detailed assessment of the coronary artery stent lumen.

Mesh:

Year:  2005        PMID: 15597014

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  18 in total

1.  Angiographic CT after intravenous contrast agent application: A noninvasive follow-up tool after intracranial angioplasty and stenting.

Authors:  M-N Psychogios; P Schramm; J-H Buhk; A Xyda; K Gröschel; K Jung; M Knauth
Journal:  AJNR Am J Neuroradiol       Date:  2010-07-15       Impact factor: 3.825

2.  Non-invasive coronary angiography using multislice computed tomography.

Authors:  Jeffrey M Schussler; Paul A Grayburn
Journal:  Heart       Date:  2005-12-30       Impact factor: 5.994

3.  Effect of contrast concentration, tube potential and reconstruction kernels on MDCT evaluation of coronary stents: an in vitro study.

Authors:  Gopi Kiran Reddy Sirineni; Mannudeep K Kalra; Krishna Pottala; Sandra Waldrop; Mushabbar Syed; Stefan Tigges
Journal:  Int J Cardiovasc Imaging       Date:  2006-07-05       Impact factor: 2.357

4.  Evaluation of in-stent restenosis in proximal coronary arteries with multidetector computed tomography (MDCT).

Authors:  V Chabbert; D Carrie; M Bennaceur; E Maupas; V Lauwers; M Mhem; T Lhermusier; M Elbaz; F Joffre; H Rousseau; J Puel
Journal:  Eur Radiol       Date:  2006-11-18       Impact factor: 5.315

Review 5.  Cardiac CT: coronary arteries and beyond.

Authors:  Andreas H Mahnken; Georg Mühlenbruch; Rolf W Günther; Joachim E Wildberger
Journal:  Eur Radiol       Date:  2006-10-26       Impact factor: 5.315

Review 6.  Flat-detector computed tomography (FD-CT).

Authors:  Willi A Kalender; Yiannis Kyriakou
Journal:  Eur Radiol       Date:  2007-06-23       Impact factor: 5.315

7.  High-resolution ex vivo imaging of coronary artery stents using 64-slice computed tomography--initial experience.

Authors:  Carsten Rist; Konstantin Nikolaou; Thomas Flohr; Bernd J Wintersperger; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2006-03-02       Impact factor: 5.315

8.  Prospective ECG-triggered axial CT at 140-kV tube voltage improves coronary in-stent restenosis visibility at a lower radiation dose compared with conventional retrospective ECG-gated helical CT.

Authors:  Jun Horiguchi; Chikako Fujioka; Masao Kiguchi; Hideya Yamamoto; Toshiro Kitagawa; Shingo Kohno; Katsuhide Ito
Journal:  Eur Radiol       Date:  2009-05-09       Impact factor: 5.315

9.  Accuracy of dual-source computed tomography in quantitative assessment of low density coronary stenosis--a motion phantom study.

Authors:  Michael Toepker; Christopher L Schlett; Thomas Irlbeck; Amir A Mahabadi; Fabian Bamberg; Christiane Leidecker; Patrick Donnelly; Udo Hoffmann
Journal:  Eur Radiol       Date:  2009-09-02       Impact factor: 5.315

Review 10.  [Flat-detector computed tomography in diagnostic and interventional neuroradiology].

Authors:  T Struffert; A Doerfler
Journal:  Radiologe       Date:  2009-09       Impact factor: 0.635

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.